Speckle plethysmography (SPG): blood flow from laser speckle
SPG shines a laser into skin and measures how fast the speckle pattern blurs. Moving red cells cause the blur, so SPG follows blood flow, while PPG follows blood volume through the light it absorbs. It is a research technique studied for pulse, perfusion and cuffless blood pressure.

SPG in one minute
Speckle plethysmography (SPG) records the pulsing of blood flow, beat by beat, from the way laser light flickers after it scatters through skin. That is the key difference from ordinary photoplethysmography (PPG): PPG tracks how much blood is under the sensor (volume) from how much light it absorbs, usually with an LED, while SPG tracks how fast the blood is moving (flow) from how a coherent laser's speckle pattern blurs. The SPG vs PPG section below compares them side by side.
Shine a laser into tissue and the light that comes back forms a grainy pattern called speckle. Red cells moving through small vessels make that pattern change, so a camera exposure sees it blurred. More flow means more blur. Track the blur many times a second and you get a flow waveform that rises and falls with every heartbeat.
The wearable form of SPG was introduced in 2018. Shining one laser through a finger produced two co-registered waveforms at once: the familiar PPG from the light's intensity, which follows blood volume, and an SPG from the speckle dynamics, which follows blood flow. In 16 people the two waveforms were offset in time and had different harmonic content (Ghijsen et al., 2018). SPG borrows its physics from laser speckle contrast imaging, a lab method for mapping blood flow in brain, skin and eye research (Boas & Dunn, 2010).
What laser speckle is
A laser emits light of one wavelength with its waves in step, which is what physicists call coherent light. When that light enters skin it scatters off cell membranes, collagen and red cells many times. Every photon takes its own route, and every route has a different length. When the returning light lands on a detector, waves that travelled different distances arrive out of step and add up or cancel each other. The result is a random map of bright and dark grains: speckle.
This only works when the light stays coherent over the whole spread of path lengths through the tissue (Boas & Dunn, 2010). That is why the choice of laser matters: many common semiconductor laser diodes have a spectrum broad enough to wash out speckle contrast, and a linewidth under about 1 nm is recommended (Postnov et al., 2019). LEDs are not coherent, and in a direct comparison LED-based speckle signals had much lower quality than laser-based ones (Herranz Olazábal et al., 2022).

Why moving blood blurs speckle
If nothing in the tissue moves, the speckle pattern holds still and a camera captures sharp grains however long it exposes. When red cells move, the path lengths change and the pattern shimmers. A camera exposure then averages several different patterns together, and the grains smear into a flatter grey. The faster the cells move, the faster the pattern changes and the stronger the blur in an exposure of a given length. Speckle contrast imaging typically uses exposures of about 1 to 10 ms (Boas & Dunn, 2010).
Speckle contrast,
Blur is measured with one number, the speckle contrast . Take a small patch of pixels, compute the spread of their brightness (the standard deviation ) and divide by their average brightness :
Sharp, fully developed speckle has close to 1: the grains are as varied as their average. Completely blurred speckle has near 0: every pixel looks the same (Boas & Dunn, 2010). How falls with exposure time depends on the speckle decorrelation time , the time it takes the pattern to change into a new one. A widely used model gives
where is a constant set by the optics (how big a speckle grain is compared with a pixel, and polarization). In this model is inversely proportional to the speed of the moving cells, so it is the link between blur and flow (Boas & Dunn, 2010). For exposures much longer than the formula simplifies to
so rises roughly in step with flow speed. Speckle contrast gives a relative flow index rather than an absolute flow in millilitres per minute, and turning it into quantitative flow takes extra modelling (Boas & Dunn, 2010). Try it below: the exposure slider integrates a simulated, changing speckle pattern for the chosen time, and the readout compares the simulated frame's contrast with the model above.
Static tissueK 1.00
Flowing · 5.0 msK 0.25Static, K = 1Flowing, Eq. 9Simulated frame
Static frameFlowing frameexp(−x), fully developed
SPG vs PPG
PPG measures blood volume; SPG measures blood flow. PPG shines light into the skin and reads how much of it the blood absorbs: with each beat the vessels swell, more light is absorbed, and the detected intensity dips. SPG shines a coherent laser and reads how fast the speckle changes, which depends on how fast the red cells move. The same laser and camera can record both at once (Ghijsen et al., 2018; Schnitzler et al., 2024).

| Aspect | PPG | SPG | Source |
|---|---|---|---|
| What it follows | Blood volume, from light intensity | Blood flow, from speckle dynamics | (Ghijsen et al., 2018) |
| Light source | Any light; only intensity is used (a laser works too) | Coherent laser with a narrow linewidth (under about 1 nm recommended) | (Ghijsen et al., 2018; Postnov et al., 2019) |
| Pulsatile amplitude in a skin simulation | 5.6% at 532 nm, 3.0% at 860 nm | 24% at 532 nm, 40% at 860 nm | (Dunn et al., 2018) |
| Longer near-infrared wavelength | Pulsatile modulation was smaller | Pulsatile modulation grew | (Dunn et al., 2018) |
| Camera signal quality, reflection mode | rPPG SQI below 0.845 | Laser rSPG SQI above 0.965 | (Herranz Olazábal et al., 2022) |
| Transmission through a finger | — | Speckle washed out and SPG began to look like PPG | (Herranz Olazábal et al., 2022) |
| Low flow (cold pressor test) | Weaker signal, lower signal-to-noise ratio than SPG | Higher, more stable signal-to-noise ratio | (Herranz Olazábal et al., 2023) |
| Pulse shape vs finger arterial pressure | Deviation 0.17 | Deviation 0.09 | (Herranz Olazábal et al., 2023) |
The two signals are complementary rather than rivals. Because one tracks vessel expansion and the other tracks flow, their waveforms are offset in time (Ghijsen et al., 2018), and a single rolling-shutter camera frame can separate blood velocity from blood volume changes (Lee et al., 2024). For the PPG side of the story, see the guides to the perfusion index and the AC/DC ratio.
What SPG can measure
Everything below comes from research studies, most of them small. None of it makes SPG a diagnostic test.
- Pulse and beat-to-beat intervals. A camera filming a laser-lit finger gave beat intervals closer to ECG than a finger-clip PPG across 3,234 beats in nine people (mean absolute deviation 10.4 vs 14.2 ms) (Herranz Olazábal et al., 2021). SPG-derived heart rate variability tracked ECG during a posture challenge in 17 people (Dunn et al., 2019).
- Microvascular flow. SPG waveform timing and shape relate to microvascular flow and resistance (Ghijsen et al., 2018). Green and infrared lasers probe different depths, and their ratio has been proposed as a micro-to-macrovascular index (Maas et al., 2026).
- Blood-pressure features (research only). Flow and pulse-arrival features from SPG have been studied for cuffless blood pressure (Lorato et al., 2024; Ellington et al., 2024; Garrett et al., 2023; Garrett et al., 2025). A review found that evaluation methods vary too much to compare optical BP devices yet (Garrett et al., 2025). For how cuffless devices are tested, see the ESH cuffless validation tests explained.
- Peripheral flow. A toe-worn laser speckle device was studied in suspected peripheral artery disease (Razavi et al., 2021), and a wearable speckle flow index tracked blood loss in a swine hemorrhage model (Bonetta-Misteli et al., 2023).
- Brain blood flow. Diffuse speckle methods read flow deep below the surface with cameras (Bi et al., 2013; Valdés et al., 2014), including changes in human cerebral blood flow through the scalp and skull (Kim et al., 2023; Kim et al., 2025; Huang et al., 2024; Favilla et al., 2024).
- The eye. Laser speckle flowgraphy reads blood flow in the optic nerve head, and its pulse-waveform shape relates to ocular circulation (Enomoto et al., 2021).
Inside an SPG sensor
An SPG sensor needs three things. First, a coherent light source with a narrow spectrum (Postnov et al., 2019). Second, a detector that can see speckle statistics: usually a camera, whose linearity, read noise and quantization decide how accurately contrast is measured in low light (Cheng et al., 2024), although an ordinary photodiode with an integrating circuit can also work (Biswas et al., 2024). Third, processing that corrects contrast for camera shot noise and dark noise (Valdés et al., 2014).
Geometry matters too. In reflection, with source and detector side by side on the skin, laser SPG kept high signal quality; in transmission through a finger the speckle washed out (Herranz Olazábal et al., 2022). A compact research wearable already pairs a laser diode with a small camera to record PPG and SPG together (Schnitzler et al., 2024).

Limits and open questions
- A single exposure can mislead. In simulations, estimating flow from one exposure time could be off by up to 90% and misjudge pulsatile flow by 22% or more, while fitting many exposures kept errors under 1% (Aleman et al., 2026).
- Relative, not absolute. Speckle contrast gives a relative flow index; absolute flow needs more modelling (Boas & Dunn, 2010).
- Small, mostly resting studies. Many of the human studies here enrolled 9 to 17 people in lab protocols (Herranz Olazábal et al., 2021; Dunn et al., 2019; Lorato et al., 2024). One of the few to include walking and running used a smartphone prototype in eight people (Ellington et al., 2024).
- Unreviewed and unvalidated results. The largest blood-pressure study is a preprint (Choo et al., 2026), and the stroke-detection model still needs an independent test set (Favilla et al., 2024).
- Skin tone. The direct evidence is one phantom study with a small clinical protocol (Rice et al., 2020). For the PPG side, see how melanin affects PPG accuracy.
Devices and companies
A short, sourced list. Regulatory facts link to the FDA database or the maker's own page, since they have no DOI. Listing is not an endorsement.
| Device | What it does | Status and source |
|---|---|---|
| FlowMet-R (Laser Associated Sciences) | Laser speckle blood-flow waveform at the toe (Razavi et al., 2021) | FDA 510(k) cleared 27 Feb 2019, K182494 |
| Openwater Open-Motion | Pulsed speckle contrast headset for brain blood flow, with an open SDK; an Openwater headset was used in a stroke study (Favilla et al., 2024) | For research use only (product page, SDK) |
| Hemorai CSI patch | Wearable patch combining speckle flow with tissue spectroscopy, aimed at hemorrhage | In development, not FDA cleared (company page) |
| Armor Medical aRMOR | Wearable laser speckle flow sensor for hemorrhage; published data so far are from swine (Bonetta-Misteli et al., 2023) | Prototype (company page) |
| ContinUse Gili Pro BioSensor | Contactless spot-check heart and breathing rate from chest-wall vibration. This is speckle vibrometry, not SPG | FDA De Novo DEN200038 |
Open datasets
- Laser speckle-contrast images of human skin under physiological tests (Dremin et al.). Skin laser speckle contrast recordings during controlled respiration, breath holding, venous occlusion and spontaneous respiration. Four MATLAB files, about 0.9 GB. The record lists the tests and files but no participant metadata. CC BY 4.0. doi:10.5281/zenodo.4359074 · ChatPPG dataset card
- High-density multi-channel SCOS cerebral blood flow data and code (BUNPC 2025-SCOS). Raw camera frames, processed data and analysis code behind the 50-channel SCOS brain-flow study: 15 people doing a Stroop task, plus a liquid flow phantom. About 3.9 GB. CC BY 4.0. doi:10.5281/zenodo.15881287 · paper: (Kim et al., 2025) · ChatPPG dataset card
For PPG data, see PPG datasets for blood pressure and the research atlas.
The evidence, paper by paper
All 33 summarized papers in the ChatPPG speckle plethysmography collection, grouped by question. Each links to its DOI.
Foundations and physics
Where SPG and its speckle-contrast relatives come from.
Wearable speckle plethysmography (SPG) for characterizing microvascular flow and resistance
Ghijsen et al. · Biomedical Optics Express 9(8):3937, 2018 · doi:10.1364/BOE.9.003937 · ChatPPG card
This paper introduced wearable speckle plethysmography (SPG). A single laser shining through a finger yields two co-registered waveforms: the classic PPG from light intensity (blood volume) and a new SPG from laser speckle dynamics (blood flow). In 16 people the two waveforms were offset in time and had different harmonic content, which the authors relate to microvascular flow and resistance.
Laser speckle contrast imaging in biomedical optics
Boas & Dunn · Journal of Biomedical Optics 15(01):1, 2010 · doi:10.1117/1.3285504 · ChatPPG card
The standard review of laser speckle contrast imaging: how blurring of a laser speckle pattern during a camera exposure encodes blood flow, how to make the measurement more quantitative, and how it has been used in brain, skin and eye research.
Valdés et al. · Biomedical Optics Express 5(8):2769, 2014 · doi:10.1364/BOE.5.002769 · ChatPPG card
This paper introduced speckle contrast optical spectroscopy (SCOS), which measures deep-tissue blood flow by combining laser speckle contrast statistics with a photon-diffusion model. It was validated on phantoms and a human forearm, and it introduced corrections for camera shot noise and dark noise that later speckle methods depend on.
Deep tissue flowmetry based on diffuse speckle contrast analysis
Bi et al. · Optics Letters 38(9):1401, 2013 · doi:10.1364/OL.38.001401 · ChatPPG card
Diffuse speckle contrast analysis (DSCA) showed that deep-tissue blood-flow information similar to diffuse correlation spectroscopy can be obtained from a simple camera image of the speckle pattern instead of expensive single-photon counters. It was demonstrated on a flow phantom and during cuff occlusion in vivo.
Dunn et al. · Biomedical Optics Express 9(9):4306, 2018 · doi:10.1364/BOE.9.004306 · ChatPPG card
Monte Carlo simulations in an eight-layer skin model plus in vivo camera measurements compared speckle plethysmography (SPG) with PPG. Simulated pulsatile amplitude was 24 % (532 nm) and 40 % (860 nm) for SPG versus 5.6 % and 3.0 % for PPG, and both simulation and measurement point to a much larger signal-to-noise ratio for SPG. Unlike PPG, the SPG modulation grows at the longer near-infrared wavelength.
Light sources, cameras and processing
What the optics and the math need to get right.
Herranz Olazábal et al. · Sensors 22(16):6059, 2022 · doi:10.3390/s22166059 · ChatPPG card
This study compared camera-derived PPG and speckle plethysmography (SPG) on ten fingers using lasers and LEDs at 639 and 850 nm, in reflection and transmission. In reflection, laser-based SPG had far better signal quality (SQI above 0.965) than LED-based signals, which shows the advantage really comes from coherent light. In transmission the speckle washed out and SPG started to look like PPG.
Choosing a laser for laser speckle contrast imaging
Postnov et al. · Scientific Reports 9(1):2542, 2019 · doi:10.1038/s41598-019-39137-x · ChatPPG card
Not all lasers are equal for speckle imaging. This study showed that many common semiconductor laser diodes have a spectrum broad enough to noticeably wash out speckle contrast and reduce signal-to-noise, and it recommends grating-stabilised diodes or other sources with linewidth under about 1 nm.
Choosing a camera and optimizing system parameters for speckle contrast optical spectroscopy
Cheng et al. · Scientific Reports 14(1):11915, 2024 · doi:10.1038/s41598-024-62106-y · ChatPPG card
A practical guide to choosing and tuning a camera for speckle contrast optical spectroscopy. Testing three commercial CMOS cameras showed that linearity, read noise and quantization distortion determine how accurately speckle contrast is measured at low light, and that some cameras that are fine for ordinary imaging are poor for speckle work.
Biswas et al. · Frontiers in Neuroergonomics 4:1288922, 2024 · doi:10.3389/fnrgo.2023.1288922 · ChatPPG card
Integrated diffuse speckle contrast spectroscopy (iDSCS) measures deep-tissue blood flow with an ordinary photodiode and a custom integrating circuit rather than an expensive camera or single-photon detector. Tested on a human forearm against diffuse correlation spectroscopy, it was sensitive to deep blood flow while being low-cost, low-power and small enough for wireless wearable probes.
Schnitzler et al. · Optical Diagnostics and Sensing: Toward Point-of-Care Diagnostics 48:21, 2024 · doi:10.1117/12.3001630 · ChatPPG card
Researchers at imec built a compact wearable that pairs a laser diode with a small camera to record PPG and speckle plethysmography (SPG) at the same time, with real-time display and software-tunable settings. It is presented as a first step toward a fully integrated, low-cost SPG/PPG wearable for continuous vital-sign monitoring.
Rolling shutter speckle plethysmography for quantitative cardiovascular monitoring
Lee et al. · Biomedical Optics Express 15(3):1540, 2024 · doi:10.1364/BOE.511755 · ChatPPG card
Rolling-shutter speckle plethysmography uses the tiny time delay between rows of an ordinary rolling-shutter camera sensor to measure how quickly laser speckle decorrelates, all from a single frame. That lets one camera separate changes in blood velocity from changes in blood volume across the cardiac cycle.
Aleman et al. · Journal of Biomedical Optics 31(02):1, 2026 · doi:10.1117/1.JBO.31.2.025002 · ChatPPG card
A simulation study compared estimating blood flow from speckle contrast at a single camera exposure (as standard SCOS and SPG do) with fitting a curve over many exposures (multi-exposure speckle imaging). Multi-exposure kept errors under 1 % even when calibration and noise were off, while single-exposure methods could be off by up to 90 % and misjudge pulsatile flow by 22 % or more.
Pulse, HRV, low perfusion and skin tone
Head-to-head comparisons with PPG and ECG.
Speckleplethysmographic (SPG) Estimation of Heart Rate Variability During an Orthostatic Challenge
Dunn et al. · Scientific Reports 9(1):14079, 2019 · doi:10.1038/s41598-019-50526-0 · ChatPPG card
Seventeen people did an orthostatic (posture) challenge while heart rate variability was measured with ECG, PPG and speckle plethysmography (SPG). SPG-derived HRV correlated highly with ECG in both time and frequency domains and was more accurate than PPG-derived HRV, which tends to degrade under physical stress.
Herranz Olazábal et al. · Computing in cardiology, 2021 · doi:10.23919/cinc53138.2021.9662819 · ChatPPG card
A camera filming a laser-lit finger (639 nm, 100 fps) produced a remote speckle plethysmogram (rSPG) whose beat-to-beat intervals were compared with ECG across 3,234 heartbeats in nine people. The contactless rSPG had smaller interval errors than a contact finger-clip PPG (mean absolute deviation 10.4 vs 14.2 ms).
Herranz Olazábal et al. · Sensors 23(11):5016, 2023 · doi:10.3390/s23115016 · ChatPPG card
Ten people did a cold pressor test (hand in ice water) to cut finger blood flow while speckle plethysmography (SPG) and PPG were recorded from the same video at 639 and 850 nm. Both signals weakened, but SPG kept a significantly higher and more stable signal-to-noise ratio than PPG throughout and showed richer harmonics.
Effect of skin optical absorption on speckleplethysmographic (SPG) signals
Rice et al. · Biomedical Optics Express 11(9):5352, 2020 · doi:10.1364/BOE.403501 · ChatPPG card
Skin-simulating phantoms and a small clinical protocol tested whether skin pigmentation affects speckle plethysmography (SPG). SPG tracked known flow linearly (R² 0.990) and stayed linear with absorbing skin layers added (R² 0.996–0.999), and clinical readings showed no significant difference between lower- and higher-melanin groups.
SPECKLE-BASED MEASUREMENT OF MICROVASCULAR FUNCTION
Maas et al. · Journal of Hypertension 44(Suppl 1):e179, 2026 · doi:10.1097/01.hjh.0001197056.64038.c5 · ChatPPG card
An imec conference abstract describes a finger speckle device using green and infrared lasers that probe different skin depths. Pressing on the finger made the shallow (green) speckle pulse shrink or vanish while the deeper infrared signal changed less, and a ratio of the two, a micro-to-macrovascular index, dropped by a median 55 %. The authors propose it as a step toward non-invasive measurement of microvascular function.
Cuffless blood pressure (research)
Flow plus volume features for blood-pressure estimation. None of these are cleared devices.
Herranz Olazábal et al. · Bioengineering 10(1):101, 2023 · doi:10.3390/bioengineering10010101 · ChatPPG card
Remote speckle plethysmography (SPG) from a camera and a 639 nm laser was compared with a clinical finger-clip PPG, using finger arterial pressure as the reference. Across 2,599 beats the SPG pulse shape matched the arterial pressure waveform about twice as closely as PPG (deviation 0.09 vs 0.17). Pulse arrival times were similar for both.
Towards continuous blood pressure monitoring using speckle plethysmography
Lorato et al. · Biophotonics in Exercise Science, Sports Medicine, Health Monitoring Technologies, and Wearables, 2024 · doi:10.1117/12.3001627 · ChatPPG card
Ten people wore a contact speckle plethysmography (SPG) sensor alongside ECG, PPG and a reference blood pressure monitor. Pulse arrival time and waveform features from SPG correlated strongly with blood pressure, and the relationship varied less between people than it did for PPG, which suggests SPG might support a more general, less calibration-heavy BP model.
Speckle Plethysmograph-Based Blood Pressure Assessment
Ellington et al. · Technologies 12(5):70, 2024 · doi:10.3390/technologies12050070 · ChatPPG card
A prototype smartphone-based speckle plethysmography (SPG) system was paired with ECG to estimate systolic blood pressure from pulse arrival time in eight people while sitting, walking and running. SPG-based estimates were slightly better than PPG-based ones (resting error 12.4 vs 13.7 mmHg), and adding the previous BP reading as an input cut the error to 3.7 mmHg.
Garrett et al. · Biomedical Optics Express 14(4):1594, 2023 · doi:10.1364/BOE.482740 · ChatPPG card
Speckle contrast optical spectroscopy (SCOS) records both a PPG-like blood-volume signal and a blood-flow index from the same light. On the finger and wrist of 13 people, blood-flow features correlated with blood pressure slightly better than PPG features, and features combining flow and volume correlated best (R = −0.59).
Garrett et al. · Biomedical Optics Express 16(8):3004, 2025 · doi:10.1364/BOE.560022 · ChatPPG card
High-speed (390 Hz) speckle contrast optical spectroscopy recorded blood flow and blood volume waveforms at two wavelengths on the wrist and finger. Feeding flow plus volume features into subject-specific machine-learning models estimated systolic and diastolic blood pressure with errors of about 2.9 and 2.1 mmHg (SD) across a wide BP range, a 31 % improvement in systolic estimation over PPG alone, and the gain held when people were re-measured weeks later.
Choo et al. · medRxiv, 2026 · doi:10.64898/2026.06.19.26356074 · ChatPPG card
A preprint from A*STAR Singapore tests diffuse speckle pulsatile flowmetry, a laser-speckle method that records deep-tissue blood flow alongside a PPG-like volume signal, for cuffless blood pressure. In 133 outpatients it reported mean absolute errors of 4.17 mmHg (systolic) and 2.42 mmHg (diastolic) against a cuff, meeting ISO 81060-2 accuracy criteria and British Hypertension Society grade A. Ablations showed the flow signal improved on PPG alone. It has not yet been peer reviewed.
Optical methods for cuffless blood pressure measurements
Garrett et al. · Biophotonics Discovery 3(01):1, 2025 · doi:10.1117/1.BIOS.3.1.010901 · ChatPPG card
A review of optical approaches to cuffless blood pressure: finger and wrist PPG, multiwavelength PPG, imaging PPG, and speckle contrast optical spectroscopy / speckle plethysmography (SCOS/SPG). It finds that most work uses finger PPG, that evaluation methods vary too much to compare devices, and that newer techniques like SCOS/SPG add physiological information that may improve accuracy if tested against consensus standards.
Peripheral flow, hemorrhage and the eye
Clinical-research uses of speckle flow outside the brain.
A Real-Time Blood Flow Measurement Device for Patients with Peripheral Artery Disease
Razavi et al. · Journal of Vascular and Interventional Radiology 32(3):453–458, 2021 · doi:10.1016/j.jvir.2020.09.006 · ChatPPG card
FlowMet, a toe-worn optical device that uses laser speckle dynamic light scattering to read blood flow, was tested in 167 limbs of 90 patients with suspected peripheral artery disease. Its waveforms resembled Doppler ultrasound, and its flow-waveform score predicted critical limb ischemia (AUC 0.83) about as well as the toe-brachial index and better than the ankle-brachial index.
Bonetta-Misteli et al. · Biomedical Optics Express 14(10):5338–5357, 2023 · doi:10.1364/BOE.494720 · ChatPPG card
A low-cost wearable laser speckle sensor measured peripheral blood flow (laser speckle flow index) to catch the vasoconstriction the body uses to compensate for bleeding. In six pigs with controlled blood loss, the flow index tracked blood-volume loss with a correlation of −0.95, outperforming the standard shock index, and tracked recovery during resuscitation. The work targets early detection of postpartum hemorrhage.
Enomoto et al. · Scientific Reports 11:6847, 2021 · doi:10.1038/s41598-021-86280-5 · ChatPPG card
Laser speckle flowgraphy is a clinical eye-imaging method that reads blood flow in the optic nerve head from laser speckle blur. In 111 eyes of 86 healthy adults, the shape of its pulse waveform (skew, flow acceleration and related indices) related to systemic factors and retinal vessel measures, which supports waveform features as markers of ocular circulation.
Brain blood flow (SCOS)
Diffuse speckle contrast through the scalp and skull.
Kim et al. · Communications Biology 6(1):844, 2023 · doi:10.1038/s42003-023-05211-4 · ChatPPG card
A fiber-based speckle contrast optical spectroscopy system using ordinary CMOS cameras measured changes in human cerebral blood flow during brain activation at 33 mm source-detector separation. Pulsing the laser to raise photon flux, plus a careful processing pipeline, gave more than ten times the signal-to-noise ratio of diffuse correlation spectroscopy at a similar cost.
Kim et al. · Communications Biology 8(1):1553, 2025 · doi:10.1038/s42003-025-08915-x · ChatPPG card
A high-density speckle contrast optical spectroscopy system with 7 laser sources and 17 CMOS cameras (50 channels) mapped changes in human cerebral blood flow over a 7.6 × 6.6 cm patch of the head. In 15 people doing a Stroop task it produced activation maps with up to a 6.6 % flow change, and the raw data and processing code are openly released.
Huang et al. · Biomedical Optics Express 15(10):6083, 2024 · doi:10.1364/boe.534796 · ChatPPG card
A portable speckle contrast optical spectroscopy device (a laser diode plus a small board camera) measured changes in brain blood flow and volume while 50 people held their breath. People at higher stroke risk showed significantly different blood-flow responses than low-risk people, which suggests a cheap, non-invasive way to assess cerebrovascular reserve.
Favilla et al. · Journal of NeuroInterventional Surgery 17(4):388, 2024 · doi:10.1136/jnis-2024-021536 · ChatPPG card
A portable Openwater headset recorded 70-second cerebral blood-flow waveforms from laser speckle contrast in 135 suspected-stroke patients. A deep-learning model on those waveforms detected large-vessel occlusion with 79 % sensitivity and 84 % specificity (AUROC 0.82), beating two standard prehospital stroke scales; it still needs validation on an independent test set.
Adjacent: speckle vibrometry (not SPG)
These read surface motion from a defocused speckle pattern, not blood-flow decorrelation. Listed for contrast.
Que et al. · Sensors 24(19):6374, 2024 · doi:10.3390/s24196374 · ChatPPG card
Speckle vibrometry uses a laser and a defocused camera to detect tiny surface vibrations from the heartbeat and breathing, even through bedding. In mechanically ventilated ICU patients measured through two to three blanket layers, it tracked instantaneous heart rate within about ±2 bpm of ECG and breathing rate within about ±0.7 breaths per minute of the ventilator.
Golberg et al. · Applied Optics 57(7):B45, 2017 · doi:10.1364/ao.57.000b45 · ChatPPG card
A remote optical system tracked laser speckle patterns reflected from the skin at several points to measure pulse transit time without contact, then converted it to systolic and diastolic blood pressure with a model compared against a reference monitor.
Frequently asked questions
What is speckle plethysmography (SPG)?
SPG is an optical way to follow blood flow with each heartbeat. A laser lights the skin, the scattered light forms a grainy speckle pattern, and moving red cells make that pattern blur. Measuring the blur many times a second gives a pulsatile flow waveform. A single laser through a finger yields both a PPG waveform from light intensity and an SPG waveform from speckle dynamics (Ghijsen et al., 2018).
Is SPG better than PPG?
It measures something different: flow rather than volume. In a skin simulation the pulsatile SPG signal was 24% and 40% at 532 and 860 nm versus 5.6% and 3.0% for PPG (Dunn et al., 2018), and SPG kept a higher signal-to-noise ratio than PPG during a cold pressor test (Herranz Olazábal et al., 2023). These are small research studies, so they show promise, not a settled advantage for every use.
Does SPG work on darker skin?
One study tested this directly. In skin-simulating phantoms SPG stayed linear with flow when absorbing skin layers were added (R² 0.996 to 0.999), and a small clinical protocol found no significant difference between lower- and higher-melanin groups (Rice et al., 2020). That is encouraging but limited evidence.
What wavelength does SPG use?
Studies have used visible and near-infrared lasers, for example 532 and 860 nm (Dunn et al., 2018) and 639 and 850 nm (Herranz Olazábal et al., 2022). Green and infrared light probe different skin depths (Maas et al., 2026). The laser's spectral width matters as much as its colour: broad laser diodes wash out speckle, and a linewidth under about 1 nm is recommended (Postnov et al., 2019).
Can SPG measure blood pressure?
Not as a cleared product. Research groups combine flow features from speckle with volume features from PPG-like signals to estimate blood pressure. Flow plus volume features correlated best with blood pressure in 13 people (Garrett et al., 2023), and subject-specific models improved systolic estimates over PPG alone (Garrett et al., 2025). A 133-patient preprint reported errors within ISO 81060-2 criteria but has not been peer reviewed (Choo et al., 2026).
Are there FDA-cleared SPG devices?
FlowMet-R from Laser Associated Sciences received FDA 510(k) clearance (K182494) on 27 February 2019 as a blood flowmeter. FlowMet is a toe-worn laser speckle device studied in patients with suspected peripheral artery disease (Razavi et al., 2021). The ContinUse Gili Pro BioSensor has an FDA De Novo authorization (DEN200038), but it uses speckle vibrometry, which reads surface motion rather than blood-flow blur.
How is SPG different from laser Doppler flowmetry?
Both read the same speckle fluctuations caused by moving cells. Laser Doppler resolves the frequency spectrum of those fluctuations, usually from a single-point fibre probe, while speckle contrast methods measure how much the pattern blurs during one camera exposure, which allows full-field imaging without scanning (Boas & Dunn, 2010).
What is SCOS?
Speckle contrast optical spectroscopy (SCOS) applies speckle contrast to light that has travelled deep through tissue, using a photon-diffusion model (Valdés et al., 2014). Fiber and high-density SCOS systems with ordinary CMOS cameras have measured changes in human brain blood flow (Kim et al., 2023; Kim et al., 2025).
Are there open SPG or speckle datasets?
Two open Zenodo records are a good start: laser speckle contrast recordings of human skin during physiological tests (doi:10.5281/zenodo.4359074) and the raw frames, processed data and code behind a 50-channel SCOS brain-flow study (doi:10.5281/zenodo.15881287). Both are CC BY 4.0.
References
- Aleman et al. (2026). Comparing multi- and single-exposure speckle contrast optical spectroscopy methods as estimators of blood flow in the diffuse regime. Journal of Biomedical Optics 31(02):1, 2026. doi:10.1117/1.JBO.31.2.025002
- Bi et al. (2013). Deep tissue flowmetry based on diffuse speckle contrast analysis. Optics Letters 38(9):1401, 2013. doi:10.1364/OL.38.001401
- Biswas et al. (2024). Non-invasive low-cost deep tissue blood flow measurement with integrated Diffuse Speckle Contrast Spectroscopy. Frontiers in Neuroergonomics 4:1288922, 2024. doi:10.3389/fnrgo.2023.1288922
- Boas & Dunn (2010). Laser speckle contrast imaging in biomedical optics. Journal of Biomedical Optics 15(01):1, 2010. doi:10.1117/1.3285504
- Bonetta-Misteli et al. (2023). Development and evaluation of a wearable peripheral vascular compensation sensor in a swine model of hemorrhage. Biomedical Optics Express 14(10):5338–5357, 2023. doi:10.1364/BOE.494720
- Cheng et al. (2024). Choosing a camera and optimizing system parameters for speckle contrast optical spectroscopy. Scientific Reports 14(1):11915, 2024. doi:10.1038/s41598-024-62106-y
- Choo et al. (2026). Clinical-grade Cuffless Blood Pressure Monitoring via Deep-tissue Diffuse Speckle Pulsatile Flowmetry. medRxiv, 2026. doi:10.64898/2026.06.19.26356074
- Dunn et al. (2018). Comparison of speckleplethysmographic (SPG) and photoplethysmographic (PPG) imaging by Monte Carlo simulations and in vivo measurements. Biomedical Optics Express 9(9):4306, 2018. doi:10.1364/BOE.9.004306
- Dunn et al. (2019). Speckleplethysmographic (SPG) Estimation of Heart Rate Variability During an Orthostatic Challenge. Scientific Reports 9(1):14079, 2019. doi:10.1038/s41598-019-50526-0
- Ellington et al. (2024). Speckle Plethysmograph-Based Blood Pressure Assessment. Technologies 12(5):70, 2024. doi:10.3390/technologies12050070
- Enomoto et al. (2021). Characterization of laser speckle flowgraphy pulse waveform parameters for the evaluation of the optic nerve head and retinal circulation. Scientific Reports 11:6847, 2021. doi:10.1038/s41598-021-86280-5
- Favilla et al. (2024). Portable cerebral blood flow monitor to detect large vessel occlusion in patients with suspected stroke. Journal of NeuroInterventional Surgery 17(4):388, 2024. doi:10.1136/jnis-2024-021536
- Garrett et al. (2023). Simultaneous photoplethysmography and blood flow measurements towards the estimation of blood pressure using speckle contrast optical spectroscopy. Biomedical Optics Express 14(4):1594, 2023. doi:10.1364/BOE.482740
- Garrett et al. (2025). Speckle contrast optical spectroscopy for cuffless blood pressure estimation based on microvascular blood flow and volume oscillations. Biomedical Optics Express 16(8):3004, 2025. doi:10.1364/BOE.560022
- Garrett et al. (2025). Optical methods for cuffless blood pressure measurements. Biophotonics Discovery 3(01):1, 2025. doi:10.1117/1.BIOS.3.1.010901
- Ghijsen et al. (2018). Wearable speckle plethysmography (SPG) for characterizing microvascular flow and resistance. Biomedical Optics Express 9(8):3937, 2018. doi:10.1364/BOE.9.003937
- Golberg et al. (2017). Large-scale clinical validation of noncontact and continuous extraction of blood pressure via multipoint defocused photonic imaging. Applied Optics 57(7):B45, 2017. doi:10.1364/ao.57.000b45
- Herranz Olazábal et al. (2021). Beat-to-beat Intervals of Speckle & Intensity-based Optical Plethysmograms compared to Electrocardiogram. Computing in cardiology, 2021. doi:10.23919/cinc53138.2021.9662819
- Herranz Olazábal et al. (2022). Camera-Derived Photoplethysmography (rPPG) and Speckle Plethysmography (rSPG): Comparing Reflective and Transmissive Mode at Various Integration Times Using LEDs and Lasers. Sensors 22(16):6059, 2022. doi:10.3390/s22166059
- Herranz Olazábal et al. (2023). Comparison between Speckle Plethysmography and Photoplethysmography during Cold Pressor Test Referenced to Finger Arterial Pressure. Sensors 23(11):5016, 2023. doi:10.3390/s23115016
- Herranz Olazábal et al. (2023). Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time. Bioengineering 10(1):101, 2023. doi:10.3390/bioengineering10010101
- Huang et al. (2024). Correlating stroke risk with non-invasive cerebrovascular perfusion dynamics using a portable speckle contrast optical spectroscopy laser device. Biomedical Optics Express 15(10):6083, 2024. doi:10.1364/boe.534796
- Kim et al. (2023). Measuring human cerebral blood flow and brain function with fiber-based speckle contrast optical spectroscopy system. Communications Biology 6(1):844, 2023. doi:10.1038/s42003-023-05211-4
- Kim et al. (2025). Mapping human cerebral blood flow with high-density, multi-channel speckle contrast optical spectroscopy. Communications Biology 8(1):1553, 2025. doi:10.1038/s42003-025-08915-x
- Lee et al. (2024). Rolling shutter speckle plethysmography for quantitative cardiovascular monitoring. Biomedical Optics Express 15(3):1540, 2024. doi:10.1364/BOE.511755
- Lorato et al. (2024). Towards continuous blood pressure monitoring using speckle plethysmography. Biophotonics in Exercise Science, Sports Medicine, Health Monitoring Technologies, and Wearables, 2024. doi:10.1117/12.3001627
- Maas et al. (2026). SPECKLE-BASED MEASUREMENT OF MICROVASCULAR FUNCTION. Journal of Hypertension 44(Suppl 1):e179, 2026. doi:10.1097/01.hjh.0001197056.64038.c5
- Postnov et al. (2019). Choosing a laser for laser speckle contrast imaging. Scientific Reports 9(1):2542, 2019. doi:10.1038/s41598-019-39137-x
- Que et al. (2024). Speckle Vibrometry for Contactless Instantaneous Heart Rate and Respiration Rate Monitoring on Mechanically Ventilated Patients. Sensors 24(19):6374, 2024. doi:10.3390/s24196374
- Razavi et al. (2021). A Real-Time Blood Flow Measurement Device for Patients with Peripheral Artery Disease. Journal of Vascular and Interventional Radiology 32(3):453–458, 2021. doi:10.1016/j.jvir.2020.09.006
- Rice et al. (2020). Effect of skin optical absorption on speckleplethysmographic (SPG) signals. Biomedical Optics Express 11(9):5352, 2020. doi:10.1364/BOE.403501
- Schnitzler et al. (2024). Wearable, compact photo- and speckle plethysmography device for continuous non-invasive vital signs monitoring. Optical Diagnostics and Sensing: Toward Point-of-Care Diagnostics 48:21, 2024. doi:10.1117/12.3001630
- Valdés et al. (2014). Speckle contrast optical spectroscopy, a non-invasive, diffuse optical method for measuring microvascular blood flow in tissue. Biomedical Optics Express 5(8):2769, 2014. doi:10.1364/BOE.5.002769
- Laser speckle-contrast images of human skin under physiological tests (Dremin et al.). Zenodo, CC BY 4.0. doi:10.5281/zenodo.4359074
- High-density multi-channel SCOS cerebral blood flow data and code (BUNPC 2025-SCOS). Zenodo, CC BY 4.0. doi:10.5281/zenodo.15881287
Educational content, not medical advice. Updated 2026-10-29.