Paper
Hilo (formerly Aktiia) / CSEM oBPM — cuffless wrist BP from green PPG pulse-wave analysis
Scientific Reports 2021 / Blood Pressure Monitoring 2021 / J Hypertens 2023 (independent) / FDA 510(k) 2025
Josep Sola et al., "Validation of the optical Aktiia bracelet in different body positions for the persistent monitoring of blood pressure", doi:10.1038/s41598-021-99294-w, CC BY 4.0
TL;DR
Hilo, formerly Aktiia, makes a wrist band that estimates blood pressure from optical PPG using pulse-wave analysis. In a Scientific Reports validation with 91 patients over four weeks, after calibration on day one, mean systolic differences from auscultation ranged from -3.02 to 0.46 mmHg across sitting, lying, lap and standing positions. The authors found it accurate across common positions, though precision varied between them.
- Protocol: extended ISO 81060-2 vs auscultation
- Result: SBP difference 0.46 ± 7.75 mmHg sitting
Abstract
The diagnosis of hypertension and the adjustment of antihypertensive drugs are evolving from isolated measurements performed at the physician offices to the full phenotyping of patients in real-life conditions. Indeed, the strongest predictor of cardiovascular risk comes from night measurements. The aim of this study was to demonstrate that a wearable device (the Aktiia Bracelet) can accurately estimate BP in the most common body positions of daily life and thus become a candidate solution for the BP phenotyping of patients. We recruited 91 patients with BP ranging from low to hypertensive levels and compared BP values from the Aktiia Bracelet against auscultatory reference values for 4 weeks according to an extended ISO 81060-2 protocol. After initializing on day one, the observed means and standard deviations of differences for systolic BP were of 0.46 ± 7.75 mmHg in the sitting position, − 2.44 ± 10.15 mmHg in the lying, − 3.02 ± 6.10 mmHg in the sitting with the device on the lap, and − 0.62 ± 12.51 mmHg in the standing position. Differences for diastolic BP readings were respectively of 0.39 ± 6.86 mmHg, − 1.93 ± 7.65 mmHg, − 4.22 ± 6.56 mmHg and − 4.85 ± 9.11 mmHg. This study demonstrates that a wearable device can accurately estimate BP in the most common body positions compared to auscultation, although precision varies across positions. While wearable persistent BP monitors have the potential to facilitate the identification of individual BP phenotypes at scale, their prognostic value for cardiovascular events and its association with target organ damage will need cross-sectional and longitudinal studies. Deploying this technology at a community level may be also useful to drive public health interventions against the epidemy of hypertension.
Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://creativecommons.org/licenses/by/4.0
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Tasks
Signals
Keywords
How it works
- Cohort
- 91 patients, 4 weeks
Cite this paper
Hilo (formerly Aktiia) / CSEM oBPM — cuffless wrist BP from green PPG pulse-wave analysis. (2021). Scientific Reports 2021 / Blood Pressure Monitoring 2021 / J Hypertens 2023 / FDA 510 2025. https://doi.org/10.1038/s41598-021-99294-w
@article{hilocsemobpm2021hilo,
title = {{Hilo (formerly Aktiia) / CSEM oBPM — cuffless wrist BP from green PPG pulse-wave analysis}},
year = {2021},
journal = {Scientific Reports 2021 / Blood Pressure Monitoring 2021 / J Hypertens 2023 / FDA 510 2025},
url = {https://doi.org/10.1038/s41598-021-99294-w}
}