Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals

Published articlePublished Jul 23, 2026Added Oct 11, 2026

  • ZZ
  • PQ
  • HT

Authors: Zichuan Zhang, Peng Qu, Hong Tang (Dalian University of Technology)

Sensors (MDPI) 26(15):4676

Key results

All results

Results as reported by the authors; not independently reproduced. A SOTA star marks the best value among results reported under the same evaluation protocol; ties go to the earlier publication.

TL;DR

Handcrafted PTT/HRV/PPG-morphology features + masked-reconstruction SSL embeddings + subject prior BP (alpha-weighted base BP from the earliest calibration file) → XGBoost. MAE 7.25/3.98 (private set) and 7.85/4.02 mmHg (UCI). Small-sample feasibility per the authors.

  • Form factor: clinical_finger

Abstract

Continuous and noninvasive blood pressure (BP) monitoring is important for hypertension screening and long-term cardiovascular management. Continuous BP estimation using photoplethysmography (PPG) and electrocardiography (ECG) has become a promising alternative to conventional cuff-based measurements. However, handcrafted feature-based methods are sensitive to fiducial point detection, while end-to-end deep models often require large labeled datasets and may ignore subject-specific BP baselines. This study proposes a personalized calibration and hybrid feature fusion framework for continuous BP estimation from ECG and PPG signals. The framework integrates physiologically interpretable handcrafted features, self-supervised waveform representations, and subject-specific prior BP information to predict systolic and diastolic BP. The handcrafted branch extracts pulse transit time, heart rate variability, and PPG morphological descriptors, whereas a masked reconstruction-based self-supervised encoder learns latent waveform embeddings without BP labels. Personalized calibration incorporates base BP from the earliest calibration file through an alpha-weighted strategy, and the fused representation is fed into XGBoost regressors. The framework was evaluated on a private dataset and the BP-UCI dataset, achieving SBP/DBP MAEs of 7.25/3.98 mmHg and 7.85/4.02 mmHg, respectively. Comparative and ablation results indicate improved baseline performance in the current experimental setting, suggesting the methodological feasibility of the proposed framework for small-sample continuous BP estimation. Further validation with larger and more diverse cohorts is still required.

Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://creativecommons.org/licenses/by/4.0/

Results on ChatPPG leaderboards · 1

All leaderboards
LeaderboardMethodResultRank
uci-bp MAE (mmHg)Cuffless blood pressure estimationHybrid features + personalized calibration7.85MAE (mmHg), lower is better#2 of 3

1 row is credited to this paper. Ranks apply within each leaderboard only, as reported by the source.

Tasks & signals

Tasks

Cuffless blood pressure estimationBlood pressurePulse arrival time

Signals

Keywords

Cuffless BPCalibrationSSLXGBoostPTT

Datasets used

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG, ECG); Form factor (clinical_finger). Method: Method (Gradient-boosted trees). Target: Blood pressure; Pulse arrival time.SetupSignals: PPG, ECGSignalsPPG, ECGForm factor: clinical_fingerForm factorclinical_fingerMethodMethodGradient-boosted treesTargetBlood pressureBlood pressurePulse arrival timePulse arrival time

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Zhang, Z., Qu, P., & Tang, H. (2026). Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals. Sensors, 26(15), 4676. https://doi.org/10.3390/s26154676
Zhang, Zichuan, Peng Qu, and Hong Tang. 2026. "Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals." Sensors 26 (15): 4676. https://doi.org/10.3390/s26154676.
Zhang, Z., Qu, P. and Tang, H. (2026) 'Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals', Sensors, 26(15), p. 4676. doi: 10.3390/s26154676.
Zhang, Zichuan, et al. "Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals." Sensors, vol. 26, no. 15, 2026, p. 4676. https://doi.org/10.3390/s26154676.
Z. Zhang, P. Qu, and H. Tang, "Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals," Sensors, vol. 26, no. 15, p. 4676, 2026, doi: 10.3390/s26154676.
@article{zhang2026personalized,
  title         = {{Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals}},
  author        = {Zhang, Zichuan and Qu, Peng and Tang, Hong},
  year          = {2026},
  journal       = {Sensors},
  volume        = {26},
  number        = {15},
  pages         = {4676},
  publisher     = {MDPI},
  doi           = {10.3390/s26154676}
}

Added to the ChatPPG research hub