Paperdoi:10.1038/s44328-026-00120-y

Sensor-skin contact pressure as a hidden confounder in wearable photoplethysmography biosensing

Venue listedDated Nov 15, 2026Added Nov 17, 2026

  • ME
  • AE
  • AA
  • SA
  • SA
  • NA
  • JA
  • RF

Authors: Mohamed Elgendi, Azza Elkhalifa, Anood Alali, Sara Alameri, Shahad Alzaabi, Najoud Alhammadi, John Allen, Richard Ribon Fletcher

npj Biosensing (review), published 2026-11-15

TL;DR

Review of 29 studies on how sensor–skin contact pressure changes PPG signals and cuffless BP estimates: too little pressure lowers optical stability and raises motion artefacts, too much compresses superficial vessels, flattens systolic peaks and removes the dicrotic notch. The best pressure depends on site, sensor and outcome; better waveform quality did not by itself mean lower BP error. Pressure is rarely measured or reported, and the authors propose pressure-aware reporting.

  • Form factor: mixed (wrist, finger and other pulse sites across the 29 reviewed studies)

Abstract

Photoplethysmography (PPG) is increasingly used for cuffless blood pressure (BP) monitoring and wearable cardiovascular sensing. However, the reliability of PPG-derived digital biomarkers is highly sensitive to sensor–skin contact pressure, which influences optical coupling, vascular compression, and motion susceptibility. Despite its importance, contact pressure is often poorly controlled or insufficiently reported, introducing a hidden source of variability that can compromise signal integrity, downstream model performance, and the reproducibility of wearable BP systems. Here, we systematically review experimental evidence on the influence of contact pressure in PPG-based cuffless BP estimation. Across 29 studies of PPG acquisition, BP estimation, and related pulse sensing, those that examined contact loading reported the following qualitative patterns: insufficient pressure reduced optical stability and increased motion artefacts, whereas excessive pressure compressed superficial vasculature, attenuated systolic peaks, and suppressed the dicrotic notch. Within studies that varied contact loading, favourable conditions depended on the anatomical site, sensor, and outcome; better waveform quality did not by itself establish lower BP-estimation error or a universal pressure target. Low, moderate, and high loading conditions are therefore used only as study-defined descriptions of signal-level effects. Because reported performance metrics represented different statistical constructs, outcomes were summarized separately for signal quality, waveform morphology, timing features, and BP-estimation error; the evidence does not support quantitative ranking of classical signal-processing, statistical, and machine-learning approaches. We further identify substantial methodological inconsistency in pressure measurement, control, and reporting, and propose standardized pressure-aware practices to improve reproducibility, benchmarking, device design, and the clinical translation of cuffless BP monitoring.

Verbatim from OpenAlex (opens in a new tab) · Abstract via OpenAlex (publisher-deposited); © the authors/publisher

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

Blood pressureSignal quality

Signals

Keywords

Contact PressureHold Down PressureStrap TensionCuffless BPWaveform MorphologyDicrotic NotchSignal QualityReporting Standards

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG); Form factor (mixed (wrist, finger and other pulse sites across the 29 reviewed studies)). Method: Method. Target: Blood pressure; Signal quality.SetupSignals: PPGSignalsPPGForm factor: mixed (wrist, finger and other pulse sites across the 29 reviewed studies)Form factormixed (wrist, finger and other pulse…MethodMethodTargetBlood pressureBlood pressureSignal qualitySignal quality

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Elgendi, M., Elkhalifa, A., Alali, A., Alameri, S., Alzaabi, S., Alhammadi, N., Allen, J., & Fletcher, R. R. (2026). Sensor-skin contact pressure as a hidden confounder in wearable photoplethysmography biosensing. npj Biosensing. https://doi.org/10.1038/s44328-026-00120-y
Elgendi, Mohamed, Azza Elkhalifa, Anood Alali, Sara Alameri, Shahad Alzaabi, Najoud Alhammadi, John Allen, and Richard Ribon Fletcher. 2026. "Sensor-skin contact pressure as a hidden confounder in wearable photoplethysmography biosensing." npj Biosensing. https://doi.org/10.1038/s44328-026-00120-y.
Elgendi, M., Elkhalifa, A., Alali, A., Alameri, S., Alzaabi, S., Alhammadi, N., Allen, J. and Fletcher, R. R. (2026) 'Sensor-skin contact pressure as a hidden confounder in wearable photoplethysmography biosensing', npj Biosensing. doi: 10.1038/s44328-026-00120-y.
Elgendi, Mohamed, et al. "Sensor-skin contact pressure as a hidden confounder in wearable photoplethysmography biosensing." npj Biosensing, 2026. https://doi.org/10.1038/s44328-026-00120-y.
M. Elgendi et al., "Sensor-skin contact pressure as a hidden confounder in wearable photoplethysmography biosensing," npj Biosensing, 2026, doi: 10.1038/s44328-026-00120-y.
@article{elgendi2026sensorskin,
  title         = {{Sensor-skin contact pressure as a hidden confounder in wearable photoplethysmography biosensing}},
  author        = {Elgendi, Mohamed and Elkhalifa, Azza and Alali, Anood and Alameri, Sara and Alzaabi, Shahad and Alhammadi, Najoud and Allen, John and Fletcher, Richard Ribon},
  year          = {2026},
  journal       = {npj Biosensing},
  doi           = {10.1038/s44328-026-00120-y}
}

Added to the ChatPPG research hub