Paperdoi:10.1109/TBCAS.2026.3736710
FiLMamba: A FiLM-Conditioned Mamba Architecture with an Energy-Efficient FPGA Accelerator for Generalizable Cuffless Blood Pressure Estimation
IEEE Transactions on Biomedical Circuits and Systems (TBCAS)
Key results
All results6.45
SD of error (mmHg) ↓
Cuffless blood pressure · MIMIC-BP (1,524 patients)
SBP · FiLM-conditioned Mamba, end-to-end
Results as reported by the authors; not independently reproduced. A SOTA star marks the best value among results reported under the same evaluation protocol; ties go to the earlier publication.
TL;DR
Addresses PPG morphology–BP ambiguity (similar waveforms → different pressures across subjects → mean regression) with a selective state-space (Mamba) model conditioned via Feature-wise Linear Modulation (FiLM) on a 4-D hemodynamic state vector, plus Cross-Domain Hemodynamic Alignment (CDHA) (unlabeled BN recalibration + per-calibration-unit bias correction). Evaluated on MIMIC-BP (1,524 patients): SBP/DBP SDE 6.45 / 4.96 mmHg, >2.5× better than unconditioned baselines; claims AAMI compliance. Training-free transfer to UCI-BP: 4.29 / 3.24 mmHg SDE — abstract claims first AAMI-compliant training-free cross-dataset cuffless BP result.
- Form factor: lab
Abstract
Cuffless blood pressure (BP) estimation from photoplethysmography (PPG) is limited by a morphology-BP ambiguity that similar waveforms map to different pressures across subjects, causing population-trained models to collapse into mean regression. We propose FiLMamba, a selective state-space model conditioned via Feature-wise Linear Modulation (FiLM) on a 4-D hemodynamic state vector, and Cross-Domain Hemodynamic Alignment (CDHA), which orthogonally decomposes residual cross-corpus shift into unlabeled batch-normalization recalibration and additive bias correction per calibration unit. On MIMIC-BP across 1,524 patients, FiLMamba achieves SBP/DBP standard deviation of error (SDE) of 6.45/4.96mmHg with >2.5× improvement over unconditioned baselines, satisfying Association for the Advancement of Medical Instrumentation (AAMI) compliance. Under training-free transfer, it further reaches 4.29/3.24mmHg on UCI-BP, the first AAMI-compliant training-free cross-dataset result in cuffless BP estimation. A dedicated FPGA accelerator specializes this stack on Zynq UltraScale+ through Heterogeneous Engine Reuse Orchestration (HERO) and Persistent Recurrent In-SRAM Mamba (PRISM), operating at 0.514W dynamic power and 150MHz with 8.23 μJ per sample, 15.3-41.1× below evaluated CPU/GPU platforms, and an energy-delay product of 0.046 J·s that is 1.87-14.8× below every general-purpose baseline.
Verbatim from Europe PMC (opens in a new tab) · Abstract via Europe PMC; © the authors/publisher
Results on ChatPPG leaderboards · 2
All leaderboards| Leaderboard | Method | Result | Rank |
|---|---|---|---|
| MIMIC-BP SDE (mmHg)Cuffless blood pressure estimation | FiLMamba | 6.45SDE (mmHg), lower is better | #1 of 1 |
| uci-bp SDE (mmHg)Cuffless blood pressure estimation | FiLMamba | 4.29SDE (mmHg), lower is better | #1 of 1 |
2 rows are credited to this paper. Ranks apply within each leaderboard only, as reported by the source.
Tasks & signals
Tasks
Signals
Keywords
Datasets used
How it works
Cite this paper
He, F., Wu, J., Li, W., Li, Y., & Zeng, X. (2026). FiLMamba: A FiLM-Conditioned Mamba Architecture with an Energy-Efficient FPGA Accelerator for Generalizable Cuffless Blood Pressure Estimation. IEEE Transactions on Biomedical Circuits and Systems. https://doi.org/10.1109/TBCAS.2026.3736710
He, Fan, Jinhong Wu, Wenxian Li, Yan Li, and Xiaoyang Zeng. 2026. "FiLMamba: A FiLM-Conditioned Mamba Architecture with an Energy-Efficient FPGA Accelerator for Generalizable Cuffless Blood Pressure Estimation." IEEE Transactions on Biomedical Circuits and Systems. https://doi.org/10.1109/TBCAS.2026.3736710.
He, F., Wu, J., Li, W., Li, Y. and Zeng, X. (2026) 'FiLMamba: A FiLM-Conditioned Mamba Architecture with an Energy-Efficient FPGA Accelerator for Generalizable Cuffless Blood Pressure Estimation', IEEE Transactions on Biomedical Circuits and Systems. doi: 10.1109/TBCAS.2026.3736710.
He, Fan, et al. "FiLMamba: A FiLM-Conditioned Mamba Architecture with an Energy-Efficient FPGA Accelerator for Generalizable Cuffless Blood Pressure Estimation." IEEE Transactions on Biomedical Circuits and Systems, 2026. https://doi.org/10.1109/TBCAS.2026.3736710.
F. He, J. Wu, W. Li, Y. Li, and X. Zeng, "FiLMamba: A FiLM-Conditioned Mamba Architecture with an Energy-Efficient FPGA Accelerator for Generalizable Cuffless Blood Pressure Estimation," IEEE Transactions on Biomedical Circuits and Systems, 2026, doi: 10.1109/TBCAS.2026.3736710.
@article{he2026filmamba,
title = {{FiLMamba: A FiLM-Conditioned Mamba Architecture with an Energy-Efficient FPGA Accelerator for Generalizable Cuffless Blood Pressure Estimation}},
author = {He, Fan and Wu, Jinhong and Li, Wenxian and Li, Yan and Zeng, Xiaoyang},
year = {2026},
journal = {IEEE Transactions on Biomedical Circuits and Systems},
doi = {10.1109/TBCAS.2026.3736710}
}Added to the ChatPPG research hub