Paperdoi:10.1088/1361-6579/ae56ae

Towards fair and trustworthy heart rate estimation from wrist-worn PPG (MW-PPG-HR)

Published articlePublished Apr 8, 2026Added Sep 27, 2026

  • DR
  • TC
  • PP

Authors: Daniel Ray, Tim Collins, Prasad V S Ponnapalli

Physiological Measurement

TL;DR

This study aimed to make wrist PPG heart rate estimates fairer across skin tones and sexes and to flag unreliable readings. The authors collected a multi-wavelength PPG dataset (blue, green, red, infrared) across activities and built a PPG-accelerometer network that estimates its own uncertainty. Error was higher for darker skin and for women at baseline, and rejecting high-uncertainty predictions removed the skin-tone gap (1.2 bpm to 0.0 bpm).

  • Result: Baseline error was higher for participants with darker skin tones and for females; uncertainty-guided rejection reduced MAE and effectively eliminated these gaps (reducing parity gaps from 1.2 BPM to 0.0 BPM for skin tone).
  • Result: Concrete dropout provided the best calibration-accuracy-efficiency trade-off, achieving the lowest average miscalibration area (MA = 0.12 ) while maintaining competitive MAE.

Abstract

Objective. To improve fairness (reduced disparities across skin tones and sexes) and trust (well-calibrated uncertainty metrics that indicate unreliable predictions) in wrist-worn photoplethysmography (PPG) heart rate estimation. While PPG offers a convenient, low-cost method for continuous heart rate monitoring, accuracy is often degraded by motion artefacts and demographic variation. Approach. We collected a multi-wavelength (blue, green, red, IR) wrist-worn PPG dataset covering diverse activities, skin tones, and sexes, and developed a multi-branch PPG-accelerometer fusion convolutional neural network quantifying both aleatoric and epistemic uncertainty (the latter via Monte Carlo dropout, concrete dropout, and deep ensembles). We compared wavelength sets, quantified demographic effects, and applied uncertainty-aware post-processing to filter out high-uncertainty predictions. Evaluation used leave-one-subject-out cross validation across both the collected multi-wavelength dataset and five commonly used single-wavelength datasets. Main results. Multi-wavelength fusion improved accuracy under motion, with blue-green-red-IR achieving the lowest mean absolute error (MAE). Baseline error was higher for participants with darker skin tones and for females; uncertainty-guided rejection reduced MAE and effectively eliminated these gaps (reducing parity gaps from 1.2 BPM to 0.0 BPM for skin tone). Concrete dropout provided the best calibration-accuracy-efficiency trade-off, achieving the lowest average miscalibration area (MA

= 0.12

) while maintaining competitive MAE. Significance. Combining calibrated uncertainty with group-aware evaluation yields a more fair and trustworthy wrist-worn PPG heart rate estimation pipeline, supporting responsible deployment in consumer and sports contexts and informing bias-aware clinical development.

Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://publishingsupport.iopscience.iop.org/iop-standard/v1

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

Heart rate

Signals

Keywords

HRFairnessUncertaintyMulti WavelengthMw PPG HR

Datasets used

How it works

How MW-PPG-HR uncertainty-aware CNN works: Setup → Method → ResultSetup: Signals (PPG, Accelerometer). Method: MW-PPG-HR uncertainty-aware CNN. Result: Result (Baseline error was higher for participants with darker skin tones and for females; uncertainty-guided rejection reduced MAE and effectively eliminated these gaps (reducing parity gaps from 1.2 BPM to 0.0 BPM for skin tone).); Result (Concrete dropout provided the best calibration-accuracy-efficiency trade-off, achieving the lowest average miscalibration area (MA = 0.12 ) while maintaining competitive MAE.).SetupSignals: PPG, AccelerometerSignalsPPG, AccelerometerMethodMW-PPG-HR uncertainty-aware CNNResultResult: Baseline error was higher for participants with darker skin tones and for females; uncertainty-guided rejection reduced MAE and effectively eliminated these gaps (reducing parity gaps from 1.2 BPM to 0.0 BPM for skin tone).ResultBaseline error was higher for…Result: Concrete dropout provided the best calibration-accuracy-efficiency trade-off, achieving the lowest average miscalibration area (MA = 0.12 ) while maintaining competitive MAE.ResultConcrete dropout provided the best…

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Ray, D., Collins, T., & Ponnapalli, P. V. S. (2026). Towards fair and trustworthy heart rate estimation from wrist-worn PPG (MW-PPG-HR). Physiological Measurement. https://doi.org/10.1088/1361-6579/ae56ae
Ray, Daniel, Tim Collins, and Prasad V S Ponnapalli. 2026. "Towards fair and trustworthy heart rate estimation from wrist-worn PPG (MW-PPG-HR)." Physiological Measurement. https://doi.org/10.1088/1361-6579/ae56ae.
Ray, D., Collins, T. and Ponnapalli, P. V. S. (2026) 'Towards fair and trustworthy heart rate estimation from wrist-worn PPG (MW-PPG-HR)', Physiological Measurement. doi: 10.1088/1361-6579/ae56ae.
Ray, Daniel, et al. "Towards fair and trustworthy heart rate estimation from wrist-worn PPG (MW-PPG-HR)." Physiological Measurement, 2026. https://doi.org/10.1088/1361-6579/ae56ae.
D. Ray, T. Collins, and P. V. S. Ponnapalli, "Towards fair and trustworthy heart rate estimation from wrist-worn PPG (MW-PPG-HR)," Physiological Measurement, 2026, doi: 10.1088/1361-6579/ae56ae.
@article{ray2026towards,
  title         = {{Towards fair and trustworthy heart rate estimation from wrist-worn PPG (MW-PPG-HR)}},
  author        = {Ray, Daniel and Collins, Tim and Ponnapalli, Prasad V. S.},
  year          = {2026},
  journal       = {Physiological Measurement},
  doi           = {10.1088/1361-6579/ae56ae}
}