Paperdoi:10.3390/s26154676
Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals
Sensors (MDPI) 26(15):4676
Key results
All results- Hybrid features + personalized calibrationAbstract (Results) (source, opens in a new tab)
7.85
MAE (mmHg) ↓
Cuffless blood pressure · BP-UCI
SBP · feature-based model with base-BP calibration
Results as reported by the authors; not independently reproduced. A SOTA star marks the best value among results reported under the same evaluation protocol; ties go to the earlier publication.
TL;DR
Handcrafted PTT/HRV/PPG-morphology features + masked-reconstruction SSL embeddings + subject prior BP (alpha-weighted base BP from the earliest calibration file) → XGBoost. MAE 7.25/3.98 (private set) and 7.85/4.02 mmHg (UCI). Small-sample feasibility per the authors.
- Form factor: clinical_finger
Abstract
Continuous and noninvasive blood pressure (BP) monitoring is important for hypertension screening and long-term cardiovascular management. Continuous BP estimation using photoplethysmography (PPG) and electrocardiography (ECG) has become a promising alternative to conventional cuff-based measurements. However, handcrafted feature-based methods are sensitive to fiducial point detection, while end-to-end deep models often require large labeled datasets and may ignore subject-specific BP baselines. This study proposes a personalized calibration and hybrid feature fusion framework for continuous BP estimation from ECG and PPG signals. The framework integrates physiologically interpretable handcrafted features, self-supervised waveform representations, and subject-specific prior BP information to predict systolic and diastolic BP. The handcrafted branch extracts pulse transit time, heart rate variability, and PPG morphological descriptors, whereas a masked reconstruction-based self-supervised encoder learns latent waveform embeddings without BP labels. Personalized calibration incorporates base BP from the earliest calibration file through an alpha-weighted strategy, and the fused representation is fed into XGBoost regressors. The framework was evaluated on a private dataset and the BP-UCI dataset, achieving SBP/DBP MAEs of 7.25/3.98 mmHg and 7.85/4.02 mmHg, respectively. Comparative and ablation results indicate improved baseline performance in the current experimental setting, suggesting the methodological feasibility of the proposed framework for small-sample continuous BP estimation. Further validation with larger and more diverse cohorts is still required.
Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://creativecommons.org/licenses/by/4.0/
Results on ChatPPG leaderboards · 1
All leaderboards| Leaderboard | Method | Result | Rank |
|---|---|---|---|
| uci-bp MAE (mmHg)Cuffless blood pressure estimation | Hybrid features + personalized calibration | 7.85MAE (mmHg), lower is better | #2 of 3 |
1 row is credited to this paper. Ranks apply within each leaderboard only, as reported by the source.
Tasks & signals
Tasks
Signals
Keywords
Datasets used
How it works
Cite this paper
Zhang, Z., Qu, P., & Tang, H. (2026). Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals. Sensors, 26(15), 4676. https://doi.org/10.3390/s26154676
Zhang, Zichuan, Peng Qu, and Hong Tang. 2026. "Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals." Sensors 26 (15): 4676. https://doi.org/10.3390/s26154676.
Zhang, Z., Qu, P. and Tang, H. (2026) 'Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals', Sensors, 26(15), p. 4676. doi: 10.3390/s26154676.
Zhang, Zichuan, et al. "Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals." Sensors, vol. 26, no. 15, 2026, p. 4676. https://doi.org/10.3390/s26154676.
Z. Zhang, P. Qu, and H. Tang, "Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals," Sensors, vol. 26, no. 15, p. 4676, 2026, doi: 10.3390/s26154676.
@article{zhang2026personalized,
title = {{Personalized Calibration and Hybrid Feature Fusion for Continuous Blood Pressure Estimation Using PPG and ECG Signals}},
author = {Zhang, Zichuan and Qu, Peng and Tang, Hong},
year = {2026},
journal = {Sensors},
volume = {26},
number = {15},
pages = {4676},
publisher = {MDPI},
doi = {10.3390/s26154676}
}Added to the ChatPPG research hub