Photoplethysmography based remote cardiac blood pressure monitoring compared with 24-h ambulatory blood pressure monitoring

Venue listed2026Added Oct 21, 2026

  • MV
  • JA
  • CH
  • AH
  • DH
  • SM
  • JS
  • ER

Authors: Mariska van Vliet, Jan J.J. Aalberts, Cora Hamelinck, Arnaud D. Hauer, Dieke Hoftijzer, Stefan H.J. Monnink, Jurjan C. Schipper, Eelko Ronner

Blood Pressure Monitoring 31(4):185–192 (2026; epub 2026-06-24)

Licence: CC BY-NC-ND 4.0

TL;DR

Prospective single-centre study, 40 adults undergoing standard 24-h ABPM (30-min interval) while wearing the CardioWatch the whole period. Compared average difference and SD over 24 h and separately for awake and asleep periods, plus the awake–asleep BP change per night-time dipper category; accuracy under motion visualised.

  • Form factor: wrist

Abstract

Objective

To evaluate the accuracy of a novel wrist-worn cuff-calibrated photoplethysmography (PPG)-based remote patient monitoring device as an alternative to ambulatory blood pressure monitoring (ABPM).

Methods

This prospective single-center study included 40 adult patients undergoing standard 24 h ABPM with a 30-min measurement interval. Participants were equipped with a wrist-worn PPG device (CardioWatch 287-2; Corsano Health B.V., The Hague, The Netherlands, manufactured by MMT, Geneva, Switzerland) throughout the ABPM period. Investigational and reference measurements were compared using the average difference and SD for the full 24-h period as well as for awake and asleep periods, individually. Differences in awake–asleep blood pressure (BP) change between the investigational and reference devices were evaluated for all patients and per night-time BP dipper category. Finally, investigational BP monitoring accuracy under motion was visualized.

Results

Throughout the 24-h period, the absolute difference in average SBP and DBP was 3.65 (SD ± 4.57) and 2.98 (SD ± 3.91), respectively. Similar results were obtained for the subsets of day- and night-time determinations. The absolute difference in average awake–asleep BP change was 2.36 (SD ± 2.40) for SBP and 2.17 (SD ± 2.13) for DBP. Comparable absolute differences were present within each of the night-time dipper categories. The performance of the investigational device remained stable under motion.

Conclusion

This study presents initial evidence supporting the accuracy of the evaluated wrist-worn PPG-based device in measuring 24-h as well as day- and night-time BP. In all cases, the average difference and SD remained below the cutoff values described by recent European Society of Hypertension guidelines. Therefore, the presented method can provide an accurate alternative to ABPM.

Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://creativecommons.org/licenses/by-nc-nd/4.0/

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

Blood pressure

Signals

Keywords

cuffless-BPABPM24hdippingnocturnal-BPCorsanowrist-PPGdevice-validationESHbp-trend

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG); Form factor (wrist). Method: Method. Target: Blood pressure.SetupSignals: PPGSignalsPPGForm factor: wristForm factorwristMethodMethodTargetBlood pressureBlood pressure

Cite this paper

APA

van Vliet, M., Aalberts, J. J., Hamelinck, C., Hauer, A. D., Hoftijzer, D., Monnink, S. H., Schipper, J. C., & Ronner, E. (2026). Photoplethysmography based remote cardiac blood pressure monitoring compared with 24-h ambulatory blood pressure monitoring. Blood Pressure Monitoring. https://doi.org/10.1097/MBP.0000000000000804

BibTeX

@article{vliet2026photoplethysmography,
  title         = {{Photoplethysmography based remote cardiac blood pressure monitoring compared with 24-h ambulatory blood pressure monitoring}},
  author        = {van Vliet, Mariska and Aalberts, Jan J.J. and Hamelinck, Cora and Hauer, Arnaud D. and Hoftijzer, Dieke and Monnink, Stefan H.J. and Schipper, Jurjan C. and Ronner, Eelko},
  year          = {2026},
  journal       = {Blood Pressure Monitoring},
  doi           = {10.1097/MBP.0000000000000804}
}

Built from the hub record. Check it against the publisher before submitting.

Added to the ChatPPG research hub