Paperdoi:10.1097/HJH.0000000000004445
Continual cuffless blood pressure monitoring in children: agreement and feasibility compared with ambulatory blood pressure monitoring
Journal of Hypertension (2026-09-11)
TL;DR
Prospective 24-h simultaneous wrist cuffless optical (Aktiia) vs ABPM in 21 children with well-controlled type 1 diabetes (mean age 14.7 y). Mean differences (Aktiia − ABPM): −4.5 mmHg 24-h SBP, −0.8 mmHg 24-h DBP; TOST 90% CIs within an exploratory ±10 mmHg margin for 24-h and night SBP and for DBP across all periods, not for daytime SBP. Bland–Altman: modest bias but wide limits of agreement, especially SBP.
- Form factor: wrist (Aktiia cuffless optical bracelet)
Abstract
Background:
Cuffless wearable technologies enable continual blood pressure (BP) monitoring and may improve tolerability compared with conventional cuff-based methods. Pediatric data comparing cuffless devices with ambulatory blood pressure monitoring (ABPM) remain limited.
Methods:
In this prospective observational study, children with well controlled type 1 diabetes mellitus and normal office BP underwent simultaneous 24-h monitoring using a wrist-worn cuffless optical device and conventional ABPM. Between-method differences were assessed for averaged 24-h, daytime, and nighttime systolic and diastolic BP. Equivalence was evaluated using the two one-sided tests procedure with a prespecified exploratory ± 10 mmHg margin, whereas individual-level agreement was assessed using Bland–Altman analysis.
Results:
Twenty-one participants were included (mean age 14.7 ± 2.6 years). Mean differences (Aktiia − ABPM) were −4.5 mmHg for 24-h systolic BP and −0.8 mmHg for 24-h diastolic BP. The 90% confidence intervals were fully contained within the ±10 mmHg margin for 24-h and nighttime systolic BP and for diastolic BP across all periods, but not for daytime systolic BP. Bland–Altman analysis showed modest mean bias but wide limits of agreement, particularly for systolic BP. Nocturnal systolic and diastolic BP dipping was attenuated with cuffless monitoring.
Conclusions:
In this selected pediatric cohort, continual cuffless BP monitoring was feasible and showed group-level comparability with ABPM for averaged 24-h BP values. However, it should not be considered interchangeable with ABPM for individual clinical decisions or assessment of nocturnal dipping. Larger studies in more diverse pediatric populations are required before broader clinical implementation.
Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: http://creativecommons.org/licenses/by/4.0
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Tasks
Signals
Keywords
How it works
Cite this paper
Huml, M., Kotková, A., Fremuthová, T., Fiklík, K., Pomahačová, R., Kormunda, S., Sýkora, J., & Fremuth, J. (2026). Continual cuffless blood pressure monitoring in children: agreement and feasibility compared with ambulatory blood pressure monitoring. Journal of Hypertension. https://doi.org/10.1097/HJH.0000000000004445
Huml, M., A. Kotková, Tereza Fremuthová, K. Fiklík, R. Pomahačová, S. Kormunda, J. Sýkora, and J. Fremuth. 2026. "Continual cuffless blood pressure monitoring in children: agreement and feasibility compared with ambulatory blood pressure monitoring." Journal of Hypertension. https://doi.org/10.1097/HJH.0000000000004445.
Huml, M., Kotková, A., Fremuthová, T., Fiklík, K., Pomahačová, R., Kormunda, S., Sýkora, J. and Fremuth, J. (2026) 'Continual cuffless blood pressure monitoring in children: agreement and feasibility compared with ambulatory blood pressure monitoring', Journal of Hypertension. doi: 10.1097/HJH.0000000000004445.
Huml, M., et al. "Continual cuffless blood pressure monitoring in children: agreement and feasibility compared with ambulatory blood pressure monitoring." Journal of Hypertension, 2026. https://doi.org/10.1097/HJH.0000000000004445.
M. Huml et al., "Continual cuffless blood pressure monitoring in children: agreement and feasibility compared with ambulatory blood pressure monitoring," Journal of Hypertension, 2026, doi: 10.1097/HJH.0000000000004445.
@article{huml2026continual,
title = {{Continual cuffless blood pressure monitoring in children: agreement and feasibility compared with ambulatory blood pressure monitoring}},
author = {Huml, M. and Kotková, A. and Fremuthová, Tereza and Fiklík, K. and Pomahačová, R. and Kormunda, S. and Sýkora, J. and Fremuth, J.},
year = {2026},
journal = {Journal of Hypertension},
doi = {10.1097/HJH.0000000000004445}
}Added to the ChatPPG research hub