Effect of skin optical absorption on speckleplethysmographic (SPG) signals
Biomedical Optics Express 11(9):5352, 2020
TL;DR
Skin-simulating phantoms and a small clinical protocol tested whether skin pigmentation affects speckle plethysmography (SPG). SPG tracked known flow linearly (R² 0.990) and stayed linear with absorbing skin layers added (R² 0.996–0.999), and clinical readings showed no significant difference between lower- and higher-melanin groups. Because SPG reads speckle decorrelation rather than absolute absorption, melanin mainly lowers signal level rather than biasing flow.
- Form factor: finger
Abstract
Recent advances in optical technology have emerged for measuring blood flow in the extremities using speckleplethysmography (SPG), which may address needs in vascular medicine and other fields. SPG has demonstrated a highly linear response with flow rate, but the susceptibility to differences in skin tone is unclear. Two validation studies using skin-simulating phantoms and a simple clinical protocol were conducted to determine the impact of absorbing skin layers on SPG measurements. Benchtop results demonstrated that the coefficient of determination between known flow rate and SPG was highly linear (R2 = 0.990) and was unaffected by the addition of skin-phantom layers with variable absorption (R2 = 0.996-0.999). Additionally, no significant trend was found between the fit residuals of SPG and flow rate with increasing skin-phantom absorption (R2=0.025, p = 0.29). In clinical testing, no significant difference was found using both a 4-way ANOVA between demographic classifications (F = 0.89, p = 0.45), and a 2-way ANOVA test between lower- and higher-melanin subclassifications (F = 0.4, p = 0.52).
Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://doi.org/10.1364/OA_License_v1#VOR-OA
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Signals
Keywords
How it works
- Wavelengths
- laser (λ not in abstract)
Cite this paper
Rice, T. B., Yang, B. Y., & White, S. (2020). Effect of skin optical absorption on speckleplethysmographic (SPG) signals. Biomedical Optics Express , 2020, 11(9), 5352. https://doi.org/10.1364/BOE.403501
Rice, Tyler B., Bruce Y. Yang, and Sean White. 2020. "Effect of skin optical absorption on speckleplethysmographic (SPG) signals." Biomedical Optics Express , 2020 11 (9): 5352. https://doi.org/10.1364/BOE.403501.
Rice, T. B., Yang, B. Y. and White, S. (2020) 'Effect of skin optical absorption on speckleplethysmographic (SPG) signals', Biomedical Optics Express , 2020, 11(9), p. 5352. doi: 10.1364/BOE.403501.
Rice, Tyler B., et al. "Effect of skin optical absorption on speckleplethysmographic (SPG) signals." Biomedical Optics Express , 2020, vol. 11, no. 9, 2020, p. 5352. https://doi.org/10.1364/BOE.403501.
T. B. Rice, B. Y. Yang, and S. White, "Effect of skin optical absorption on speckleplethysmographic (SPG) signals," Biomedical Optics Express , 2020, vol. 11, no. 9, p. 5352, 2020, doi: 10.1364/BOE.403501.
@article{rice2020effect,
title = {{Effect of skin optical absorption on speckleplethysmographic (SPG) signals}},
author = {Rice, Tyler B. and Yang, Bruce Y. and White, Sean},
year = {2020},
journal = {Biomedical Optics Express , 2020},
volume = {11},
number = {9},
pages = {5352},
doi = {10.1364/BOE.403501}
}Added to the ChatPPG research hub