Device comparisons

Fitbit vs Apple Watch: Heart Rate and Sleep Accuracy from Published Studies

Independent studies favor Apple Watch for exercise heart rate (concordance with ECG 0.92 vs 0.67 for Fitbit Blaze in Gillinov 2017). For sleep staging, Fitbit Sense (κ 0.42) and Charge 5 (κ 0.41) came close to Apple Watch Series 8 (κ 0.53) in Schyvens 2025. Both brands ran large AFib studies, but with different definitions, so their headline numbers are not directly comparable.

Published study results

MetricFitbit (Sense 2 / Charge 6)Apple Watch Series 10
Exercise heart rateConcordance vs ECGGillinov et al. 2017, n=500.67Fitbit Blaze0.92Apple WatchCloser in this study
Sleep stagesCohen's κ vs polysomnographySchyvens et al. 2025, n=620.42 / 0.41Sense / Charge 50.53Series 8Closer in this study

Each row is one statistic from a study already cited on this page. The link names the sample size when this page already states it. This is not a new measurement.

Devices

ChatPPG comparison of Fitbit (Sense 2 / Charge 6) and Apple Watch Series 10. Product photographs, not a measurement.

Fitbit Charge 6 with a coral band: product photograph from the Google Store. Apple Watch Series 10, jet black aluminum with a black sport band: product photograph from Apple Newsroom, September 2024.

Quick answer

Apple Watch has the stronger evidence. During exercise its heart rate reached 0.92 concordance with ECG versus 0.67 for the Fitbit Blaze. For sleep stages against polysomnography, Apple Watch Series 8 scored κ 0.53 versus 0.42 for Fitbit Sense and 0.41 for Charge 5. Both tested older models.

Sources: Gillinov et al. 2017, Med Sci Sports Exerc; Schyvens et al. 2025, Sleep Advances

At a glance

Gillinov et al. 2017 measured exercise heart rate in 50 adults. Schyvens et al. 2025 measured sleep stages in 62 adults. The Fitbit devices were Blaze, Sense and Charge 5; the Apple Watch sleep result is Series 8.

SpecificationFitbit (Sense 2 / Charge 6)Apple Watch Series 10
Exercise heart rateConcordance vs ECG0.67Fitbit Blaze0.92Apple WatchCloser in this study
Sleep stagesCohen's κ vs polysomnography0.42 / 0.41Sense / Charge 50.53Series 8Closer in this study

Primary rows only. The study tables further down list more results from the accuracy database.

Gillinov et al. 2017, n=50; Schyvens et al. 2025, n=62

Hardware specs

SpecificationFitbit (Sense 2 / Charge 6)Apple Watch Series 10
Battery LifeDiffers6+ days (Sense 2); up to 7 days (Charge 6)Up to 18 hours; 36 hours Low Power Mode
Water ResistanceDiffers50 m50 m (ISO 22810:2010)
WeightDiffers37.64 g (Sense 2); 30 g (Charge 6)36.4 g (46 mm aluminum, GPS)
ECGDiffersECG app (not in all countries)ECG app; irregular rhythm notifications
SpO2DiffersRed and infrared sensors (Sense 2)Blood Oxygen app (not for medical use)

Hardware only. Study results are not repeated in this table. Differs marks values that are not the same. An em dash means this page did not publish that figure.

Specs verified on October 17, 2026: Gillinov et al. 2017, Med Sci Sports Exerc (exercise HR vs ECG, n=50), Schyvens et al. 2025, Sleep Advances (six devices vs PSG, n=62), Perez et al. 2019, NEJM (Apple Heart Study), Lubitz et al. 2022, Circulation (Fitbit Heart Study), Google Store: Fitbit Sense 2 tech specs, Google Store: Fitbit Charge 6 tech specs, Apple Watch Series 10 technical specifications, FDA De Novo DEN180042: Apple irregular rhythm notification.

Pros and cons

Fitbit (Sense 2 / Charge 6)

Pros

  • Battery 6+ days (Sense 2), up to 7 days (Charge 6)
  • ECG app; Sense 2 adds cEDA skin-conductance sensing
  • Fitbit Heart Study: 98.2% PPV for concurrent AF (Lubitz 2022)

Cons

  • Fitbit Blaze exercise HR concordance 0.67 vs ECG (Gillinov 2017)
  • Sleep-stage κ 0.42/0.41 (Sense/Charge 5) vs 0.53 for Apple Watch S8 (Schyvens 2025)

Apple Watch Series 10

Pros

  • Exercise HR concordance 0.92 with ECG (Gillinov 2017)
  • Highest sleep-stage κ (0.53, Series 8) of six devices in Schyvens 2025
  • ECG app and FDA-authorized irregular rhythm notifications

Cons

  • Up to 18 hours battery (36 hours Low Power Mode)
  • Requires an iPhone

What published studies show

One table per metric. These rows come from the accuracy database and are not repeated as a second card list. From the wearable accuracy database (v1.0.0).

Heart rate at rest4 studies

Device testedResultReferenceSampleStudy
Fitbit Charge 47.8 bpm (seated rest)Mean absolute errorHolter ECGn = 23, Rest, sedentary, low- and high-intensity phasesNissen et al. (2022)Independent
Fitbit Charge 4ICC 0.80 (COPD) / 0.79 (controls)Intraclass correlationECG chest strap (Polar H7/H10)n = 55, 5 days free-livingHermans et al. (2024)Funding not reported
Apple Watch Series 9 / Ultra 25.91% (MAE 3.73 bpm)Mean absolute percentage errorPolar H10 + Kubios HRVn = 39, Serial morning measurements over 14 days, 316 readingsO'Grady et al. (2024)Independent
Apple Watch Series 6R-R intervals and BPM at rest1.15%Mean absolute percentage errorECGn = 78, Lab: rest, talking, movie, before/after walkingBonneval et al. (2025)Independent

Heart rate accuracy across devices →

Heart rate during exercise11 studies

Device testedResultReferenceSampleStudy
Fitbit Charge 6mixed rest/stress/walking4.5 bpmMean absolute errorECG chest strap (Zephyr BioHarness / Faros)n = 45, Rest, cognitive stress, steady and intermittent walking in neutral/hot/cold chambersGielen et al. (2026)Funding not reported
Fitbit Inspire 3mixed rest/stress/walking9–14 bpm (group range)Mean absolute errorECG chest strap (Zephyr BioHarness / Faros)n = 45, Rest, cognitive stress, steady and intermittent walking in neutral/hot/cold chambersGielen et al. (2026)Funding not reported
Fitbit Charge 4squats15.7 bpm (squats)Mean absolute errorHolter ECGn = 23, Rest, sedentary, low- and high-intensity phasesNissen et al. (2022)Independent
Fitbit Versa4.1%–19.1% (range by intensity)Coefficient of variationECG chest strap (Polar H7/H10)n = 25, Sitting, walking, running, sprintsDüking et al. (2020)Independent
More results (2)
Device testedResultReferenceSampleStudy
Fitbit Sense4.14–10.76%Coefficient of variationECG chest strap (Polar H7/H10)n = 60, Sitting, walking, running, resistance, cyclingHajj-Boutros et al. (2023)Funding not reported
Fitbit Blazeall exercise conditionsrc = 0.67Lin's concordance correlationECGn = 50, Gillinov: treadmill, bike, ellipticalGillinov et al. (2017)Funding not reported

9 more results in the database. Heart rate accuracy across devices →

Overnight heart rate1 study

Device testedResultReferenceSampleStudy
Apple Watch Series 6ICC 0.96Intraclass correlationECGn = 53, Overnight, sleep laboratoryMiller et al. (2022)Other maker funded
Apple Watch Series 61.5 bpm (absolute bias)Mean absolute errorECGn = 53, Overnight, sleep laboratoryMiller et al. (2022)Other maker funded

Heart rate accuracy across devices →

HRV (RMSSD)2 studies

Device testedResultReferenceSampleStudy
Apple Watch Series 9 / Ultra 228.88% (MAE 20.46 ms)Mean absolute percentage errorPolar H10 + Kubios HRVn = 39, Serial morning measurements over 14 days, 316 readingsO'Grady et al. (2024)Independent
Apple Watch Series 6ICC 0.67Intraclass correlationECGn = 53, Overnight, sleep laboratoryMiller et al. (2022)Other maker funded
Apple Watch Series 622.5 ms (absolute bias)Mean absolute error (absolute bias)ECGn = 53, Overnight, sleep laboratoryMiller et al. (2022)Other maker funded

HRV accuracy across devices →

Sleep stages (4-stage)4 studies

Device testedResultReferenceSampleStudy
Fitbit Senseκ = 0.42Cohen's kappaPolysomnography (PSG)n = 37, Sleep laboratory, 1 nightSchyvens et al. (2025)Funding not reported
Fitbit Charge 5κ = 0.41Cohen's kappaPolysomnography (PSG)n = 39, Sleep laboratory, 1 nightSchyvens et al. (2025)Funding not reported
Fitbit Sense 261.7–78.0% (range across stages)SensitivityPolysomnography (PSG)n = 35, Inpatient, 1 nightRobbins et al. (2024)Other maker funded
Fitbit Sense 2deep sleep durationDeep sleep −15 minMean bias (signed)Polysomnography (PSG)n = 35, Inpatient, 1 nightRobbins et al. (2024)Other maker funded
More results (2)
Device testedResultReferenceSampleStudy
Fitbit Sense 2κ = 0.42Cohen's kappaPolysomnography (PSG)n = 26, Two Korean sleep centres, in-lab PSGLee et al. (2023)Independent
Fitbit Sense 264.6%Epoch-by-epoch agreementPolysomnography (PSG)n = 26, Two Korean sleep centres, in-lab PSGLee et al. (2023)Independent

7 more results in the database. Sleep accuracy across devices →

Sleep vs wake (2-stage)4 studies

Device testedResultReferenceSampleStudy
Fitbit Charge 5> 90% (all devices)SensitivityPolysomnography (PSG)n = 62, Sleep laboratory, 1 nightSchyvens et al. (2025)Funding not reported
Fitbit Sense> 90% (all devices)SensitivityPolysomnography (PSG)n = 62, Sleep laboratory, 1 nightSchyvens et al. (2025)Funding not reported
Fitbit Sense 2≥ 95%SensitivityPolysomnography (PSG)n = 35, Inpatient, 1 nightRobbins et al. (2024)Other maker funded
Fitbit Alta HRAs good as or better than actigraphyStudy finding (no single number)Polysomnography (PSG)n = 34, Sleep laboratory, 3 nights incl. disrupted nightChinoy et al. (2021)Funding not reported
More results (2)
Device testedResultReferenceSampleStudy
Apple Watch Series 8> 90% (all devices)SensitivityPolysomnography (PSG)n = 62, Sleep laboratory, 1 nightSchyvens et al. (2025)Funding not reported
Apple Watch Series 8≥ 95%SensitivityPolysomnography (PSG)n = 35, Inpatient, 1 nightRobbins et al. (2024)Other maker funded

2 more results in the database. Sleep accuracy across devices →

Total sleep time1 study

Device testedResultReferenceSampleStudy
Fitbit Sense 2−74.5 min (older adults)Mean bias (signed)Polysomnography (PSG)n = 19, 1 nightSearles et al. (2026)Funding not reported

Sleep accuracy across devices →

Blood oxygen (SpO2)9 studies

Device testedResultReferenceSampleStudy
Fitbit Charge 4ICC 0.32 (COPD, overnight)Intraclass correlationMedical-grade pulse oximetern = 30, 4 nightsHermans et al. (2024)Funding not reported
Apple Watch Series 73.5% (rest)Root-mean-square errorMedical-grade pulse oximetern = 36, Rest, COPD outpatientsStøve et al. (2023)Funding not reported
Apple Watch Series 6r = 0.89Pearson correlationMedical-grade pulse oximetern = 265, Spot readingsSpaccarotella et al. (2022)Independent
Apple Watch Series 6−0.23 (LoA −3.49 to 3.04)Mean bias (signed)Medical-grade pulse oximetern = 265, Spot readingsSpaccarotella et al. (2022)Independent
More results (2)
Device testedResultReferenceSampleStudy
Apple Watch Series 72.2% (RMSE 2.9%)Mean absolute errorMedical-grade pulse oximetern = 49, Spot readings, with and without chronic pulmonary diseaseJiang et al. (2023)Independent
Apple Watch Series 72.9%Root-mean-square errorMedical-grade pulse oximetern = 49, Spot readings, with and without chronic pulmonary diseaseJiang et al. (2023)Independent

5 more results in the database. SpO2 accuracy across devices →

Atrial fibrillation detection6 studies

Device testedResultReferenceSampleStudy
Fitbit Sensesmartwatch ECG app, inconclusive tracings counted66% (spec 79%)Sensitivity12-lead ECGn = 201, Tertiary cardiology service, single-lead ECG algorithmsMannhart et al. (2023)Funding not reported
Fitbit Sensesmartwatch ECG app, inconclusive tracings counted79%Specificity12-lead ECGn = 201, Tertiary cardiology service, single-lead ECG algorithmsMannhart et al. (2023)Funding not reported
Fitbit (PPG irregular rhythm algorithm)PPG irregular rhythm detectionPPV 98.2% (95% CI 95.5–99.5%)Positive predictive valueECG patchn = 455699, Remote trial; 1-week ECG patch after detectionLubitz et al. (2022)Maker funded
Fitbit Ionic + FibriCheck algorithmthird-party PPG algorithm (FibriCheck), not Fitbit's ownSens 97% / spec 100%Sensitivity12-lead ECGn = 73, Around cardioversion, 1-min PPGSelder et al. (2023)Independent
More results (2)
Device testedResultReferenceSampleStudy
Apple Watch Series 6smartwatch ECG app, inconclusive tracings counted85% (spec 75%)Sensitivity12-lead ECGn = 201, Tertiary cardiology service, single-lead ECG algorithmsMannhart et al. (2023)Funding not reported
Apple Watch Series 6smartwatch ECG app, inconclusive tracings counted75%Specificity12-lead ECGn = 201, Tertiary cardiology service, single-lead ECG algorithmsMannhart et al. (2023)Funding not reported

3 more results in the database. AFib accuracy across devices →

VO2max estimate3 studies

Device testedResultReferenceSampleStudy
Apple Watch Series 1013.2% (MAE 6.79 mL/kg/min)Mean absolute percentage errorIndirect calorimetryn = 40, Maximal treadmill test (modified Åstrand)Lambe et al. (2026)Independent
Apple Watch13.31% (MAE 6.92 mL/kg/min)Mean absolute percentage errorIndirect calorimetryn = 30, Maximal treadmill test (modified Åstrand)Lambe et al. (2025)Independent
Apple Watch Series 715.79% (ICC 0.47)Mean absolute percentage errorIndirect calorimetryn = 19, Graded cycle ergometer testCaserman et al. (2024)Independent

VO2max accuracy across devices →

Energy expenditure (calories)6 studies

Device testedResultReferenceSampleStudy
Fitbit Versa8.0%–32.3% (range by intensity)Coefficient of variationIndirect calorimetryn = 25, Sitting, walking, running, sprintsDüking et al. (2020)Independent
Fitbit Sense13.44–29.66% (poor in all activities)Coefficient of variationIndirect calorimetryn = 60, Sitting, walking, running, resistance, cyclingHajj-Boutros et al. (2023)Funding not reported
Fitbit Sense 2physical activity energy expenditure, 10 min+3.14 kcalMean bias (signed)Indirect calorimetryn = 44, 10-min recumbent cycling intervalsKostrna et al. (2026)Independent
Fitbit SurgeNo device achieved EE error below 20%Study finding (no single number)Indirect calorimetryn = 60, Sitting, walking, running, cyclingShcherbina et al. (2017)Independent
More results (2)
Device testedResultReferenceSampleStudy
Apple Watch Series 9Inconsistent across all groupsStudy finding (no single number)Indirect calorimetryn = 30, 10-min treadmillChase et al. (2024)Independent
Apple Watch Series 413.5%–27.1% (range by intensity)Coefficient of variationIndirect calorimetryn = 25, Sitting, walking, running, sprintsDüking et al. (2020)Independent

4 more results in the database. Calories accuracy across devices →

Respiratory rate1 study

Device testedResultReferenceSampleStudy
Fitbit Charge 4ICC 0.84 (COPD) / 0.77 (controls)Intraclass correlationECG chest strap (Polar H7/H10)n = 55, 5 days free-livingHermans et al. (2024)Funding not reported

Verdict

Apple Watch has the stronger independent evidence for exercise heart rate and was slightly ahead for sleep staging; Fitbit offers multi-day battery and works with Android and iPhone. Both have published large-scale AFib studies.

Frequently asked questions

Is Fitbit or Apple Watch more accurate for heart rate?

In Gillinov et al. (2017), Apple Watch reached 0.92 concordance with ECG during exercise versus 0.67 for the Fitbit Blaze. Newer models have not been compared head to head.

Can Fitbit and Apple Watch detect AFib?

Both have irregular rhythm features backed by large studies. The Apple Heart Study reported a PPV of 0.84 (Perez 2019) and the Fitbit Heart Study 98.2% (Lubitz 2022), but the studies used different definitions, so the numbers should not be ranked.

Which tracks sleep better: Fitbit or Apple Watch?

In Schyvens et al. (2025), sleep-stage agreement with polysomnography was κ 0.53 for Apple Watch Series 8, 0.42 for Fitbit Sense and 0.41 for Fitbit Charge 5, all fair to moderate.

Device research records

Datasets, papers and validation results for the models named on this page.

Other comparisons

View all comparisons