Device research record
Apple Watch Series 6
Signals and sensors
From the device record. Fields the source does not state are left out.
- Maker
- Apple
- Form factor
- Wrist
- Series
- Series 6
- Family
- Apple Watch
- Signals recorded
- PPG, Accelerometer, ECG
- Also called
- Apple Watch S6, Watch Series 6
What published studies show
Results quoted from 9 published validation studies, each against a clinical reference. A result applies to the model and protocol tested, not to every model the brand sells. From the wearable accuracy database (v1.0.0).
Heart rate at rest1 study
| Device tested | Result | Reference | Sample | Study |
|---|---|---|---|---|
| Apple Watch Series 6R-R intervals and BPM at rest | 1.15%Mean absolute percentage error | ECG | n = 78, Lab: rest, talking, movie, before/after walking | Bonneval et al. (2025)Independent |
- Apple Watch Series 61.15%
Mean absolute percentage error, R-R intervals and BPM at rest · vs ECG · n = 78, Lab: rest, talking, movie, before/after walking
Bonneval et al. (2025) Independent
Heart rate during exercise2 studies
| Device tested | Result | Reference | Sample | Study |
|---|---|---|---|---|
| Apple Watch Series 6 | High agreement in squat, deadlift, cable row (rho > 0.832)Study finding (no single number) | ECG chest strap (Polar H7/H10) | n = 29, Resistance exercise (5 exercises) | Støve et al. (2022)Funding not reported |
| Apple Watch Series 6 | < 5% in all 5 activitiesCoefficient of variation | ECG chest strap (Polar H7/H10) | n = 60, Sitting, walking, running, resistance, cycling | Hajj-Boutros et al. (2023)Funding not reported |
- Apple Watch Series 6High agreement in squat, deadlift, cable row (rho > 0.832)
Study finding (no single number) · vs ECG chest strap (Polar H7/H10) · n = 29, Resistance exercise (5 exercises)
Støve et al. (2022) Funding not reported
- Apple Watch Series 6< 5% in all 5 activities
Coefficient of variation · vs ECG chest strap (Polar H7/H10) · n = 60, Sitting, walking, running, resistance, cycling
Hajj-Boutros et al. (2023) Funding not reported
Overnight heart rate1 study
| Device tested | Result | Reference | Sample | Study |
|---|---|---|---|---|
| Apple Watch Series 6 | ICC 0.96Intraclass correlation | ECG | n = 53, Overnight, sleep laboratory | Miller et al. (2022)Other maker funded |
| Apple Watch Series 6 | 1.5 bpm (absolute bias)Mean absolute error | ECG | n = 53, Overnight, sleep laboratory | Miller et al. (2022)Other maker funded |
- Apple Watch Series 6ICC 0.96
Intraclass correlation · vs ECG · n = 53, Overnight, sleep laboratory
Miller et al. (2022) Other maker funded
- Apple Watch Series 61.5 bpm (absolute bias)
Mean absolute error · vs ECG · n = 53, Overnight, sleep laboratory
Miller et al. (2022) Other maker funded
HRV (RMSSD)1 study
| Device tested | Result | Reference | Sample | Study |
|---|---|---|---|---|
| Apple Watch Series 6 | ICC 0.67Intraclass correlation | ECG | n = 53, Overnight, sleep laboratory | Miller et al. (2022)Other maker funded |
| Apple Watch Series 6 | 22.5 ms (absolute bias)Mean absolute error (absolute bias) | ECG | n = 53, Overnight, sleep laboratory | Miller et al. (2022)Other maker funded |
- Apple Watch Series 6ICC 0.67
Intraclass correlation · vs ECG · n = 53, Overnight, sleep laboratory
Miller et al. (2022) Other maker funded
- Apple Watch Series 622.5 ms (absolute bias)
Mean absolute error (absolute bias) · vs ECG · n = 53, Overnight, sleep laboratory
Miller et al. (2022) Other maker funded
Sleep stages (4-stage)1 study
| Device tested | Result | Reference | Sample | Study |
|---|---|---|---|---|
| Apple Watch Series 6 | 53%Epoch-by-epoch agreement | Polysomnography (PSG) | n = 53, Sleep laboratory, 1 night, 9 h in bed, all devices worn together | Miller et al. (2022)Other maker funded |
| Apple Watch Series 6 | κ = 0.20Cohen's kappa | Polysomnography (PSG) | n = 53, Sleep laboratory, 1 night, 9 h in bed, all devices worn together | Miller et al. (2022)Other maker funded |
Epoch-by-epoch agreement · vs Polysomnography (PSG) · n = 53, Sleep laboratory, 1 night, 9 h in bed, all devices worn together
Miller et al. (2022) Other maker funded
- Apple Watch Series 6κ = 0.20
Cohen's kappa · vs Polysomnography (PSG) · n = 53, Sleep laboratory, 1 night, 9 h in bed, all devices worn together
Miller et al. (2022) Other maker funded
Sleep vs wake (2-stage)1 study
| Device tested | Result | Reference | Sample | Study |
|---|---|---|---|---|
| Apple Watch Series 6 | 88%Epoch-by-epoch agreement | Polysomnography (PSG) | n = 53, Sleep laboratory, 1 night, 9 h in bed, all devices worn together | Miller et al. (2022)Other maker funded |
| Apple Watch Series 6 | κ = 0.30Cohen's kappa | Polysomnography (PSG) | n = 53, Sleep laboratory, 1 night, 9 h in bed, all devices worn together | Miller et al. (2022)Other maker funded |
Epoch-by-epoch agreement · vs Polysomnography (PSG) · n = 53, Sleep laboratory, 1 night, 9 h in bed, all devices worn together
Miller et al. (2022) Other maker funded
- Apple Watch Series 6κ = 0.30
Cohen's kappa · vs Polysomnography (PSG) · n = 53, Sleep laboratory, 1 night, 9 h in bed, all devices worn together
Miller et al. (2022) Other maker funded
Blood oxygen (SpO2)3 studies
| Device tested | Result | Reference | Sample | Study |
|---|---|---|---|---|
| Apple Watch Series 6 | r = 0.89Pearson correlation | Medical-grade pulse oximeter | n = 265, Spot readings | Spaccarotella et al. (2022)Independent |
| Apple Watch Series 6 | −0.23 (LoA −3.49 to 3.04)Mean bias (signed) | Medical-grade pulse oximeter | n = 265, Spot readings | Spaccarotella et al. (2022)Independent |
| Apple Watch Series 6 | Not yet up to the medical standard of pulse oximetryStudy finding (no single number) | Medical-grade pulse oximeter | n = 508, Congenital heart disease clinic | Pätz et al. (2023)Funding not reported |
| Apple Watch Series 6 | 0.46% vs ABG (LoA −5.91 to 5.00)Mean bias (signed) | Arterial blood gas (SaO2) | n = 167, Emergency department, COPD exacerbation | Arslan et al. (2024)Funding not reported |
- Apple Watch Series 6r = 0.89
Pearson correlation · vs Medical-grade pulse oximeter · n = 265, Spot readings
Spaccarotella et al. (2022) Independent
- Apple Watch Series 6−0.23 (LoA −3.49 to 3.04)
Mean bias (signed) · vs Medical-grade pulse oximeter · n = 265, Spot readings
Spaccarotella et al. (2022) Independent
- Apple Watch Series 6Not yet up to the medical standard of pulse oximetry
Study finding (no single number) · vs Medical-grade pulse oximeter · n = 508, Congenital heart disease clinic
Pätz et al. (2023) Funding not reported
- Apple Watch Series 60.46% vs ABG (LoA −5.91 to 5.00)
Mean bias (signed) · vs Arterial blood gas (SaO2) · n = 167, Emergency department, COPD exacerbation
Arslan et al. (2024) Funding not reported
Atrial fibrillation detection1 study
| Device tested | Result | Reference | Sample | Study |
|---|---|---|---|---|
| Apple Watch Series 6smartwatch ECG app, inconclusive tracings counted | 85% (spec 75%)Sensitivity | 12-lead ECG | n = 201, Tertiary cardiology service, single-lead ECG algorithms | Mannhart et al. (2023)Funding not reported |
| Apple Watch Series 6smartwatch ECG app, inconclusive tracings counted | 75%Specificity | 12-lead ECG | n = 201, Tertiary cardiology service, single-lead ECG algorithms | Mannhart et al. (2023)Funding not reported |
- Apple Watch Series 685% (spec 75%)
Sensitivity, smartwatch ECG app, inconclusive tracings counted · vs 12-lead ECG · n = 201, Tertiary cardiology service, single-lead ECG algorithms
Mannhart et al. (2023) Funding not reported
Specificity, smartwatch ECG app, inconclusive tracings counted · vs 12-lead ECG · n = 201, Tertiary cardiology service, single-lead ECG algorithms
Mannhart et al. (2023) Funding not reported
Energy expenditure (calories)2 studies
| Device tested | Result | Reference | Sample | Study |
|---|---|---|---|---|
| Apple Watch Series 6 | 14.68–24.85% (poor in all activities)Coefficient of variation | Indirect calorimetry | n = 60, Sitting, walking, running, resistance, cycling | Hajj-Boutros et al. (2023)Funding not reported |
| Apple Watch Series 6walking | 19.8% walk / 24.4% runMean absolute percentage error | Indirect calorimetry | n = 20, Outdoor 2-km walk (6 km/h) and run (10 km/h) | Le et al. (2022)Independent |
- Apple Watch Series 614.68–24.85% (poor in all activities)
Coefficient of variation · vs Indirect calorimetry · n = 60, Sitting, walking, running, resistance, cycling
Hajj-Boutros et al. (2023) Funding not reported
- Apple Watch Series 619.8% walk / 24.4% run
Mean absolute percentage error, walking · vs Indirect calorimetry · n = 20, Outdoor 2-km walk (6 km/h) and run (10 km/h)
Le et al. (2022) Independent
Comparisons
Head-to-head pages that name Apple Watch Series 6 come first, then other comparisons for the brand.
- Whoop vs Apple Watch: What the Sleep-Lab Studies FoundMentions Apple Watch Series 6
- Garmin vs Apple Watch: PPG, ECG & Battery ComparedMentions Apple Watch Series 6
- Oura Ring vs Apple Watch: Finger PPG vs Wrist PPG AccuracyMentions Apple Watch Series 6
- Apple Watch Ultra 2 vs Garmin Fenix 8: Specs Guide
- Fitbit vs Apple Watch: PPG Sensor Accuracy and Health Features
- Samsung Galaxy Watch vs Apple Watch: PPG Sensor and Health Features
- Google Pixel Watch vs Apple Watch: PPG Sensor and Health Ecosystem
- Withings vs Apple Watch: ScanWatch 2 ECG, SpO2 and PPG Compared