Speckle plethysmography (SPG): blood flow from laser speckle

SPG shines a laser into skin and measures how fast the speckle pattern blurs. Moving red cells cause the blur, so SPG follows blood flow, while PPG follows blood volume through the light it absorbs. It is a research technique studied for pulse, perfusion and cuffless blood pressure.

A simulated laser speckle pattern: random bright and dark grains on a dark red background.
Simulated laser speckle from static tissue.Moving blood blurs this pattern; SPG measures how much.

SPG in one minute

Speckle plethysmography (SPG) records the pulsing of blood flow, beat by beat, from the way laser light flickers after it scatters through skin. That is the key difference from ordinary photoplethysmography (PPG): PPG tracks how much blood is under the sensor (volume) from how much light it absorbs, usually with an LED, while SPG tracks how fast the blood is moving (flow) from how a coherent laser's speckle pattern blurs. The SPG vs PPG section below compares them side by side.

Shine a laser into tissue and the light that comes back forms a grainy pattern called speckle. Red cells moving through small vessels make that pattern change, so a camera exposure sees it blurred. More flow means more blur. Track the blur many times a second and you get a flow waveform that rises and falls with every heartbeat.

The wearable form of SPG was introduced in 2018. Shining one laser through a finger produced two co-registered waveforms at once: the familiar PPG from the light's intensity, which follows blood volume, and an SPG from the speckle dynamics, which follows blood flow. In 16 people the two waveforms were offset in time and had different harmonic content (Ghijsen et al., 2018). SPG borrows its physics from laser speckle contrast imaging, a lab method for mapping blood flow in brain, skin and eye research (Boas & Dunn, 2010).

What laser speckle is

A laser emits light of one wavelength with its waves in step, which is what physicists call coherent light. When that light enters skin it scatters off cell membranes, collagen and red cells many times. Every photon takes its own route, and every route has a different length. When the returning light lands on a detector, waves that travelled different distances arrive out of step and add up or cancel each other. The result is a random map of bright and dark grains: speckle.

This only works when the light stays coherent over the whole spread of path lengths through the tissue (Boas & Dunn, 2010). That is why the choice of laser matters: many common semiconductor laser diodes have a spectrum broad enough to wash out speckle contrast, and a linewidth under about 1 nm is recommended (Postnov et al., 2019). LEDs are not coherent, and in a direct comparison LED-based speckle signals had much lower quality than laser-based ones (Herranz Olazábal et al., 2022).

Where speckle comes from. Laser light enters the skin from the sensor side and scatters many times before some of it comes back out next to the source. Because every path is a different length, the returning waves add up into random bright and dark grains: a speckle pattern. The glow and photon paths come from a simple Monte Carlo light-scattering model, and the inset is simulated speckle, shown changing far more slowly than real speckle so the motion is visible. Conceptual drawing, not to scale.

Why moving blood blurs speckle

If nothing in the tissue moves, the speckle pattern holds still and a camera captures sharp grains however long it exposes. When red cells move, the path lengths change and the pattern shimmers. A camera exposure then averages several different patterns together, and the grains smear into a flatter grey. The faster the cells move, the faster the pattern changes and the stronger the blur in an exposure of a given length. Speckle contrast imaging typically uses exposures of about 1 to 10 ms (Boas & Dunn, 2010).

Speckle contrast, K=σ/⟨I⟩K = \sigma/\langle I\rangle

Blur is measured with one number, the speckle contrast KK. Take a small patch of pixels, compute the spread of their brightness (the standard deviation σ\sigma) and divide by their average brightness ⟨I⟩\langle I\rangle:

K=σ⟨I⟩K = \frac{\sigma}{\langle I \rangle}

Sharp, fully developed speckle has KK close to 1: the grains are as varied as their average. Completely blurred speckle has KK near 0: every pixel looks the same (Boas & Dunn, 2010). How KK falls with exposure time TT depends on the speckle decorrelation time τc\tau_c, the time it takes the pattern to change into a new one. A widely used model gives

K(T)=β1/2{τcT+τc22T2[exp⁡ ⁣(−2Tτc)−1]}1/2K(T) = \beta^{1/2}\left\{\frac{\tau_c}{T} + \frac{\tau_c^{2}}{2T^{2}}\left[\exp\!\left(-\frac{2T}{\tau_c}\right) - 1\right]\right\}^{1/2}

where β\beta is a constant set by the optics (how big a speckle grain is compared with a pixel, and polarization). In this model τc\tau_c is inversely proportional to the speed of the moving cells, so it is the link between blur and flow (Boas & Dunn, 2010). For exposures much longer than τc\tau_c the formula simplifies to

K2≈β τcT(T≫τc)K^{2} \approx \beta\,\frac{\tau_c}{T}\quad (T \gg \tau_c)

so 1/K21/K^{2} rises roughly in step with flow speed. Speckle contrast gives a relative flow index rather than an absolute flow in millilitres per minute, and turning it into quantitative flow takes extra modelling (Boas & Dunn, 2010). Try it below: the exposure slider integrates a simulated, changing speckle pattern for the chosen time, and the readout compares the simulated frame's contrast with the model above.

Static tissueK 1.00
Static tissueFully developed speckle, sharp at any exposure
Flowing · 5.0 msK 0.25
Flowing red cellsK = 0.25 simulated · 0.25 from Eq. 9
Speckle contrast K against exposure time. Static tissue stays at K = 1. For flowing cells with a simulated decorrelation time of 316 µs, K falls from about 1 at short exposures toward 0 at long exposures; the current exposure of 5.0 ms gives K = 0.25.00.51100 µs1.0 ms10 msExposure time T (log scale)Contrast K

Static, K = 1Flowing, Eq. 9Simulated frame

Histogram of pixel intensity divided by the mean. The static frame follows the negative exponential curve of fully developed speckle, with many dark pixels and a long bright tail. The flowing frame, averaged over the exposure, narrows around the mean.01234Pixel intensity ÷ meanShare

Static frameFlowing frameexp(−x), fully developed

Static preview. The simulator loads when this figure is on screen.
Exposure time decides how much blur you see. Left, static tissue keeps sharp speckle at any exposure. Middle, flowing red cells: the same pattern integrated over the chosen exposure. Right, contrast K against exposure from the model above, for the chosen flow speed, and below it the spread of pixel brightness: static speckle follows the negative-exponential curve of fully developed speckle, while averaging over the exposure pulls flowing pixels toward the mean. Each frame is simulated from random-phase speckle that decorrelates over τc; it is an educational simulation, not a measurement.

SPG vs PPG

PPG measures blood volume; SPG measures blood flow. PPG shines light into the skin and reads how much of it the blood absorbs: with each beat the vessels swell, more light is absorbed, and the detected intensity dips. SPG shines a coherent laser and reads how fast the speckle changes, which depends on how fast the red cells move. The same laser and camera can record both at once (Ghijsen et al., 2018; Schnitzler et al., 2024).

Simulated pulsatile amplitude in a skin model, % (Dunn et al., 2018)

  • PPG · 532 nm5.6%
  • PPG · 860 nm3.0%
  • SPG · 532 nm24%
  • SPG · 860 nm40%
Same skin, two different questions. PPG asks how much light the blood absorbs as its volume swells with each beat. SPG asks how fast the speckle pattern changes, which depends on how fast red cells move. In a skin simulation, the pulsatile SPG signal was several times larger than PPG at both wavelengths (Dunn et al., 2018). Waveform shapes and timing are illustrative; measured SPG and PPG pulses differ in timing and harmonics (Ghijsen et al., 2018).
Sourced differences between SPG and PPG. Each row cites the study behind it.
AspectPPGSPGSource
What it followsBlood volume, from light intensityBlood flow, from speckle dynamics(Ghijsen et al., 2018)
Light sourceAny light; only intensity is used (a laser works too)Coherent laser with a narrow linewidth (under about 1 nm recommended)(Ghijsen et al., 2018; Postnov et al., 2019)
Pulsatile amplitude in a skin simulation5.6% at 532 nm, 3.0% at 860 nm24% at 532 nm, 40% at 860 nm(Dunn et al., 2018)
Longer near-infrared wavelengthPulsatile modulation was smallerPulsatile modulation grew(Dunn et al., 2018)
Camera signal quality, reflection moderPPG SQI below 0.845Laser rSPG SQI above 0.965(Herranz Olazábal et al., 2022)
Transmission through a finger—Speckle washed out and SPG began to look like PPG(Herranz Olazábal et al., 2022)
Low flow (cold pressor test)Weaker signal, lower signal-to-noise ratio than SPGHigher, more stable signal-to-noise ratio(Herranz Olazábal et al., 2023)
Pulse shape vs finger arterial pressureDeviation 0.17Deviation 0.09(Herranz Olazábal et al., 2023)

The two signals are complementary rather than rivals. Because one tracks vessel expansion and the other tracks flow, their waveforms are offset in time (Ghijsen et al., 2018), and a single rolling-shutter camera frame can separate blood velocity from blood volume changes (Lee et al., 2024). For the PPG side of the story, see the guides to the perfusion index and the AC/DC ratio.

What SPG can measure

Everything below comes from research studies, most of them small. None of it makes SPG a diagnostic test.

Inside an SPG sensor

An SPG sensor needs three things. First, a coherent light source with a narrow spectrum (Postnov et al., 2019). Second, a detector that can see speckle statistics: usually a camera, whose linearity, read noise and quantization decide how accurately contrast is measured in low light (Cheng et al., 2024), although an ordinary photodiode with an integrating circuit can also work (Biswas et al., 2024). Third, processing that corrects contrast for camera shot noise and dark noise (Valdés et al., 2014).

Geometry matters too. In reflection, with source and detector side by side on the skin, laser SPG kept high signal quality; in transmission through a finger the speckle washed out (Herranz Olazábal et al., 2022). A compact research wearable already pairs a laser diode with a small camera to record PPG and SPG together (Schnitzler et al., 2024).

A plausible wrist layout. The coherent source and the detector both look down through a window that touches the skin, with an opaque barrier between them so only light that has travelled through tissue is measured. Research prototypes have paired a laser diode with a small camera for simultaneous PPG and SPG (Schnitzler et al., 2024), used a plain photodiode instead of a camera (Biswas et al., 2024), and compared lasers with LEDs at the same wavelengths (Herranz Olazábal et al., 2022). Conceptual drawing, not a specific device, not to scale; light paths come from a simplified Monte Carlo model.

Limits and open questions

Devices and companies

A short, sourced list. Regulatory facts link to the FDA database or the maker's own page, since they have no DOI. Listing is not an endorsement.

DeviceWhat it doesStatus and source
FlowMet-R (Laser Associated Sciences)Laser speckle blood-flow waveform at the toe (Razavi et al., 2021)FDA 510(k) cleared 27 Feb 2019, K182494
Openwater Open-MotionPulsed speckle contrast headset for brain blood flow, with an open SDK; an Openwater headset was used in a stroke study (Favilla et al., 2024)For research use only (product page, SDK)
Hemorai CSI patchWearable patch combining speckle flow with tissue spectroscopy, aimed at hemorrhageIn development, not FDA cleared (company page)
Armor Medical aRMORWearable laser speckle flow sensor for hemorrhage; published data so far are from swine (Bonetta-Misteli et al., 2023)Prototype (company page)
ContinUse Gili Pro BioSensorContactless spot-check heart and breathing rate from chest-wall vibration. This is speckle vibrometry, not SPGFDA De Novo DEN200038

Open datasets

  • Laser speckle-contrast images of human skin under physiological tests (Dremin et al.). Skin laser speckle contrast recordings during controlled respiration, breath holding, venous occlusion and spontaneous respiration. Four MATLAB files, about 0.9 GB. The record lists the tests and files but no participant metadata. CC BY 4.0. doi:10.5281/zenodo.4359074 · ChatPPG dataset card
  • High-density multi-channel SCOS cerebral blood flow data and code (BUNPC 2025-SCOS). Raw camera frames, processed data and analysis code behind the 50-channel SCOS brain-flow study: 15 people doing a Stroop task, plus a liquid flow phantom. About 3.9 GB. CC BY 4.0. doi:10.5281/zenodo.15881287 · paper: (Kim et al., 2025) · ChatPPG dataset card

For PPG data, see PPG datasets for blood pressure and the research atlas.

The evidence, paper by paper

All 33 summarized papers in the ChatPPG speckle plethysmography collection, grouped by question. Each links to its DOI.

Foundations and physics

Where SPG and its speckle-contrast relatives come from.

  • Wearable speckle plethysmography (SPG) for characterizing microvascular flow and resistance

    Ghijsen et al. · Biomedical Optics Express 9(8):3937, 2018 · doi:10.1364/BOE.9.003937 · ChatPPG card

    This paper introduced wearable speckle plethysmography (SPG). A single laser shining through a finger yields two co-registered waveforms: the classic PPG from light intensity (blood volume) and a new SPG from laser speckle dynamics (blood flow). In 16 people the two waveforms were offset in time and had different harmonic content, which the authors relate to microvascular flow and resistance.

  • Laser speckle contrast imaging in biomedical optics

    Boas & Dunn · Journal of Biomedical Optics 15(01):1, 2010 · doi:10.1117/1.3285504 · ChatPPG card

    The standard review of laser speckle contrast imaging: how blurring of a laser speckle pattern during a camera exposure encodes blood flow, how to make the measurement more quantitative, and how it has been used in brain, skin and eye research.

  • Speckle contrast optical spectroscopy, a non-invasive, diffuse optical method for measuring microvascular blood flow in tissue

    Valdés et al. · Biomedical Optics Express 5(8):2769, 2014 · doi:10.1364/BOE.5.002769 · ChatPPG card

    This paper introduced speckle contrast optical spectroscopy (SCOS), which measures deep-tissue blood flow by combining laser speckle contrast statistics with a photon-diffusion model. It was validated on phantoms and a human forearm, and it introduced corrections for camera shot noise and dark noise that later speckle methods depend on.

  • Deep tissue flowmetry based on diffuse speckle contrast analysis

    Bi et al. · Optics Letters 38(9):1401, 2013 · doi:10.1364/OL.38.001401 · ChatPPG card

    Diffuse speckle contrast analysis (DSCA) showed that deep-tissue blood-flow information similar to diffuse correlation spectroscopy can be obtained from a simple camera image of the speckle pattern instead of expensive single-photon counters. It was demonstrated on a flow phantom and during cuff occlusion in vivo.

  • Comparison of speckleplethysmographic (SPG) and photoplethysmographic (PPG) imaging by Monte Carlo simulations and in vivo measurements

    Dunn et al. · Biomedical Optics Express 9(9):4306, 2018 · doi:10.1364/BOE.9.004306 · ChatPPG card

    Monte Carlo simulations in an eight-layer skin model plus in vivo camera measurements compared speckle plethysmography (SPG) with PPG. Simulated pulsatile amplitude was 24 % (532 nm) and 40 % (860 nm) for SPG versus 5.6 % and 3.0 % for PPG, and both simulation and measurement point to a much larger signal-to-noise ratio for SPG. Unlike PPG, the SPG modulation grows at the longer near-infrared wavelength.

Light sources, cameras and processing

What the optics and the math need to get right.

Pulse, HRV, low perfusion and skin tone

Head-to-head comparisons with PPG and ECG.

  • Speckleplethysmographic (SPG) Estimation of Heart Rate Variability During an Orthostatic Challenge

    Dunn et al. · Scientific Reports 9(1):14079, 2019 · doi:10.1038/s41598-019-50526-0 · ChatPPG card

    Seventeen people did an orthostatic (posture) challenge while heart rate variability was measured with ECG, PPG and speckle plethysmography (SPG). SPG-derived HRV correlated highly with ECG in both time and frequency domains and was more accurate than PPG-derived HRV, which tends to degrade under physical stress.

  • Beat-to-beat Intervals of Speckle & Intensity-based Optical Plethysmograms compared to Electrocardiogram

    Herranz Olazábal et al. · Computing in cardiology, 2021 · doi:10.23919/cinc53138.2021.9662819 · ChatPPG card

    A camera filming a laser-lit finger (639 nm, 100 fps) produced a remote speckle plethysmogram (rSPG) whose beat-to-beat intervals were compared with ECG across 3,234 heartbeats in nine people. The contactless rSPG had smaller interval errors than a contact finger-clip PPG (mean absolute deviation 10.4 vs 14.2 ms).

  • Comparison between Speckle Plethysmography and Photoplethysmography during Cold Pressor Test Referenced to Finger Arterial Pressure

    Herranz Olazábal et al. · Sensors 23(11):5016, 2023 · doi:10.3390/s23115016 · ChatPPG card

    Ten people did a cold pressor test (hand in ice water) to cut finger blood flow while speckle plethysmography (SPG) and PPG were recorded from the same video at 639 and 850 nm. Both signals weakened, but SPG kept a significantly higher and more stable signal-to-noise ratio than PPG throughout and showed richer harmonics.

  • Effect of skin optical absorption on speckleplethysmographic (SPG) signals

    Rice et al. · Biomedical Optics Express 11(9):5352, 2020 · doi:10.1364/BOE.403501 · ChatPPG card

    Skin-simulating phantoms and a small clinical protocol tested whether skin pigmentation affects speckle plethysmography (SPG). SPG tracked known flow linearly (R² 0.990) and stayed linear with absorbing skin layers added (R² 0.996–0.999), and clinical readings showed no significant difference between lower- and higher-melanin groups.

  • SPECKLE-BASED MEASUREMENT OF MICROVASCULAR FUNCTION

    Maas et al. · Journal of Hypertension 44(Suppl 1):e179, 2026 · doi:10.1097/01.hjh.0001197056.64038.c5 · ChatPPG card

    An imec conference abstract describes a finger speckle device using green and infrared lasers that probe different skin depths. Pressing on the finger made the shallow (green) speckle pulse shrink or vanish while the deeper infrared signal changed less, and a ratio of the two, a micro-to-macrovascular index, dropped by a median 55 %. The authors propose it as a step toward non-invasive measurement of microvascular function.

Cuffless blood pressure (research)

Flow plus volume features for blood-pressure estimation. None of these are cleared devices.

  • Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time

    Herranz Olazábal et al. · Bioengineering 10(1):101, 2023 · doi:10.3390/bioengineering10010101 · ChatPPG card

    Remote speckle plethysmography (SPG) from a camera and a 639 nm laser was compared with a clinical finger-clip PPG, using finger arterial pressure as the reference. Across 2,599 beats the SPG pulse shape matched the arterial pressure waveform about twice as closely as PPG (deviation 0.09 vs 0.17). Pulse arrival times were similar for both.

  • Towards continuous blood pressure monitoring using speckle plethysmography

    Lorato et al. · Biophotonics in Exercise Science, Sports Medicine, Health Monitoring Technologies, and Wearables, 2024 · doi:10.1117/12.3001627 · ChatPPG card

    Ten people wore a contact speckle plethysmography (SPG) sensor alongside ECG, PPG and a reference blood pressure monitor. Pulse arrival time and waveform features from SPG correlated strongly with blood pressure, and the relationship varied less between people than it did for PPG, which suggests SPG might support a more general, less calibration-heavy BP model.

  • Speckle Plethysmograph-Based Blood Pressure Assessment

    Ellington et al. · Technologies 12(5):70, 2024 · doi:10.3390/technologies12050070 · ChatPPG card

    A prototype smartphone-based speckle plethysmography (SPG) system was paired with ECG to estimate systolic blood pressure from pulse arrival time in eight people while sitting, walking and running. SPG-based estimates were slightly better than PPG-based ones (resting error 12.4 vs 13.7 mmHg), and adding the previous BP reading as an input cut the error to 3.7 mmHg.

  • Simultaneous photoplethysmography and blood flow measurements towards the estimation of blood pressure using speckle contrast optical spectroscopy

    Garrett et al. · Biomedical Optics Express 14(4):1594, 2023 · doi:10.1364/BOE.482740 · ChatPPG card

    Speckle contrast optical spectroscopy (SCOS) records both a PPG-like blood-volume signal and a blood-flow index from the same light. On the finger and wrist of 13 people, blood-flow features correlated with blood pressure slightly better than PPG features, and features combining flow and volume correlated best (R = −0.59).

  • Speckle contrast optical spectroscopy for cuffless blood pressure estimation based on microvascular blood flow and volume oscillations

    Garrett et al. · Biomedical Optics Express 16(8):3004, 2025 · doi:10.1364/BOE.560022 · ChatPPG card

    High-speed (390 Hz) speckle contrast optical spectroscopy recorded blood flow and blood volume waveforms at two wavelengths on the wrist and finger. Feeding flow plus volume features into subject-specific machine-learning models estimated systolic and diastolic blood pressure with errors of about 2.9 and 2.1 mmHg (SD) across a wide BP range, a 31 % improvement in systolic estimation over PPG alone, and the gain held when people were re-measured weeks later.

  • Clinical-grade Cuffless Blood Pressure Monitoring via Deep-tissue Diffuse Speckle Pulsatile Flowmetry

    Choo et al. · medRxiv, 2026 · doi:10.64898/2026.06.19.26356074 · ChatPPG card

    A preprint from A*STAR Singapore tests diffuse speckle pulsatile flowmetry, a laser-speckle method that records deep-tissue blood flow alongside a PPG-like volume signal, for cuffless blood pressure. In 133 outpatients it reported mean absolute errors of 4.17 mmHg (systolic) and 2.42 mmHg (diastolic) against a cuff, meeting ISO 81060-2 accuracy criteria and British Hypertension Society grade A. Ablations showed the flow signal improved on PPG alone. It has not yet been peer reviewed.

  • Optical methods for cuffless blood pressure measurements

    Garrett et al. · Biophotonics Discovery 3(01):1, 2025 · doi:10.1117/1.BIOS.3.1.010901 · ChatPPG card

    A review of optical approaches to cuffless blood pressure: finger and wrist PPG, multiwavelength PPG, imaging PPG, and speckle contrast optical spectroscopy / speckle plethysmography (SCOS/SPG). It finds that most work uses finger PPG, that evaluation methods vary too much to compare devices, and that newer techniques like SCOS/SPG add physiological information that may improve accuracy if tested against consensus standards.

Peripheral flow, hemorrhage and the eye

Clinical-research uses of speckle flow outside the brain.

  • A Real-Time Blood Flow Measurement Device for Patients with Peripheral Artery Disease

    Razavi et al. · Journal of Vascular and Interventional Radiology 32(3):453–458, 2021 · doi:10.1016/j.jvir.2020.09.006 · ChatPPG card

    FlowMet, a toe-worn optical device that uses laser speckle dynamic light scattering to read blood flow, was tested in 167 limbs of 90 patients with suspected peripheral artery disease. Its waveforms resembled Doppler ultrasound, and its flow-waveform score predicted critical limb ischemia (AUC 0.83) about as well as the toe-brachial index and better than the ankle-brachial index.

  • Development and evaluation of a wearable peripheral vascular compensation sensor in a swine model of hemorrhage

    Bonetta-Misteli et al. · Biomedical Optics Express 14(10):5338–5357, 2023 · doi:10.1364/BOE.494720 · ChatPPG card

    A low-cost wearable laser speckle sensor measured peripheral blood flow (laser speckle flow index) to catch the vasoconstriction the body uses to compensate for bleeding. In six pigs with controlled blood loss, the flow index tracked blood-volume loss with a correlation of −0.95, outperforming the standard shock index, and tracked recovery during resuscitation. The work targets early detection of postpartum hemorrhage.

  • Characterization of laser speckle flowgraphy pulse waveform parameters for the evaluation of the optic nerve head and retinal circulation

    Enomoto et al. · Scientific Reports 11:6847, 2021 · doi:10.1038/s41598-021-86280-5 · ChatPPG card

    Laser speckle flowgraphy is a clinical eye-imaging method that reads blood flow in the optic nerve head from laser speckle blur. In 111 eyes of 86 healthy adults, the shape of its pulse waveform (skew, flow acceleration and related indices) related to systemic factors and retinal vessel measures, which supports waveform features as markers of ocular circulation.

Brain blood flow (SCOS)

Diffuse speckle contrast through the scalp and skull.

Adjacent: speckle vibrometry (not SPG)

These read surface motion from a defocused speckle pattern, not blood-flow decorrelation. Listed for contrast.

Frequently asked questions

What is speckle plethysmography (SPG)?

SPG is an optical way to follow blood flow with each heartbeat. A laser lights the skin, the scattered light forms a grainy speckle pattern, and moving red cells make that pattern blur. Measuring the blur many times a second gives a pulsatile flow waveform. A single laser through a finger yields both a PPG waveform from light intensity and an SPG waveform from speckle dynamics (Ghijsen et al., 2018).

Is SPG better than PPG?

It measures something different: flow rather than volume. In a skin simulation the pulsatile SPG signal was 24% and 40% at 532 and 860 nm versus 5.6% and 3.0% for PPG (Dunn et al., 2018), and SPG kept a higher signal-to-noise ratio than PPG during a cold pressor test (Herranz Olazábal et al., 2023). These are small research studies, so they show promise, not a settled advantage for every use.

Does SPG work on darker skin?

One study tested this directly. In skin-simulating phantoms SPG stayed linear with flow when absorbing skin layers were added (R² 0.996 to 0.999), and a small clinical protocol found no significant difference between lower- and higher-melanin groups (Rice et al., 2020). That is encouraging but limited evidence.

What wavelength does SPG use?

Studies have used visible and near-infrared lasers, for example 532 and 860 nm (Dunn et al., 2018) and 639 and 850 nm (Herranz Olazábal et al., 2022). Green and infrared light probe different skin depths (Maas et al., 2026). The laser's spectral width matters as much as its colour: broad laser diodes wash out speckle, and a linewidth under about 1 nm is recommended (Postnov et al., 2019).

Can SPG measure blood pressure?

Not as a cleared product. Research groups combine flow features from speckle with volume features from PPG-like signals to estimate blood pressure. Flow plus volume features correlated best with blood pressure in 13 people (Garrett et al., 2023), and subject-specific models improved systolic estimates over PPG alone (Garrett et al., 2025). A 133-patient preprint reported errors within ISO 81060-2 criteria but has not been peer reviewed (Choo et al., 2026).

Are there FDA-cleared SPG devices?

FlowMet-R from Laser Associated Sciences received FDA 510(k) clearance (K182494) on 27 February 2019 as a blood flowmeter. FlowMet is a toe-worn laser speckle device studied in patients with suspected peripheral artery disease (Razavi et al., 2021). The ContinUse Gili Pro BioSensor has an FDA De Novo authorization (DEN200038), but it uses speckle vibrometry, which reads surface motion rather than blood-flow blur.

How is SPG different from laser Doppler flowmetry?

Both read the same speckle fluctuations caused by moving cells. Laser Doppler resolves the frequency spectrum of those fluctuations, usually from a single-point fibre probe, while speckle contrast methods measure how much the pattern blurs during one camera exposure, which allows full-field imaging without scanning (Boas & Dunn, 2010).

What is SCOS?

Speckle contrast optical spectroscopy (SCOS) applies speckle contrast to light that has travelled deep through tissue, using a photon-diffusion model (Valdés et al., 2014). Fiber and high-density SCOS systems with ordinary CMOS cameras have measured changes in human brain blood flow (Kim et al., 2023; Kim et al., 2025).

Are there open SPG or speckle datasets?

Two open Zenodo records are a good start: laser speckle contrast recordings of human skin during physiological tests (doi:10.5281/zenodo.4359074) and the raw frames, processed data and code behind a 50-channel SCOS brain-flow study (doi:10.5281/zenodo.15881287). Both are CC BY 4.0.

References

  1. Aleman et al. (2026). Comparing multi- and single-exposure speckle contrast optical spectroscopy methods as estimators of blood flow in the diffuse regime. Journal of Biomedical Optics 31(02):1, 2026. doi:10.1117/1.JBO.31.2.025002
  2. Bi et al. (2013). Deep tissue flowmetry based on diffuse speckle contrast analysis. Optics Letters 38(9):1401, 2013. doi:10.1364/OL.38.001401
  3. Biswas et al. (2024). Non-invasive low-cost deep tissue blood flow measurement with integrated Diffuse Speckle Contrast Spectroscopy. Frontiers in Neuroergonomics 4:1288922, 2024. doi:10.3389/fnrgo.2023.1288922
  4. Boas & Dunn (2010). Laser speckle contrast imaging in biomedical optics. Journal of Biomedical Optics 15(01):1, 2010. doi:10.1117/1.3285504
  5. Bonetta-Misteli et al. (2023). Development and evaluation of a wearable peripheral vascular compensation sensor in a swine model of hemorrhage. Biomedical Optics Express 14(10):5338–5357, 2023. doi:10.1364/BOE.494720
  6. Cheng et al. (2024). Choosing a camera and optimizing system parameters for speckle contrast optical spectroscopy. Scientific Reports 14(1):11915, 2024. doi:10.1038/s41598-024-62106-y
  7. Choo et al. (2026). Clinical-grade Cuffless Blood Pressure Monitoring via Deep-tissue Diffuse Speckle Pulsatile Flowmetry. medRxiv, 2026. doi:10.64898/2026.06.19.26356074
  8. Dunn et al. (2018). Comparison of speckleplethysmographic (SPG) and photoplethysmographic (PPG) imaging by Monte Carlo simulations and in vivo measurements. Biomedical Optics Express 9(9):4306, 2018. doi:10.1364/BOE.9.004306
  9. Dunn et al. (2019). Speckleplethysmographic (SPG) Estimation of Heart Rate Variability During an Orthostatic Challenge. Scientific Reports 9(1):14079, 2019. doi:10.1038/s41598-019-50526-0
  10. Ellington et al. (2024). Speckle Plethysmograph-Based Blood Pressure Assessment. Technologies 12(5):70, 2024. doi:10.3390/technologies12050070
  11. Enomoto et al. (2021). Characterization of laser speckle flowgraphy pulse waveform parameters for the evaluation of the optic nerve head and retinal circulation. Scientific Reports 11:6847, 2021. doi:10.1038/s41598-021-86280-5
  12. Favilla et al. (2024). Portable cerebral blood flow monitor to detect large vessel occlusion in patients with suspected stroke. Journal of NeuroInterventional Surgery 17(4):388, 2024. doi:10.1136/jnis-2024-021536
  13. Garrett et al. (2023). Simultaneous photoplethysmography and blood flow measurements towards the estimation of blood pressure using speckle contrast optical spectroscopy. Biomedical Optics Express 14(4):1594, 2023. doi:10.1364/BOE.482740
  14. Garrett et al. (2025). Speckle contrast optical spectroscopy for cuffless blood pressure estimation based on microvascular blood flow and volume oscillations. Biomedical Optics Express 16(8):3004, 2025. doi:10.1364/BOE.560022
  15. Garrett et al. (2025). Optical methods for cuffless blood pressure measurements. Biophotonics Discovery 3(01):1, 2025. doi:10.1117/1.BIOS.3.1.010901
  16. Ghijsen et al. (2018). Wearable speckle plethysmography (SPG) for characterizing microvascular flow and resistance. Biomedical Optics Express 9(8):3937, 2018. doi:10.1364/BOE.9.003937
  17. Golberg et al. (2017). Large-scale clinical validation of noncontact and continuous extraction of blood pressure via multipoint defocused photonic imaging. Applied Optics 57(7):B45, 2017. doi:10.1364/ao.57.000b45
  18. Herranz Olazábal et al. (2021). Beat-to-beat Intervals of Speckle & Intensity-based Optical Plethysmograms compared to Electrocardiogram. Computing in cardiology, 2021. doi:10.23919/cinc53138.2021.9662819
  19. Herranz Olazábal et al. (2022). Camera-Derived Photoplethysmography (rPPG) and Speckle Plethysmography (rSPG): Comparing Reflective and Transmissive Mode at Various Integration Times Using LEDs and Lasers. Sensors 22(16):6059, 2022. doi:10.3390/s22166059
  20. Herranz Olazábal et al. (2023). Comparison between Speckle Plethysmography and Photoplethysmography during Cold Pressor Test Referenced to Finger Arterial Pressure. Sensors 23(11):5016, 2023. doi:10.3390/s23115016
  21. Herranz Olazábal et al. (2023). Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time. Bioengineering 10(1):101, 2023. doi:10.3390/bioengineering10010101
  22. Huang et al. (2024). Correlating stroke risk with non-invasive cerebrovascular perfusion dynamics using a portable speckle contrast optical spectroscopy laser device. Biomedical Optics Express 15(10):6083, 2024. doi:10.1364/boe.534796
  23. Kim et al. (2023). Measuring human cerebral blood flow and brain function with fiber-based speckle contrast optical spectroscopy system. Communications Biology 6(1):844, 2023. doi:10.1038/s42003-023-05211-4
  24. Kim et al. (2025). Mapping human cerebral blood flow with high-density, multi-channel speckle contrast optical spectroscopy. Communications Biology 8(1):1553, 2025. doi:10.1038/s42003-025-08915-x
  25. Lee et al. (2024). Rolling shutter speckle plethysmography for quantitative cardiovascular monitoring. Biomedical Optics Express 15(3):1540, 2024. doi:10.1364/BOE.511755
  26. Lorato et al. (2024). Towards continuous blood pressure monitoring using speckle plethysmography. Biophotonics in Exercise Science, Sports Medicine, Health Monitoring Technologies, and Wearables, 2024. doi:10.1117/12.3001627
  27. Maas et al. (2026). SPECKLE-BASED MEASUREMENT OF MICROVASCULAR FUNCTION. Journal of Hypertension 44(Suppl 1):e179, 2026. doi:10.1097/01.hjh.0001197056.64038.c5
  28. Postnov et al. (2019). Choosing a laser for laser speckle contrast imaging. Scientific Reports 9(1):2542, 2019. doi:10.1038/s41598-019-39137-x
  29. Que et al. (2024). Speckle Vibrometry for Contactless Instantaneous Heart Rate and Respiration Rate Monitoring on Mechanically Ventilated Patients. Sensors 24(19):6374, 2024. doi:10.3390/s24196374
  30. Razavi et al. (2021). A Real-Time Blood Flow Measurement Device for Patients with Peripheral Artery Disease. Journal of Vascular and Interventional Radiology 32(3):453–458, 2021. doi:10.1016/j.jvir.2020.09.006
  31. Rice et al. (2020). Effect of skin optical absorption on speckleplethysmographic (SPG) signals. Biomedical Optics Express 11(9):5352, 2020. doi:10.1364/BOE.403501
  32. Schnitzler et al. (2024). Wearable, compact photo- and speckle plethysmography device for continuous non-invasive vital signs monitoring. Optical Diagnostics and Sensing: Toward Point-of-Care Diagnostics 48:21, 2024. doi:10.1117/12.3001630
  33. Valdés et al. (2014). Speckle contrast optical spectroscopy, a non-invasive, diffuse optical method for measuring microvascular blood flow in tissue. Biomedical Optics Express 5(8):2769, 2014. doi:10.1364/BOE.5.002769
  34. Laser speckle-contrast images of human skin under physiological tests (Dremin et al.). Zenodo, CC BY 4.0. doi:10.5281/zenodo.4359074
  35. High-density multi-channel SCOS cerebral blood flow data and code (BUNPC 2025-SCOS). Zenodo, CC BY 4.0. doi:10.5281/zenodo.15881287

Educational content, not medical advice. Updated 2026-10-29.