Paperdoi:10.1016/j.jvir.2020.09.006

A Real-Time Blood Flow Measurement Device for Patients with Peripheral Artery Disease

Venue listed2021Added Oct 29, 2026

  • MR
  • DF
  • SW
  • TA

Authors: Mahmood K. Razavi, D. P. T. Flanigan, Sean M. White, Tyler B. Rice (with Laser Associated Sciences co-authors)

Journal of Vascular and Interventional Radiology 32(3):453–458, 2021

TL;DR

FlowMet, a toe-worn optical device that uses laser speckle dynamic light scattering to read blood flow, was tested in 167 limbs of 90 patients with suspected peripheral artery disease. Its waveforms resembled Doppler ultrasound, and its flow-waveform score predicted critical limb ischemia (AUC 0.83) about as well as the toe-brachial index and better than the ankle-brachial index.

  • Form factor: other

Abstract

PurposeTo evaluate the feasibility of a new optical device that measures peripheral blood flow as a diagnostic and monitoring tool for patients with peripheral artery disease (PAD).Materials and methodsIn this prospective study, 167 limbs of 90 patients (mean age, 76 y; 53% men) with suspected PAD were evaluated with the FlowMet device, which uses a new type of dynamic light-scattering technology to assess blood flow in real time. Measurements of magnitude and phasicity of blood flow were combined into a single-value flow-waveform score and compared vs ankle-brachial index (ABI), toe-brachial index (TBI), and clinical presentation of patients per Rutherford category (RC). Receiver operating characteristic curves were constructed to predict RC. Area under the curve (AUC), sensitivity, and specificity were compared among flow-waveform score, ABI, and TBI.ResultsQualitatively, the FlowMet waveforms were analogous to Doppler velocity measurements, and degradation of waveform phasicity and amplitude were observed with increasing PAD severity. Quantitatively, the flow, waveform, and composite flow-waveform scores decreased significantly with decreasing TBI. In predicting RC ≥ 4, the flow-waveform score (AUC = 0.83) showed a linear decrease with worsening patient symptoms and power comparable to that of TBI (AUC = 0.82) and better than that of ABI (AUC = 0.71). Optimal sensitivity and specificity pairs were found to be 56%/83%, 72%/81%, and 89%/74% for ABI, TBI, and flow-waveform score, respectively.ConclusionsThe technology tested in this pilot study showed a high predictive value for diagnosis of critical limb ischemia. The device showed promise as a diagnostic tool capable of providing clinical feedback in real time.

Verbatim from Europe PMC (opens in a new tab) · Licence: cc by-nc-nd

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Keywords

Speckle PlethysmographySpgFlowmetPadClinicalToeLaser Associated SciencesMedtronic

How it works

How Method works: Setup → Method → TargetSetup: Signals (SPG/DLS, ABI, TBI); Wavelengths (laser (λ not in abstract)); Form factor (other). Method: Method.SetupSignals: SPG/DLS, ABI, TBISignalsSPG/DLS, ABI, TBIWavelengths: laser (λ not in abstract)Wavelengthslaser (λ not in abstract)Form factor: otherForm factorotherMethodMethod
Wavelengths
laser (λ not in abstract)

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Razavi, M. K., Flanigan, D. P. T., White, S. M., & Rice, T. B. (2021). A Real-Time Blood Flow Measurement Device for Patients with Peripheral Artery Disease. Journal of Vascular and Interventional Radiology , 2021, 32(3), 453–458. https://doi.org/10.1016/j.jvir.2020.09.006
Razavi, Mahmood K., D. P. T. Flanigan, Sean M. White, and Tyler B. Rice. 2021. "A Real-Time Blood Flow Measurement Device for Patients with Peripheral Artery Disease." Journal of Vascular and Interventional Radiology , 2021 32 (3): 453–458. https://doi.org/10.1016/j.jvir.2020.09.006.
Razavi, M. K., Flanigan, D. P. T., White, S. M. and Rice, T. B. (2021) 'A Real-Time Blood Flow Measurement Device for Patients with Peripheral Artery Disease', Journal of Vascular and Interventional Radiology , 2021, 32(3), pp. 453–458. doi: 10.1016/j.jvir.2020.09.006.
Razavi, Mahmood K., et al. "A Real-Time Blood Flow Measurement Device for Patients with Peripheral Artery Disease." Journal of Vascular and Interventional Radiology , 2021, vol. 32, no. 3, 2021, pp. 453–458. https://doi.org/10.1016/j.jvir.2020.09.006.
M. K. Razavi, D. P. T. Flanigan, S. M. White, and T. B. Rice, "A Real-Time Blood Flow Measurement Device for Patients with Peripheral Artery Disease," Journal of Vascular and Interventional Radiology , 2021, vol. 32, no. 3, pp. 453–458, 2021, doi: 10.1016/j.jvir.2020.09.006.
@article{razavi2021realtime,
  title         = {{A Real-Time Blood Flow Measurement Device for Patients with Peripheral Artery Disease}},
  author        = {Razavi, Mahmood K. and Flanigan, D. P. T. and White, Sean M. and Rice, Tyler B.},
  year          = {2021},
  journal       = {Journal of Vascular and Interventional Radiology , 2021},
  volume        = {32},
  number        = {3},
  pages         = {453--458},
  doi           = {10.1016/j.jvir.2020.09.006}
}

Added to the ChatPPG research hub