Paperdoi:10.1016/j.ahjo.2026.100837

Arterial blood pressure waveform reconstruction estimation from PPG using TCN–BiLSTM

Published articlePublished Aug 2026Added Oct 11, 2026

  • HL
  • AB
  • NM
  • HC

Authors: Hicham Loumissi, Adil Barra, Najat Messaoudi, Hicham Medromi (University of Hassan II Casablanca)

American Heart Journal Plus: Cardiology Research and Practice

Key results

All results

Results as reported by the authors; not independently reproduced. A SOTA star marks the best value among results reported under the same evaluation protocol; ties go to the earlier publication.

TL;DR

TCN + BiLSTM on derivative-enhanced PPG reconstructs ABP. VitalDB, 2,226 patients: MAE/RMSE SBP 3.61/5.24, DBP 1.79/2.37 mmHg. Tested on a MIMIC-IV cohort without retraining (external numbers not in abstract).

  • Form factor: clinical_finger

Abstract

Photoplethysmography PPG is a non-invasive and effective method for continuously measuring blood pressure BP without the use of a cuff. However, its accuracy can be affected by noise, motion artifacts, and inter-individual variability. We implement a deep learning framework that utilizes the PPG signal to accurately reconstruct arterial blood pressure ABP waves, enabling reliable estimation of systolic and diastolic blood pressure SBP/DBP values from the reconstructed ABP waves. To capture temporal and morphological cardiovascular dynamics, the TCN-BiLSTM model combines temporal convolutional networks and bidirectional long short-term memory networks within a framework of derivative-enhanced PPG representations. The Mean Absolute Error and Root Mean Square Error MAE/RMSE values of the proposed method were 3.61/5.24 mmHg for SBP and 1.79/2.37 mmHg for DBP when using 2226 patients in VitalDB database. The completed model was tested on an independent the MIMIC-IV database cohort as an additional external assessment, without retraining. These findings are consistent with those of previous studies, suggesting that using PPG to reconstruct the ABP waveform provides an accurate and physiologically valid method for estimating arterial pressure. This study is feasible for continuous blood pressure monitoring without the use of a cuff.

Verbatim from Europe PMC (opens in a new tab) · Licence: cc by-nc-nd

Results on ChatPPG leaderboards · 1

All leaderboards
LeaderboardMethodResultRank
VitalDB MAE (mmHg)Cuffless blood pressure estimationTCN–BiLSTM PPG→ABP3.61MAE (mmHg), lower is better#1 of 2

1 row is credited to this paper. Ranks apply within each leaderboard only, as reported by the source.

Tasks & signals

Signals

Keywords

Cuffless BPABP WaveformTCNBiLSTMVitalDBExternal Validation

Datasets used

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG); Form factor (clinical_finger). Method: Method. Target: Blood pressure.SetupSignals: PPGSignalsPPGForm factor: clinical_fingerForm factorclinical_fingerMethodMethodTargetBlood pressureBlood pressure

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Loumissi, H., Barra, A., Messaoudi, N., & Medromi, H. (2026). Arterial blood pressure waveform reconstruction estimation from PPG using TCN–BiLSTM. American Heart Journal Plus: Cardiology Research and Practice. https://doi.org/10.1016/j.ahjo.2026.100837
Loumissi, Hicham, Adil Barra, Najat Messaoudi, and Hicham Medromi. 2026. "Arterial blood pressure waveform reconstruction estimation from PPG using TCN–BiLSTM." American Heart Journal Plus: Cardiology Research and Practice. https://doi.org/10.1016/j.ahjo.2026.100837.
Loumissi, H., Barra, A., Messaoudi, N. and Medromi, H. (2026) 'Arterial blood pressure waveform reconstruction estimation from PPG using TCN–BiLSTM', American Heart Journal Plus: Cardiology Research and Practice. doi: 10.1016/j.ahjo.2026.100837.
Loumissi, Hicham, et al. "Arterial blood pressure waveform reconstruction estimation from PPG using TCN–BiLSTM." American Heart Journal Plus: Cardiology Research and Practice, 2026. https://doi.org/10.1016/j.ahjo.2026.100837.
H. Loumissi, A. Barra, N. Messaoudi, and H. Medromi, "Arterial blood pressure waveform reconstruction estimation from PPG using TCN–BiLSTM," American Heart Journal Plus: Cardiology Research and Practice, 2026, doi: 10.1016/j.ahjo.2026.100837.
@article{loumissi2026arterial,
  title         = {{Arterial blood pressure waveform reconstruction estimation from PPG using TCN–BiLSTM}},
  author        = {Loumissi, Hicham and Barra, Adil and Messaoudi, Najat and Medromi, Hicham},
  year          = {2026},
  journal       = {American Heart Journal Plus: Cardiology Research and Practice},
  doi           = {10.1016/j.ahjo.2026.100837}
}

Added to the ChatPPG research hub