Paperdoi:10.1007/s44200-026-00120-3

A Dual-Output Physiology-Informed Neural Network Architecture for Continuous Cuffless Blood Pressure Waveform Estimation: A Proof-of-Concept

Venue listed2026Added Oct 11, 2026

  • ZA
  • CS

Authors: Zaineb Aloui, Cederick Landry (Université de Sherbrooke)

Artery Research

TL;DR

Subject-specific NARX networks on ECG + PPG; the physiology-informed variant adds a personalized 4-parameter BP–PPG sigmoidal layer and a dual-output loss (BP and PPG waveforms). n = 5, >6 h of daily-living activities. With 15 min training it matches the 30-min baseline.

  • Form factor: other

Abstract

This pilot study introduces the first subject‑specific, physiology‑informed neural network approach for accurate estimation of the full blood pressure (BP) waveform. The method builds on a nonlinear autoregressive model with exogenous inputs (NARX), implemented through artificial neural networks trained on subject‑specific electrocardiography (ECG) and photoplethysmography (PPG) signals. Three model configurations are evaluated: (1) a reference NARX model trained on 30 min of data, (2) a reduced‑data NARX model trained on 15 min, and (3) a physiology‑informed model (NARX physio ), also trained on 15 min, which incorporates a personalized four‑parameter BP–PPG sigmoidal layer into the training process through a dual‑output loss function that simultaneously estimates the BP and PPG waveforms. By embedding BP–PPG physiology directly into training, the architecture ensures internal physiological consistency—something not achieved in existing PPG‑based or purely data‑driven machine learning approaches. All three models were assessed on a small ( n  = 5), but rich dataset, covering over six hours of activities of daily living. Despite using only half the training duration of the baseline NARX model, the physiology‑informed approach maintains accurate BP estimation. Beyond improved accuracy, the sigmoidal layer provides physiologically interpretable parameters, including estimated arterial‑compliance curve width with an average of 36.5 mmHg and an estimated PPG contact pressure with an average of 59 mmHg in this study, offering insights unavailable in standard neural network models. Together, these preliminary results suggest that the NARX physio model enables effective personalization of cuffless BP waveform estimation with limited training data, highlighting its potential to improve both the reliability and the clinical relevance of continuous cuffless BP monitoring.

Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://creativecommons.org/licenses/by/4.0

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

Blood pressure

Signals

Keywords

cuffless-BPABP-waveformphysics-informedNARXpersonalizedpilot

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG, ECG); Form factor (other). Method: Method. Target: Blood pressure.SetupSignals: PPG, ECGSignalsPPG, ECGForm factor: otherForm factorotherMethodMethodTargetBlood pressureBlood pressure

Cite this paper

APA

Aloui, Z., & Landry, C. (2026). A Dual-Output Physiology-Informed Neural Network Architecture for Continuous Cuffless Blood Pressure Waveform Estimation: A Proof-of-Concept. Artery Research. https://doi.org/10.1007/s44200-026-00120-3

BibTeX

@article{aloui2026dualoutput,
  title         = {{A Dual-Output Physiology-Informed Neural Network Architecture for Continuous Cuffless Blood Pressure Waveform Estimation: A Proof-of-Concept}},
  author        = {Aloui, Zaineb and Landry, Cederick},
  year          = {2026},
  journal       = {Artery Research},
  doi           = {10.1007/s44200-026-00120-3}
}

Built from the hub record. Check it against the publisher before submitting.

Added to the ChatPPG research hub