Paperdoi:10.64898/2026.06.19.26356074
Clinical-grade Cuffless Blood Pressure Monitoring via Deep-tissue Diffuse Speckle Pulsatile Flowmetry
medRxiv, 2026
TL;DR
A preprint from ASTAR Singapore tests diffuse speckle pulsatile flowmetry, a laser-speckle method that records deep-tissue blood flow alongside a PPG-like volume signal, for cuffless blood pressure. In 133 outpatients it reported mean absolute errors of 4.17 mmHg (systolic) and 2.42 mmHg (diastolic) against a cuff, meeting ISO 81060-2 accuracy criteria and British Hypertension Society grade A. Ablations showed the flow signal improved on PPG alone.
- Form factor: other
Abstract
Blood pressure (BP) is a vital sign which is measured to diagnose and manage hypertension. However, current methods to measure BP use inflatable cuffs which cause discomfort and limit the frequency at which measurements can be made, or intra-arterial catheters which are invasive and pose infection risks. Here, we propose and evaluate the use of Diffuse Speckle Pulsatile Flowmetry (DSPF) as a cuffless BP measurement method to address these limitations. DSPF is a laser speckle-based technique which simultaneously records blood flow rate and blood volume (i.e. photoplethysmography or PPG) signals from relatively deep vascular tissue. Using information from these signals, we studied DSPF’s effectiveness in measuring systolic BP (SBP) and diastolic BP (DBP) through an outpatient study in which 133 patients were recruited, and in measuring beat-to-beat BP waveforms through an inpatient study in which two patients were recruited. In the outpatient study, the DSPF method was able to achieve mean absolute errors (MAEs) of 4.17 mmHg and 2.42 mmHg for SBP and DBP respectively compared to conventional cuff-based methods. It was also able to fulfil the requirements of the AAMI/ESH/ISO 81060-2:2018 standard for BP measurement devices and attain an ‘A’ grade according to the British Hypertension Society’s grading scheme. For the inpatient study, it produced BP waveforms which had MAEs of 2.35 mmHg and 3.06 mmHg compared to arterial-line measurements for the two patients, respectively. Compared to PPG which has been studied more extensively as a cuffless BP measurement method, we found through ablation studies that DSPF was able to reach significantly lower MAEs and hence better accuracies. DSPF augments the performance of PPG-only methods by leveraging additional information from the blood flow rate signal, and we therefore find it to be a superior cuffless BP measurement method which can potentially be used in outpatient, inpatient, and remote settings.
Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://www.medrxiv.org/about/FAQ#license
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Tasks
Signals
Keywords
How it works
- Wavelengths
- laser (λ not in abstract)
Cite this paper
Choo, T. W. J., Yap, A. A. M., Kawaja, A., Chao, V. T. T., Ng, C. J., Olivo, M., & Bi, R. (2026). Clinical-grade Cuffless Blood Pressure Monitoring via Deep-tissue Diffuse Speckle Pulsatile Flowmetry. medRxiv, 2026. https://doi.org/10.64898/2026.06.19.26356074
Choo, Tristan Wen Jie, Alicia Ann May Yap, Ariffin Kawaja, Victor T. T. Chao, Chirk Jenn Ng, Malini Olivo, and Renzhe Bi. 2026. "Clinical-grade Cuffless Blood Pressure Monitoring via Deep-tissue Diffuse Speckle Pulsatile Flowmetry." medRxiv, 2026. https://doi.org/10.64898/2026.06.19.26356074.
Choo, T. W. J., Yap, A. A. M., Kawaja, A., Chao, V. T. T., Ng, C. J., Olivo, M. and Bi, R. (2026) 'Clinical-grade Cuffless Blood Pressure Monitoring via Deep-tissue Diffuse Speckle Pulsatile Flowmetry', medRxiv, 2026. doi: 10.64898/2026.06.19.26356074.
Choo, Tristan Wen Jie, et al. "Clinical-grade Cuffless Blood Pressure Monitoring via Deep-tissue Diffuse Speckle Pulsatile Flowmetry." medRxiv, 2026, 2026. https://doi.org/10.64898/2026.06.19.26356074.
T. W. J. Choo et al., "Clinical-grade Cuffless Blood Pressure Monitoring via Deep-tissue Diffuse Speckle Pulsatile Flowmetry," medRxiv, 2026, 2026, doi: 10.64898/2026.06.19.26356074.
@misc{choo2026clinicalgrade,
title = {{Clinical-grade Cuffless Blood Pressure Monitoring via Deep-tissue Diffuse Speckle Pulsatile Flowmetry}},
author = {Choo, Tristan Wen Jie and Yap, Alicia Ann May and Kawaja, Ariffin and Chao, Victor T. T. and Ng, Chirk Jenn and Olivo, Malini and Bi, Renzhe},
year = {2026},
howpublished = {medRxiv, 2026},
doi = {10.64898/2026.06.19.26356074}
}Added to the ChatPPG research hub