BVP Meaning: Blood Volume Pulse vs PPG, and Why It Flips

BVP meaning: blood volume pulse is a PPG waveform under another name. How Empatica makes it, why some exports are upside down, and how to fix polarity.

ChatPPG Team
Published
Updated
21 min read
  • PPG & Signal Processing
  • PPG
Real wrist PPG from the GalaxyPPG dataset (Park et al. 2025, CC BY 4.0): raw Galaxy Watch5 counts, the same samples inverted and band-passed, and Empatica E4 BVP. Inversion and filtering by ChatPPG
Real wrist PPG from the GalaxyPPG dataset (Park et al. 2025, CC BY 4.0): raw Galaxy Watch5 counts, the same samples inverted and band-passed, and Empatica E4 BVP. Inversion and filtering by ChatPPG.

Short answer: BVP (blood volume pulse) is the pulsatile waveform recorded by a photoplethysmography (PPG) sensor, so a BVP signal is a PPG signal under another name. The label comes from psychophysiology and biofeedback. Empatica uses it for the processed PPG output of its E4 and EmbracePlus wristbands, which is why datasets such as WESAD and PPG-DaLiA store wrist PPG in files called BVP.csv.

The name says little about the processing. A file labelled BVP may be centred on zero or sit on a large DC level, may be filtered or raw, and may even be stored upside down. Below we cover what BVP means, how Empatica produces it, and how to check the sign of any PPG or BVP trace, using real recordings from public datasets.

This article is educational and is not medical advice.

Key facts

Scroll horizontally if needed

Fact Detail Source
What BVP is The pulsatile PPG waveform, labelled by its physiological meaning (a pulse of blood volume) rather than its method (photoplethysmography) Peper et al. 2007; Thought Technology HRV manual
Empatica E4 BVP 64 Hz, made by a proprietary algorithm that combines green and red light; exported as BVP.csv Empatica blog and E4 spec sheet
Empatica EmbracePlus BVP Stored in Avro files under rawData.bvp, values documented as "light absorption (nW)", with a separate systolic-peaks stream Empatica Data Access Documentation
Units Relative. Not millilitres of blood and not mmHg Thought Technology manual; Peper et al. 2007
Polarity Detected light falls as blood volume rises, so raw PPG dips at systole; most displays invert it so systole points up Reisner et al. 2008
Real exports differ The E4 BVP.csv we checked (GalaxyPPG) was systole-up; an EmbracePlus export was systole-down in all 18 sessions of one public dataset ChatPPG analysis of GalaxyPPG; Em and Teoh dataset
Typical pulsatile share In one raw Galaxy Watch5 recording, the beat was 0.67% of the DC level ChatPPG analysis of GalaxyPPG

What BVP means

A PPG sensor shines light into skin and measures how much reaches a photodiode. Each heartbeat pushes a small extra volume of arterial blood into the tissue under the sensor, which changes how much light is absorbed, and the resulting waveform rises and falls once per beat. Engineers call it the photoplethysmogram, or PPG. Psychophysiologists named the same trace the blood volume pulse, after what it tracks: a pulse in local blood volume (Peper et al. 2007).

So "PPG vs BVP" is mostly a naming difference:

  • PPG names the method: optical (photo) measurement of volume change (plethysmography).
  • BVP names the quantity the waveform represents: the blood volume pulse at the measurement site.

Thought Technology's HRV manual says this directly. Heartbeats can be detected with ECG or with "Blood Volume Pulse (BVP, also called photoplethysmography)", and the BVP sensor's unit of measure is "relative (%)" (Thought Technology, HRV applied to psychophysiology).

In practice, a file labelled "BVP" usually came from a PPG sensor, and someone has already turned the raw light reading into a pulse waveform. How they did it varies, so check.

Where the name came from

Plethysmography means measuring volume change. Alrick Hertzman's 1937 paper on "photoelectric plethysmography of the fingers and toes in man" is the usual starting point for optical plethysmography (Hertzman 1937). John Allen's 2007 review traces its later clinical use, from pulse oximetry to vascular assessment (Allen 2007; ChatPPG paper card).

The BVP label took hold in biofeedback and psychophysiology, where clinical systems sold finger "BVP sensors" for heart rate variability training, peripheral vasoconstriction and relaxation work. Peper and colleagues described BVP as the signal derived from a PPG sensor and argued that its amplitude and shape carry information beyond heart rate (Peper et al. 2007). Vendors still use the name, for example Mind Media, PLUX and Thought Technology. So do research papers, such as Kushki et al.'s study of BVP and skin conductance at different body sites (Kushki et al. 2011).

The term then crossed into camera-based PPG. Poh, McDuff and Picard's 2010 webcam study validated video-derived pulse rate against "an FDA-approved finger blood volume pulse (BVP) sensor" (Poh et al. 2010). Since then, much of the remote PPG literature has called the waveform it recovers from video the BVP. Open-source tools do the same: rPPG-Toolbox reads the UBFC-rPPG ground-truth pulse oximeter trace through a function documented as reading "a bvp signal file" (rPPG-Toolbox UBFC loader), and the iBVP dataset is named after imaging BVP (Joshi and Cho 2024).

PPG vs BVP at a glance

Scroll horizontally if needed

PPG (photoplethysmogram) BVP (blood volume pulse)
What the name describes The optical method The physiological quantity the waveform tracks
Where the term is common Engineering, clinical monitoring, smartwatches, pulse oximetry Biofeedback, psychophysiology, affective computing, Empatica devices, rPPG papers
Typical content Anything from raw photodiode counts (with DC) to a cleaned pulse Usually a pulse waveform that has already been processed in some way
Units ADC counts, current, voltage, or arbitrary units Relative or arbitrary units; Empatica EmbracePlus documents nW of light absorption
DC component Often present in raw device data Often removed (E4 BVP.csv values centre near zero)
Sign Raw detector output dips at systole Varies by vendor and export; check every file
Typical outputs derived from it Heart rate, IBI, SpO2 (needs two wavelengths), perfusion, morphology Heart rate, IBI and HRV, pulse amplitude, morphology

Physiologically, BVP and PPG are the same signal. A BVP file is a PPG waveform that has been labelled, and often preprocessed, by whoever made it.

How Empatica produces BVP

Empatica E4

The E4 was a research wristband used widely in stress and affect studies. Empatica describes its BVP as "the primary output from the E4 PPG sensor", obtained "by a proprietary algorithm that combines the light signals observed during both green and red exposure", at a fixed 64 Hz (Empatica blog). The 2014 E4 specification lists two green and two red LEDs, two photodiodes with 14 mm² total sensitive area, a sensor output resolution of 0.9 nW per digit and a motion artefact removal algorithm that "combines different light wavelengths" (E4 tech specs). For why devices pair wavelengths, see green vs red vs infrared PPG.

The E4 export is a folder of CSV files. BVP.csv holds one column: the Unix start time in the first row, the sampling rate (64.0) in the second, then one value per sample. The same export includes IBI.csv (inter-beat intervals derived from BVP), HR.csv, ACC.csv (32 Hz), EDA.csv and TEMP.csv (4 Hz) (Empatica blog).

Before you analyse E4 BVP, know four things:

  • It is processed, not raw. You get one combined channel instead of separate green and red channels, and the combination algorithm is proprietary.
  • The export does not label the unit. The spec sheet gives 0.9 nW per digit, while at least one third-party importer describes the data as volts (MindWare KB0053). For anything that crosses devices or people, treat the values as arbitrary units.
  • It is centred near zero, so there is no DC level to compute an AC/DC ratio or perfusion index from.
  • In the E4 BVP.csv we checked, from the GalaxyPPG dataset, it is systole-up: the fast upstroke rises (details below).

Empatica EmbracePlus

EmbracePlus is Empatica's current wristband. Its raw data arrive as Avro files of about 30 minutes each. The BVP stream sits under rawData.bvp with a start timestamp in microseconds, a sampling frequency field and an array of values documented as "light absorption (nW)". A separate systolicPeaks stream holds the timestamps at which the device detected systolic peaks (Empatica Data Access Documentation; Empatica Academic Support). For other devices that expose raw signals, see the raw data access table.

The word "absorption" has a polarity implication: more blood means more absorption, so an absorption-like signal should rise at systole. One public EmbracePlus dataset found the opposite in its exports. Em and Teoh recorded 18 real-world driving sessions on EmbracePlus firmware 4.19.2, and their README states that the exported BVP "sign is inverted relative to the waveform on which the device marks its systolic peaks, so multiply by -1 before peak detection" (Em and Teoh dataset, Zenodo 23718697). Their polarity check found the exported signal below its mean at the device's own peak timestamps in all 18 of 18 sessions. Figure 3 shows one of those sessions.

We have seen this in only one dataset and one firmware version. Treat it as a reason to check your own export, not as a rule for every EmbracePlus file.

Why PPG and BVP are sometimes upside down

The physics: systole is a dip in detected light

Blood absorbs light. When a pulse of blood arrives under the sensor, more light is absorbed along the path and less reaches the photodiode, so the raw detector signal falls during systole and recovers during diastole. Reisner and colleagues put it plainly: a pulse of blood "decreases the light intensity at the photodetector", and "by convention, the photoplethysmogram is inverted so that it correlates positively with blood volume" (Reisner et al. 2008).

This holds for both sensor layouts:

  • Transmission PPG (finger clip, earlobe): the LED and photodiode sit on opposite sides of the tissue. More blood in the path means less transmitted light.
  • Reflectance PPG (wrist, ring, forehead): the LED and photodiode sit side by side, and the detector collects light scattered back from tissue. More blood still means less detected light in nearly all recordings (transmission vs reflection).

There are rare exceptions. Jerve and colleagues studied reflectance PPG above the dorsalis pedis artery in the foot, where detected intensity sometimes rises at systole, an anomaly they call PPG signal inversion (Jerve et al. 2024). Kamshilin and colleagues saw counter-phase imaging PPG waveforms in adjacent areas of the palm and wrist (Kamshilin et al. 2015). These cases are uncommon, but they are one more reason to check the sign on the data itself.

Figure 1 shows the convention on real data. The top panel is a raw Galaxy Watch5 PPG recording in which every heartbeat is a sharp dip. Inverting and band-pass filtering the same samples gives the familiar upright pulse. The bottom panel is an Empatica E4 BVP.csv from the same participant and session, which arrives already systole-up and centred on zero.

Three panels of real wrist PPG from the GalaxyPPG dataset. A: raw Galaxy Watch5 counts where each beat is a dip. B: the same samples inverted and band-passed, with systole pointing up. C: Empatica E4 BVP as exported, already systole-up.

Figure 1. Real data from GalaxyPPG (Park et al. 2025; Zenodo, CC BY 4.0), participant P02, seated baseline. The watch and the E4 were on different wrists with unsynchronised clocks, so the panels are not sample-aligned. Inversion and filtering by ChatPPG.

AC and DC: why the dip is hard to see in raw data

Most of the detected light never meets pulsing arterial blood. Skin, bone, venous blood and the steady part of arterial blood absorb or scatter it, and together they form the large DC level. The heartbeat is a small AC ripple on top (AC/DC ratio explained). In the Galaxy Watch5 recording in Figure 2, the median beat was about 13,620 counts on a DC level of about 2,026,154 counts, or 0.67%. On a zero-based axis the pulse is almost invisible.

Two panels of raw Galaxy Watch5 PPG. A: 60 seconds on a zero-based axis showing the pulse as a tiny ripple on a large DC level. B: an 8-second zoom with the mean removed, showing beats of about 13,620 counts.

Figure 2. Real data from GalaxyPPG (Park et al. 2025; Zenodo, CC BY 4.0), P02 seated baseline. AC is the median peak-to-peak beat amplitude in an 8 s window and DC is the window mean, both in raw ADC counts. One recording, not a population value. Analysis by ChatPPG.

The same physics explains why BVP files often look nothing like raw device data. Removing the DC level (baseline wander removal), inverting and band-pass filtering are usually the first three steps in turning raw PPG into a "BVP" trace.

Who flips the sign

There is no single standard, so the sign depends on who produced the file:

Scroll horizontally if needed

Signal Stored polarity How we know
Raw photodiode or ADC output (most devices) Systole down Optics (Reisner et al. 2008); Galaxy Watch5 raw values, Figure 1
Clinical pulse oximeter pleth display Systole up Inverted by convention (Reisner et al. 2008); BIDMC monitor PLETH, Figure 4
Empatica E4 BVP.csv Systole up Checked in GalaxyPPG (Figure 1 and the table below)
Empatica EmbracePlus BVP export (firmware 4.19.2, one dataset) Systole down Em and Teoh README; device peak timestamps, Figure 3
Biofeedback BVP sensors Displayed systole up Vendor manuals describe the trace rising with each beat (Thought Technology)
Camera rPPG output Method dependent; ICA output sign is arbitrary ICA cannot determine the sign of its components (Hyvärinen and Oja 2000)

A caveat on biofeedback manuals: some describe the sensor as reflecting more light when blood volume rises. The optics literature describes the opposite at the detector (Reisner et al. 2008). Whichever description is right for a given sensor, what users see on those systems is a trace that rises with each beat.

Raw multi-wavelength exports need the same check. Sensor Bio V2, for example, exports unfiltered green, infrared and red amplitudes in arbitrary units with nanosecond timestamps, plus raw accelerometer data (raw data format). Its documentation does not state a sign convention, so run the slope check below on each channel before peak detection.

For camera PPG, the raw colour traces behave like any other PPG, but the methods that turn them into a pulse do not all preserve sign. Independent component analysis is the clearest case, since ICA recovers components only up to an unknown sign and scale (Hyvärinen and Oja 2000; see also ICA for PPG). Deep rPPG models trained against a ground-truth BVP file learn whatever sign that file uses, which is another reason to standardise polarity before training.

EmbracePlus BVP as exported with device systolic-peak timestamps marked. A: the timestamps fall on troughs. B: after multiplying by -1 and band-pass filtering, the timestamps fall on peaks.

Figure 3. Real data from Em and Teoh's EmbracePlus driving dataset (Zenodo 23718697, CC BY 4.0), session S01. Green dashes are the device's own systolic-peak timestamps. Inversion and filtering by ChatPPG.

Try it on real signals

Why is my PPG upside down? Flip real signals yourself

Pick a recording, then toggle inversion and filtering. In an upright PPG the fast upstroke (systole) rises and the slower decay falls. Green dashes on the EmbracePlus trace are the device's own systolic-peak timestamps.

Stored polarity: systole-down (raw photodiode counts). Slope check (steepest rise / steepest fall) = 0.36, so systole points down as shown (unfiltered drift can bias this).

01234567seconds
Real data: GalaxyPPG P02 seated baseline. Sources: GalaxyPPG (Park et al. 2025) (CC BY 4.0); BIDMC PPG and Respiration Dataset (Pimentel et al. 2017, PhysioNet) (ODC-By 1.0); EmbracePlus real-world driving dataset (Em and Teoh, Zenodo) (CC BY 4.0). Zero-phase 2nd-order Butterworth band-pass (0.5-8 Hz wrist, 0.5-15 Hz finger) over a 28 s context, then cut to 8 s. Windows picked by a template-correlation quality score. Modified by ChatPPG.

Interactive: pick a recording (Galaxy Watch5 raw, Empatica E4 BVP, EmbracePlus BVP or BIDMC finger PLETH), then toggle inversion and band-pass filtering. The slope check updates live. All four traces are real samples, with sources and licences cited under the chart.

What a real BVP waveform looks like

An upright PPG or BVP beat has a recognisable shape (PPG waveform basics; morphology):

  • Foot and upstroke. A fast rise as the pulse arrives, the steepest part of the beat.
  • Systolic peak. The maximum, shortly after the upstroke.
  • Dicrotic notch. A dip or shoulder on the way down, linked to the end of ventricular ejection and to reflected pressure waves (dicrotic notch physiology; notch detection).
  • Diastolic wave. A smaller second bump after the notch, followed by a slow decay to the next foot.

How much of this you can see depends on age, site and processing. The notch and diastolic wave are usually clear in young adults and fade or disappear in older adults as arteries stiffen (Charlton et al. 2022, paper card; Peper et al. 2007). Figure 4 shows both cases. The finger PLETH trace from an 88-year-old ICU patient has sharp peaks and no visible notch. In the median wrist beats from a 25-year-old participant, the E4 BVP shows a clear notch about 0.16 s after the systolic peak and a diastolic wave at about 0.25 s, and the inverted Galaxy Watch5 signal has a similar shoulder.

Two panels. A: 8 seconds of finger PPG from an 88-year-old ICU patient, sharp peaks with no visible notch. B: median beat shapes for finger PPG and two wrist signals, with the notch and diastolic wave marked on the E4 trace.

Figure 4. Real data from the BIDMC PPG and Respiration Dataset (Pimentel et al. 2017; PhysioNet, ODC-By 1.0), subject 01, and GalaxyPPG (Park et al. 2025, CC BY 4.0), P02. Right panel: median of 89 finger beats and 194 beats per wrist device, each peak-aligned and scaled 0 to 1; shading is the interquartile range. These are single participants and do not represent age groups or devices in general. Alignment and scaling by ChatPPG.

Finger and wrist signals also differ in practice. Finger sites have dense vascular beds and give large, clean pulses, while wrist reflectance signals are smaller and far more exposed to motion. Figure 5 shows what jogging does to both wrist signals: step-rate motion shows up in the E4 BVP and in the Galaxy Watch5 signal, and the pulse can no longer be read beat by beat. Empatica's own E4 spec says heart rate data estimated from its PPG sensor are "not suitable for running or physical activity" (E4 tech specs).

Three panels during jogging: E4 accelerometer magnitude with regular step peaks, E4 BVP dominated by motion, and Galaxy Watch5 PPG with irregular waves.

Figure 5. Real data from GalaxyPPG (Park et al. 2025; Zenodo, CC BY 4.0), P02, 10 s starting 30 s into the jogging session. See motion artifact detection.

How to detect and fix polarity in code

Check the sign before you detect peaks, compute features or train a model. On an upside-down trace your "systolic peaks" land on the troughs, every beat time shifts by the length of the upstroke, and morphology features come out scrambled.

Method 1: compare the steepest rise with the steepest fall. In an upright beat the upstroke is steeper than the decay, so in a band-passed trace the largest positive slope should be bigger than the largest negative slope. The function below takes the median of that ratio over 2-second windows.

import numpy as np
from scipy.signal import butter, filtfilt

def bandpass(x, fs, lo=0.5, hi=8.0, order=2):
    b, a = butter(order, [lo / (fs / 2), hi / (fs / 2)], btype="band")
    return filtfilt(b, a, x)

def slope_ratio(x, fs, win_s=2.0):
    """Median of (steepest rise / steepest fall) over windows. >1 means systole points up."""
    d = np.diff(bandpass(np.asarray(x, float), fs)) * fs
    w = int(win_s * fs)
    r = [d[i:i + w].max() / abs(d[i:i + w].min()) for i in range(0, len(d) - w, w)]
    return float(np.median(r))

def to_systole_up(x, fs):
    """Return the signal with systole pointing up (flip if needed) and the ratio used."""
    x = np.asarray(x, float)
    r = slope_ratio(x, fs)
    return (x if r > 1 else -x), r

Use at least 30 seconds at rest. The check can fail during heavy motion, with arrhythmias, or on unusual waveforms such as the inverted pulses described above, so log the ratio and look at a plot whenever it is close to 1.

Here are the values for the recordings shown in this article (0.5 to 8 Hz band-pass for wrist, 2-second windows, a few minutes of rest):

Scroll horizontally if needed

Recording Slope ratio Reading
E4 BVP.csv, GalaxyPPG P02 2.30 Systole up
BIDMC finger PLETH, subject 01 1.74 Systole up
Galaxy Watch5 raw values, GalaxyPPG P02 0.45 Systole down
EmbracePlus BVP export, Em and Teoh S01 0.69 Systole down

Method 2: use a timing reference. If the device gives beat timestamps (EmbracePlus systolic peaks) or you have a simultaneous ECG, average the signal around those times. Em and Teoh's polarity check does exactly this with the EmbracePlus peak stream (Zenodo 23718697). With ECG, each R wave should be followed by a pulse upstroke at the wrist, not a downstroke.

Method 3, with care: skewness. Skewness is a useful signal quality index (Elgendi 2016; paper card), but we do not rely on it as a sign test on its own. Treat it as a quality check and confirm polarity with Method 1 or 2.

Once the sign is fixed, record it in your metadata, for example polarity: systole_up, method: slope_ratio, value: 2.30. Peak detectors that expect upright pulses, such as those in the peak detection guide and peak-valley detection, will then behave as designed.

Datasets that label PPG as BVP

Many public datasets inherit the BVP name from Empatica hardware or from rPPG conventions. Check the sign and preprocessing of each one rather than assuming they match.

Scroll horizontally if needed

Dataset What is called BVP Rate Notes
WESAD Empatica E4 wrist BVP.csv 64 Hz Readme describes BVP.csv as "Data from photoplethysmograph (PPG)"; chest ECG at 700 Hz (Schmidt et al. 2018; UCI)
PPG-DaLiA Empatica E4 wrist BVP 64 Hz Chest ECG reference; built for heart rate under daily-life motion (Reiss et al. 2019)
GalaxyPPG Empatica E4 wrist BVP, next to raw Galaxy Watch5 PPG 64 Hz (E4), 25 Hz (watch) Polar H10 ECG reference; CC BY 4.0 (Park et al. 2025)
UBFC-Phys Wristband BVP alongside face video Not confirmed here; check the dataset documentation Files named bvp_s*_T*.csv (Sabour et al. 2023)
EmbracePlus driving EmbracePlus BVP export plus device systolic peaks 64 Hz Exported sign inverted per its README; CC BY 4.0 (Zenodo 23,718,697)
Wearable Device Dataset (E4 stress and exercise) Empatica E4 BVP 64 Hz PhysioNet
UBFC-rPPG Finger pulse oximeter waveform (CMS50E) used as rPPG ground truth See dataset The file calls it PPG; rPPG-Toolbox reads it as BVP (UBFC-rPPG)
iBVP PPG ground truth for RGB and thermal video See paper Name means imaging BVP; check the paper for the reference device (Joshi and Cho 2024)

Licences vary a lot across these datasets. Some allow any use with attribution and some are restricted to non-commercial research, so read each licence before reuse.

If you plan to train or evaluate models on these files, see BVP datasets for wrist PPG models for polarity, preprocessing and evaluation pitfalls.

From BVP to heart rate, IBI and HRV

The main use of BVP is beat timing. Find each pulse, measure the time between consecutive pulses (the inter-beat interval, IBI), and compute heart rate and HRV from the IBI series.

Empatica does this on board for the E4. Its IBI.csv lists beat times and intervals at 1/64 s resolution. Data are reported "only when BVP signal is clear", non-prototypical beats are discarded and nothing is smoothed, so gaps are expected (E4 tech specs; Empatica blog). Kubios recommends importing the E4 BVP.csv rather than IBI.csv, so you can run your own beat detection, edit detections and mark noisy segments (Kubios guide). The WESAD readme says to ignore the derived HR.csv and IBI.csv files in that dataset.

Three cautions apply:

  • Strictly, it is pulse rate variability. Intervals between pulses at the wrist are not identical to ECG R-R intervals, because pulse arrival time varies from beat to beat (PRV vs HRV; pulse rate variability).
  • 64 Hz limits timing resolution. One sample is 15.6 ms, so beat times need interpolation for fine-grained HRV.
  • Polarity affects the fiducial point. Detecting peaks on an upside-down trace marks the foot instead of the peak, which shifts every beat by the length of the upstroke.

BVP amplitude

The size of each pulse, often called blood volume amplitude (BVA) or pulse amplitude, is the other classic BVP measure. Biofeedback uses it as an index of peripheral vasoconstriction: sympathetic arousal narrows skin arterioles and shrinks the pulse, and relaxation widens them (Peper et al. 2007). Peper and colleagues express changes as a percentage within a session.

Pulse amplitude is relative. It depends on sensor contact, skin temperature, site, LED drive current and device processing, so compare it within one person and session, not across people or devices. With a DC level available you can normalise amplitude to the perfusion index; with a zero-centred BVP export you cannot.

Common misconceptions

This article is educational and is not medical advice.

  • "BVP measures blood volume." It tracks relative change in local blood volume under the sensor. It does not report millilitres of blood or total blood volume.
  • "BVP is blood pressure." The letters overlap with BP and the waveform shape relates to arterial mechanics, but a BVP signal does not give mmHg on its own.
  • "BVP and PPG are different sensors." Usually not. A BVP sensor is a PPG sensor, and the label describes the output.
  • "BVP is the same as IBI or heart rate." IBI and heart rate are derived from BVP by beat detection.
  • "BVP and plethysmogram mean different things." A plethysmogram is any volume-change recording, a photoplethysmogram is the optical kind, and BVP is the pulsatile part of it.
  • "BVP means something only in wearables." The acronym is crowded. In mathematics BVP usually means boundary value problem, and in finance it often means Bessemer Venture Partners.
  • "Upside-down data means a broken sensor." Usually it is a sign convention, which you can fix once and record in your metadata.

References

Frequently Asked Questions

What is BVP in PPG?
BVP (blood volume pulse) is the pulsatile waveform a PPG sensor records as each heartbeat changes the blood volume under the sensor. The two terms describe the same signal: BVP names what it represents and PPG names how it is measured.
Is BVP the same as PPG?
Physiologically, yes. In files, BVP usually means a PPG signal that has already been processed, for example filtered, centred on zero or combined across wavelengths, so check the preprocessing and the sign before comparing it with raw PPG.
What is the Empatica BVP signal?
On the E4 it is a 64 Hz single-channel waveform produced by a proprietary algorithm that combines green and red light. On EmbracePlus it is the bvp stream in the raw Avro files, documented as light absorption in nW, next to a separate systolic-peaks stream.
Why is my PPG or BVP signal upside down?
Blood absorbs light, so the raw detector output dips at each heartbeat. Many devices and monitors invert it so systole points up, but not every export does. One public EmbracePlus dataset reports that its exported BVP has to be multiplied by -1.
How do I check if my BVP is inverted?
Band-pass filter it and compare the steepest rise with the steepest fall. In an upright beat the upstroke is steeper, so the ratio is above 1. If you have beat timestamps or an ECG, check that pulses peak just after them.
What units is BVP in?
Relative units. Biofeedback systems report percentage change. Empatica documents EmbracePlus values as nW of light absorption and gives the E4 sensor resolution as 0.9 nW per digit. None of these are blood volume or pressure units.
Can I calculate HRV from BVP?
Yes, from the inter-beat intervals between detected pulses, ideally at rest. Strictly, this is pulse rate variability, and motion, low sampling rates and wrong polarity all add error. This is educational information, not medical advice.
What does BVP amplitude mean?
It is the height of each pulse, often used as a relative index of peripheral vasoconstriction or vasodilation. Compare it within one person and one session, not across people or devices.
Which datasets use BVP?
WESAD, PPG-DaLiA, GalaxyPPG and other Empatica E4 datasets store wrist PPG as BVP at 64 Hz. UBFC-Phys includes wristband BVP files, and rPPG datasets and toolboxes often call their ground-truth pulse waveform BVP.