Paperdoi:10.1136/jnis-2024-021536
Portable cerebral blood flow monitor to detect large vessel occlusion in patients with suspected stroke
Journal of NeuroInterventional Surgery 17(4):388, 2024
TL;DR
A portable Openwater headset recorded 70-second cerebral blood-flow waveforms from laser speckle contrast in 135 suspected-stroke patients. A deep-learning model on those waveforms detected large-vessel occlusion with 79 % sensitivity and 84 % specificity (AUROC 0.82), beating two standard prehospital stroke scales; it still needs validation on an independent test set.
- Form factor: head
Abstract
Background
Early detection of large vessel occlusion (LVO) facilitates triage to an appropriate stroke center to reduce treatment times and improve outcomes. Prehospital stroke scales are not sufficiently sensitive, so we investigated the ability of the portable Openwater optical blood flow monitor to detect LVO.
Methods
Patients were prospectively enrolled at two comprehensive stroke centers during stroke alert evaluation within 24 hours of onset with National Institutes of Health Stroke Scale (NIHSS) score ≥2. A 70 s bedside optical blood flow scan generated cerebral blood flow waveforms based on relative changes in speckle contrast. Anterior circulation LVO was determined by CT angiography. A deep learning model trained on all patient data using fivefold cross-validation and learned discriminative representations from the raw speckle contrast waveform data. Receiver operating characteristic (ROC) analysis compared the Openwater diagnostic performance (ie, LVO detection) with prehospital stroke scales.
Results
Among 135 patients, 52 (39%) had an anterior circulation LVO. The median NIHSS score was 8 (IQR 4–14). The Openwater instrument had 79% sensitivity and 84% specificity for the detection of LVO. The rapid arterial occlusion evaluation (RACE) scale had 60% sensitivity and 81% specificity and the Los Angeles motor scale (LAMS) had 50% sensitivity and 81% specificity. The binary Openwater classification (high-likelihood vs low-likelihood) had an area under the ROC (AUROC) of 0.82 (95% CI 0.75 to 0.88), which outperformed RACE (AUC 0.70; 95% CI 0.62 to 0.78; P=0.04) and LAMS (AUC 0.65; 95% CI 0.57 to 0.73; P=0.002).
Conclusions
The Openwater optical blood flow monitor outperformed prehospital stroke scales for the detection of LVO in patients undergoing acute stroke evaluation in the emergency department. These encouraging findings need to be validated in an independent test set and the prehospital environment.
Verbatim from the publisher's DOI record (opens in a new tab) · Abstract as deposited by the publisher with Crossref; © the authors/publisher
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Signals
Keywords
How it works
- Wavelengths
- NIR laser (λ not in abstract)
Cite this paper
Favilla, C. G., Baird, G. L., Grama, K., Konecky, S. D., Carter, S., Smith, W., Gitlevich, R., Lebron-Cruz, A., Yodh, A. G., & McTaggart, R. (2024). Portable cerebral blood flow monitor to detect large vessel occlusion in patients with suspected stroke. Journal of NeuroInterventional Surgery , 2024, 17(4), 388. https://doi.org/10.1136/jnis-2024-021536
Favilla, Christopher G., Grayson Luderman Baird, Kedar Grama, Soren D. Konecky, Sarah Carter, Wendy Smith, Rebecca Gitlevich, Alexa Lebron-Cruz, Arjun G. Yodh, and Ryan McTaggart. 2024. "Portable cerebral blood flow monitor to detect large vessel occlusion in patients with suspected stroke." Journal of NeuroInterventional Surgery , 2024 17 (4): 388. https://doi.org/10.1136/jnis-2024-021536.
Favilla, C. G., Baird, G. L., Grama, K., Konecky, S. D., Carter, S., Smith, W., Gitlevich, R., Lebron-Cruz, A., Yodh, A. G. and McTaggart, R. (2024) 'Portable cerebral blood flow monitor to detect large vessel occlusion in patients with suspected stroke', Journal of NeuroInterventional Surgery , 2024, 17(4), p. 388. doi: 10.1136/jnis-2024-021536.
Favilla, Christopher G., et al. "Portable cerebral blood flow monitor to detect large vessel occlusion in patients with suspected stroke." Journal of NeuroInterventional Surgery , 2024, vol. 17, no. 4, 2024, p. 388. https://doi.org/10.1136/jnis-2024-021536.
C. G. Favilla et al., "Portable cerebral blood flow monitor to detect large vessel occlusion in patients with suspected stroke," Journal of NeuroInterventional Surgery , 2024, vol. 17, no. 4, p. 388, 2024, doi: 10.1136/jnis-2024-021536.
@article{favilla2024portable,
title = {{Portable cerebral blood flow monitor to detect large vessel occlusion in patients with suspected stroke}},
author = {Favilla, Christopher G. and Baird, Grayson Luderman and Grama, Kedar and Konecky, Soren D. and Carter, Sarah and Smith, Wendy and Gitlevich, Rebecca and Lebron-Cruz, Alexa and Yodh, Arjun G. and McTaggart, Ryan},
year = {2024},
journal = {Journal of NeuroInterventional Surgery , 2024},
volume = {17},
number = {4},
pages = {388},
doi = {10.1136/jnis-2024-021536}