Paperdoi:10.64898/2026.11.18.767743

Noninvasive Cardiac Arrest Recognition Using Wearable Sensors in a Porcine Model of Ventricular Fibrillation

Preprint2026Added Nov 20, 2026

  • ZE
  • MS
  • AR
  • IA
  • AB
  • SP
  • RG
  • JS
  • +1

Authors: Zachary Englhardt, Moaz Bin Saeed, Armin Razi, Iyad Ayoub, Alvin Baetiong, Shwetak Patel, Raúl J. Gazmuri, Jacob E. Sunshine, Vikram Iyer

bioRxiv (bioengineering) v1, posted 2026-11-20; preprint, not peer reviewed

TL;DR

VF electrically induced in 9 anaesthetised, supine pigs; custom wearable-class MAX30101 green (537 nm) PPG at ~36 Hz on ear and hoof, plus limb clinical PPG, ECG, and catheter pressures (RA, LV, PA, aorta). Band-pass filtered PPG (and the clinical pulse-ox trace) just goes flat after VF. The raw DC intensity instead shows a reproducible exponential decay: pooled τ = 10.53 s, between-animal CV 8.9 % at 5 s, 17.0 % at 10 s, new steady state after ~1 min. Standard band-pass + automatic gain control hides it. Mechanism: arterial→venous blood redistribution.

  • Form factor: ear + hoof (porcine, wearable-class MAX30101); human FP cohort = wrist Fossil Sport WearOS watch (MOODS field study)

Abstract

Background: Wearable detection of out-of-hospital cardiac arrest (OHCA) could improve survival, given the prognostic importance of time to resuscitation and that 50% to 75% of events are unwitnessed. Existing systems may be limited by their form factor and by specificity achieved at the cost of sensitivity. We investigated whether the raw photoplethysmography (PPG) signal from low-cost wearable sensors contains a distinct signature of circulatory arrest. Methods: Ventricular fibrillation (VF) was electrically induced in 9 invasively monitored pigs while raw PPG was recorded with a wearable-class sensor. PPG dynamics after VF onset were related to aortic and right atrial pressures. A heuristic detection algorithm was evaluated for sensitivity and latency by leave-one-out cross-validation in the porcine data and for false positives in 57,857 hours of free-living PPG from 115 human participants. Results: After VF onset, the raw PPG showed a reproducible exponential decay (time constant 10.53 s; coefficient of variation 8.9% to 17.0% over the first 10 s) that tracked rising right atrial pressure (r = 0.90), reflecting arterial-to-venous blood redistribution. The algorithm detected VF in 8 of 9 animals (sensitivity 0.889; 95% CI, 0.518-0.997) with a median latency of 57.2 s. In the human cohort, it produced 2.73 false positives per user-year (95% CI, 1.62-4.31), comparable to current systems. Conclusions: Raw wearable PPG encodes circulatory collapse during VF, as supported by invasive hemodynamic measurements. While preliminary, these results suggest the potential for wearable-based OHCA detection with improved sensitivity, reduced latency, and comparable specificity.

Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: http://creativecommons.org/licenses/by-nd/4.0/

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

AF / rhythm

Signals

Keywords

Cardiac ArrestLoss of PulseVentricular FibrillationOHCARaw PPGDC ComponentGreen PPGMAX30101

Datasets used

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG); Wavelengths (green); Form factor (ear + hoof (porcine, wearable-class MAX30101); human FP cohort = wrist Fossil Sport WearOS watch (MOODS field study)). Method: Method. Target: AF / rhythm.SetupSignals: PPGSignalsPPGWavelengths: greenWavelengthsgreenForm factor: ear + hoof (porcine, wearable-class MAX30101); human FP cohort = wrist Fossil Sport WearOS watch (MOODS field study)Form factorear + hoof (porcine, wearable-class…MethodMethodTargetAF / rhythmAF / rhythm
Wavelengths
green

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Englhardt, Z., Bin Saeed, M., Razi, A., Ayoub, I., Baetiong, A., Patel, S., Gazmuri, R. J., Sunshine, J. E., & Iyer, V. (2026). Noninvasive Cardiac Arrest Recognition Using Wearable Sensors in a Porcine Model of Ventricular Fibrillation. bioRxiv v1, posted. https://doi.org/10.64898/2026.11.18.767743
Englhardt, Zachary, Moaz Bin Saeed, Armin Razi, Iyad Ayoub, Alvin Baetiong, Shwetak Patel, Raúl J. Gazmuri, Jacob E. Sunshine, and Vikram Iyer. 2026. "Noninvasive Cardiac Arrest Recognition Using Wearable Sensors in a Porcine Model of Ventricular Fibrillation." bioRxiv v1, posted. https://doi.org/10.64898/2026.11.18.767743.
Englhardt, Z., Bin Saeed, M., Razi, A., Ayoub, I., Baetiong, A., Patel, S., Gazmuri, R. J., Sunshine, J. E. and Iyer, V. (2026) 'Noninvasive Cardiac Arrest Recognition Using Wearable Sensors in a Porcine Model of Ventricular Fibrillation', bioRxiv v1, posted. doi: 10.64898/2026.11.18.767743.
Englhardt, Zachary, et al. "Noninvasive Cardiac Arrest Recognition Using Wearable Sensors in a Porcine Model of Ventricular Fibrillation." bioRxiv v1, posted, 2026. https://doi.org/10.64898/2026.11.18.767743.
Z. Englhardt et al., "Noninvasive Cardiac Arrest Recognition Using Wearable Sensors in a Porcine Model of Ventricular Fibrillation," bioRxiv v1, posted, 2026, doi: 10.64898/2026.11.18.767743.
@misc{englhardt2026noninvasive,
  title         = {{Noninvasive Cardiac Arrest Recognition Using Wearable Sensors in a Porcine Model of Ventricular Fibrillation}},
  author        = {Englhardt, Zachary and Bin Saeed, Moaz and Razi, Armin and Ayoub, Iyad and Baetiong, Alvin and Patel, Shwetak and Gazmuri, Raúl J. and Sunshine, Jacob E. and Iyer, Vikram},
  year          = {2026},
  howpublished  = {bioRxiv v1, posted},
  doi           = {10.64898/2026.11.18.767743}
}

Added to the ChatPPG research hub