What Is the Oxygen Desaturation Index (ODI)?

The oxygen desaturation index (ODI) is the number of times blood oxygen saturation (SpO2) drops by a set amount per hour of an overnight recording. ODI3 counts drops of 3 percentage points or more from the level just before the dip; ODI4 counts drops of 4 points or more, so ODI4 is never higher than ODI3 on the same night. It comes from a pulse oximeter alone. AHI is a different number: it counts apneas and hypopneas scored from airflow and breathing effort, and under the AASM recommended rule some hypopneas count with an arousal and no oxygen drop at all. The two usually move together, but they are not interchangeable, and an ODI on its own is not a diagnosis.

A watercolor of a person asleep with a plain fingertip oximeter clip wired to a small wrist recorder, and a blue ribbon above that runs level and dips in a few separate notches, the deepest washed in coral
An oximeter counts the dips in blood oxygen; ODI is how many per hour.

What the oxygen desaturation index measures

ODI is a rate of oxygen dips. An oximeter on a finger records SpO2 through the night, software marks every place where the reading falls by at least a set amount and then recovers, and the count is divided by the hours recorded:

ODI=desaturationshours of recording\mathrm{ODI} = \frac{\text{desaturations}}{\text{hours of recording}}

That is how Ng and colleagues describe it: the average number of desaturation episodes per hour of recording, usually reported as ODI3 (drops of 3% or more) and ODI4 (drops of 4% or more) (Ng et al., Journal of Clinical Sleep Medicine, 2017). The percentages are percentage points of saturation, so a fall from 95% to 92% is a 3% desaturation.

A worked example with made-up numbers: 48 dips of 3% or more over 8 hours of recording is an ODI3 of 6. If 32 of those dips were 4% or deeper, the ODI4 is 4. Every 4% dip is also a 3% dip, so on one recording ODI4 can match ODI3 but never exceed it.

Interactive · illustrative, synthetic trace, not a recording

Which oxygen dips count toward ODI3 and ODI4?

Threshold
Averaging
88%90%92%94%96%98%0 min5 min10 min15 min20 min25 min30 min12345678910
Dips counted10 in 30 min
ODI3 for this half hour20 per hour
Deepest drop seen7.0 points

10 dips of 3% or more counted in this made-up half hour with 1-second averaging, an ODI3 of 20 per hour. The same trace gives 6 dips at the 4% threshold.

Synthetic 1 Hz SpO2 for one made-up half hour, drawn to show how counting works. It is not patient data. The counting rule is simplified for teaching: the baseline is the highest value in the previous 2 minutes, a dip counts when SpO2 falls the threshold or more below it, and the dip ends when SpO2 comes back within 1 point of that baseline. Real oximetry software uses its own rules, which are usually not published (Ng et al. 2017). The averaging buttons apply a trailing moving average, the way an oximeter setting would; Farré et al. (1998) found that longer averaging times made measured desaturations up to 60% shallower.

How ODI is calculated from oximetry

Three choices decide what gets counted.

  • The baseline. A dip is measured from the level before it. For scoring hypopneas, the AASM defines the pre-event baseline as the closest stable oximetry reading before the event or, when events come so close together that oxygen never recovers, the highest value before the event (AASM, Scoring Respiratory Events). ODI software makes its own version of that call.

  • The threshold, 3% or 4%. A lower threshold also catches the smaller dips.

  • The oximeter's own processing. Oximeters average SpO2 over a few seconds before reporting it. In 12 patients with severe sleep apnea, Farré and colleagues ran two identical oximeters side by side; stretching one oximeter's averaging time from 3 seconds to 12 or 21 seconds made the measured desaturations up to 60% shallower (Farré et al., Sleep, 1998). A dip that is short and sharp can drop under the threshold just from smoothing.

There is no standard rulebook for any of this. Ng and colleagues point out that each oximeter or sleep-system maker uses its own algorithm, that the baseline level and how long a drop must last both change the count, and that the algorithms are not usually available for clinicians to check (Ng et al. 2017). AHI, by contrast, has published scoring rules.

ODI3 vs ODI4 vs AHI vs T90

MetricWhat it countsSignals neededDivided byWhat it can miss
ODI3Oxygen drops of 3% or more from the pre-dip levelPulse oximetry onlyHours of recordingBreathing events with little or no oxygen drop; it cannot tell sleep from wake
ODI4Oxygen drops of 4% or morePulse oximetry onlyHours of recordingSmaller dips, so it is never higher than ODI3 on the same recording
AHIApneas plus hypopneas, scored under rule 1A or 1BAirflow, breathing effort and oximetry, plus EEG for sleep and arousals in a lab studyHours of sleepHow deep or long the oxygen drops were
REIThe same events on a home sleep apnea testAirflow, breathing effort and oximetryHours of recordingHypopneas marked only by an arousal
T90 (CT90)Share of the night with SpO2 under 90%Pulse oximetry onlyPercent of recording timeRepeated dips that stay above 90%

ODI's denominator is from Ng et al. 2017. AHI, REI and why a home test cannot see arousal-only hypopneas are from the 2017 AASM diagnostic guideline (Kapur et al., Journal of Clinical Sleep Medicine, 2017); AHI covers them in detail. Chung and colleagues define CT90 as the cumulative time percentage with SpO2 under 90% (Chung et al., Anesthesia & Analgesia, 2012).

How ODI relates to AHI

The two numbers overlap because many breathing events end in an oxygen dip. They split apart for a few reasons.

AHI counts breathing events, not dips. An apnea counts whether or not oxygen falls much. A hypopnea needs a 30% or greater drop in airflow for at least 10 seconds plus a consequence, and the consequence depends on the rule. Under the recommended rule, called 1A, it is a 3% oxygen desaturation or an arousal. Under the acceptable rule, 1B, it must be a 4% desaturation, and arousals do not count. AASM-accredited sleep centers have to state in the report which rule they used (AASM, 2013; Berry et al., Journal of Clinical Sleep Medicine, 2012). A hypopnea scored from an arousal alone adds to AHI and adds nothing to ODI. The AASM's position is that these arousal-based events still matter, because they can cause sleep apnea symptoms without much desaturation (Malhotra et al., Journal of Clinical Sleep Medicine, 2018).

The denominators differ too. AHI divides by hours of sleep. ODI divides by hours of recording, so time spent awake in bed dilutes it.

Studies that measured both on the same night show how close they can get and where they part:

  • In 475 surgical patients, ODI predicted an AHI over 5, 15 and 30 with areas under the curve of 0.908, 0.931 and 0.958. ODI over 10 picked up 93% of people with an AHI over 15, at 75% specificity (Chung et al. 2012).

  • In 3,413 portable recordings, an ODI under 5 ruled out an AHI of 5 or more with 97.7% sensitivity and 97.0% specificity in the test set. Here ODI and AHI came from the same device and oximeter (Fabius et al., Sleep and Breathing, 2019).

  • In 296 adults from a population cohort, agreement between ODI3 and a flow-based AHI on severity category was only fair (kappa 0.32), and near chance in people with a BMI of 35 or more (kappa 0.11). The median ODI3 was 9 and the median AHI was 5 (Senaratna et al., Journal of Sleep Research, 2019).

Interactive · published results

How closely does ODI track AHI when both are measured on the same night?

475 surgical patients

AHI over 5 (cut-off ODI over 5)0.908
AHI over 15 (cut-off ODI over 15)0.931
AHI over 30 (cut-off ODI over 30)0.958

In this group ODI sorted people by AHI well, and best at the severe end. The authors also report that ODI over 10 caught 93% of people with an AHI over 15, with 75% specificity.

Chung et al., Anesthesia & Analgesia, 2012 (doi:10.1213/ANE.0b013e318248f4f5): 475 surgical patients with a portable sleep study at home and a separate high-resolution oximeter; AUC is the area under the ROC curve (1.0 is perfect separation) and accuracy is at the best cut-offs, ODI over 5, 15 and 30. Fabius et al., Sleep and Breathing, 2019 (doi:10.1007/s11325-018-1654-2): 3,413 portable recordings, rule tested on 1,132; ODI and AHI came from the same device and oximeter. Senaratna et al., Journal of Sleep Research, 2019 (doi:10.1111/jsr.12804): 296 adults from a population cohort, two-channel home study, ODI3 vs a flow-only AHI; kappa measures agreement on severity category (0 is chance, 1 is perfect). Each study used its own devices, rules and population, so the numbers do not combine.

A 2021 systematic review found 8 studies with 1,924 patients. Sensitivity ranged from 32% to 98.5% and specificity from 47.7% to 98%. The authors did not compare ODI and AHI directly across studies, because the studies used different hypopnea rules, desaturation thresholds and severity cut-offs (Rashid et al., Laryngoscope, 2021). Overnight oximetry vs a sleep study covers what that means when oximetry is used for screening.

Is there a normal ODI?

There is no official ODI scale the way there is for AHI. Papers often reuse AHI's 5, 15 and 30 cut points, and in Chung's data the most accurate ODI cut-offs did land on 5, 15 and 30 (Chung et al. 2012). The Rashid review suggested considering a 4% ODI of 15 or more per hour for diagnosing adult sleep apnea, and 10 or more as a reason for further evaluation, while saying more study was needed before a firm recommendation (Rashid et al. 2021).

Those are research thresholds, and they depend on the oximeter and software behind the number. The 2017 AASM guideline recommends against diagnosing sleep apnea from clinical tools, questionnaires or prediction algorithms without polysomnography or a home sleep apnea test (Kapur et al. 2017). For what oxygen usually does overnight in healthy sleepers, see normal SpO2 while sleeping.

Why two oximeters can give different ODIs

The same person on the same night can get two different ODIs. Ng and colleagues had 106 adults wear two oximetry systems at once during a sleep study. The two ODIs correlated well (r = 0.96 for ODI4) but agreed poorly: one system read 4.4 events per hour higher on average for ODI4 and 7.1 higher for ODI3, with wide limits. When one recording was run through both programs, the difference nearly vanished, which points at the hardware and its signal processing (Ng et al. 2017).

Interactive · published results

Two oximetry systems, same people, same night: how far apart were their ODIs?

-10-50+5+10+15+20+25Two systems: each with its own oximeter and software-5.8 to +14.6One recording (AL oximetry) scored by both programs-2.7 to +2.8

Difference in ODI4, events per hour (ApneaLink minus Compumedics)

ODI4: on average the ApneaLink read 4.4 events per hour higher, and for 95% of people the gap fell between 5.8 lower and 14.6 higher. Scored from one recording, the two programs differed by 0.03 on average.

Ng et al., Journal of Clinical Sleep Medicine, 2017 (doi:10.5664/jcsm.6552): 106 adults referred for a sleep study wore a ResMed ApneaLink Plus and the lab's Compumedics system at the same time. The dot is the mean difference (ApneaLink minus Compumedics) and the bar is the 95% limits of agreement, in events per hour. When the ApneaLink recording was exported and scored by both programs, the gap almost disappeared, so the authors put the difference down to how each device acquires and processes the signal rather than to the counting software. The category view uses ODI3 cut-offs the authors applied for their analysis, not a clinical standard.

The oximeter matters for AHI as well, since hypopneas depend on a desaturation. In 113 patients wearing two of four common oximeters during polysomnography, three gave a similar AHI under the 4% rule, but the fourth raised it by 3.7 events per hour on average. That was enough to move 7 people across the 15-per-hour line (Zafar et al., Chest, 2005).

Other limits of ODI

  • Movement or a sensor that slips can add false dips or leave gaps, and oximeters differ in how they reject those artifacts. Poor blood flow to the finger also lowers SpO2 accuracy, which Ng and colleagues list as one reason two oximeters on different fingers can disagree.

  • Because ODI divides by recording time, a long stretch awake lowers it, and dips during wake still count.

  • In Senaratna's cohort, ODI3 and AHI agreed least in people with a BMI of 35 or more. A separate study of 1,333 patient records reported that the agreement between AHI and ODI differed between people of normal weight and people with obesity (Ernst et al., Sleep and Breathing, 2016).

  • ODI has no breathing signal, so it cannot say whether a dip came from an obstructive or a central event, or from something other than breathing. Central sleep apnea explains why that distinction needs effort sensors.

  • A single night is a sample, and nights vary.

Where wearables fit

Some watches and rings estimate SpO2 from wrist or finger PPG and report overnight dips. How a wearable computes an ODI-like number, and what makes it more or less trustworthy, is covered in PPG oxygen desaturation index in wearables. How PPG models build an AHI-like estimate is in wearable AHI estimation, and published results are on the sleep apnea AHI leaderboard. The sleep apnea condition page covers PPG detection more broadly, and Sensor Bio keeps a review of wearable sleep apnea studies. The sleep apnea hub puts AHI, ODI, home tests and lab studies side by side.

This page explains a number. It is not a diagnosis and not medical advice. An ODI, from a clinic oximeter or a wearable, is read by a clinician alongside symptoms and, when needed, a sleep study.

References

  • AASM. AASM clarifies hypopnea scoring criteria (rules 1A and 1B). 2013.

  • AASM Inter-scorer Reliability Program. Scoring respiratory events (pre-event oximetry baseline).

  • Berry RB, et al. Rules for scoring respiratory events in sleep: update of the 2007 AASM manual. J Clin Sleep Med. 2012. doi:10.5664/jcsm.2172

  • Chung F, et al. Oxygen desaturation index from nocturnal oximetry: a sensitive and specific tool to detect sleep-disordered breathing in surgical patients. Anesth Analg. 2012. doi:10.1213/ANE.0b013e318248f4f5

  • Ernst G, et al. Difference between apnea-hypopnea index (AHI) and oxygen desaturation index (ODI): proportional increase associated with degree of obesity. Sleep Breath. 2016. doi:10.1007/s11325-016-1330-3

  • Fabius TM, et al. Validation of the oxygen desaturation index in the diagnostic workup of obstructive sleep apnea. Sleep Breath. 2019. doi:10.1007/s11325-018-1654-2

  • Farré R, et al. Importance of the pulse oximeter averaging time when measuring oxygen desaturation in sleep apnea. Sleep. 1998. doi:10.1093/sleep/21.4.386

  • Kapur VK, et al. Clinical practice guideline for diagnostic testing for adult obstructive sleep apnea: an AASM clinical practice guideline. J Clin Sleep Med. 2017. doi:10.5664/jcsm.6506

  • Malhotra RK, et al. Polysomnography for obstructive sleep apnea should include arousal-based scoring: an AASM position statement. J Clin Sleep Med. 2018. doi:10.5664/jcsm.7234

  • Ng Y, et al. Oxygen desaturation index differs significantly between types of sleep software. J Clin Sleep Med. 2017. doi:10.5664/jcsm.6552

  • Rashid NH, et al. The value of oxygen desaturation index for diagnosing obstructive sleep apnea: a systematic review. Laryngoscope. 2021. doi:10.1002/lary.28663

  • Senaratna CV, et al. Comparison of apnoea-hypopnoea index and oxygen desaturation index when identifying obstructive sleep apnoea using type-4 sleep studies. J Sleep Res. 2019. doi:10.1111/jsr.12804

  • Zafar S, et al. Choice of oximeter affects apnea-hypopnea index. Chest. 2005. doi:10.1378/chest.127.1.80

Go deeper

Frequently Asked Questions

What is the oxygen desaturation index?

The oxygen desaturation index (ODI) is the number of drops in blood oxygen saturation per hour of an overnight oximetry recording. A drop counts when SpO2 falls a set amount below the level before it, usually 3 percentage points (ODI3) or 4 (ODI4) (Ng et al. 2017, doi:10.5664/jcsm.6552).

What is the difference between ODI3 and ODI4?

ODI3 counts oxygen drops of 3% or more and ODI4 counts drops of 4% or more. Every 4% drop is also a 3% drop, so on one recording ODI4 is equal to or lower than ODI3. Reports should say which threshold was used, because the numbers are not the same.

Is ODI the same as AHI?

No. AHI counts apneas and hypopneas per hour of sleep from airflow, effort and oxygen signals. ODI counts oxygen drops per hour of recording from an oximeter alone. Under the AASM recommended rule (1A), a hypopnea can be scored from an arousal with no oxygen drop, which raises AHI but not ODI. The two correlate but are not interchangeable; one population study found only fair agreement on severity (kappa 0.32; Senaratna et al. 2019, doi:10.1111/jsr.12804).

What is a normal ODI?

There is no official ODI scale. Studies often reuse AHI's 5, 15 and 30 per hour cut points, and a 2021 systematic review suggested considering a 4% ODI of 15 or more for diagnosing adult sleep apnea and 10 or more for further evaluation, while calling for more research (Rashid et al. 2021, doi:10.1002/lary.28663). These are research thresholds, not a self-test, and they depend on the oximeter used.

Why do two oximeters give different ODIs?

Oximeters process the signal differently, including how long they average SpO2, and each maker uses its own dip-counting rules. Worn together by 106 adults, two systems differed by 4.4 events per hour on average for ODI4, with limits of agreement from 5.8 lower to 14.6 higher (Ng et al. 2017, doi:10.5664/jcsm.6552). Longer averaging can make desaturations up to 60% shallower (Farré et al. 1998, doi:10.1093/sleep/21.4.386).

Can a smartwatch measure ODI?

Some watches and rings estimate overnight SpO2 dips from PPG, which is related to ODI, but the result depends on fit, motion and the device's own processing. It is not a diagnosis. How wearables compute it is covered in our wearable ODI guide, and a sleep study is how sleep apnea is diagnosed (Kapur et al. 2017, doi:10.5664/jcsm.6506).