SleepImage Ring: How the CPC Sleep Study Ring Works
The SleepImage ring is a finger PPG and oximetry recorder for an FDA-cleared CPC sleep study (K182618). How it works, how to wear it, what it reports, and the evidence.
- Sleep & Sleep Apnea
Cardiopulmonary coupling: stable vs unstable sleep
Illustrative, simulated data · not patient dataNight B: same stages, unstable breathing (sleep-apnea-like)
Illustrative CPC sleep spectrogram rendered by ChatPPG from simulated data. Not patient data and not a SleepImage report.
Short answer: The SleepImage ring is a finger pulse oximeter used for a home sleep study. It records your pulse waveform, oxygen level and movement overnight, and SleepImage's cloud software analyzes the night with cardiopulmonary coupling (CPC), a method that reads sleep stability from how heart rate and breathing move together. The software is FDA cleared (K182618) for prescription use in children and adults, and with oximetry it reports an apnea-hypopnea index for a clinician to review.
This page explains how the system works and what the published evidence says. It is not a review, and it is not medical advice. ChatPPG has no relationship with SleepImage.
What the SleepImage ring is
SleepImage is the brand of MyCardio LLC, based in Denver. Its patient resources page lists three recorders that work with the app: the SleepImage Ring PO2 (large), the SleepImage Ring PO4 (small) and the SleepImage Fingertip PO6, a clip-style sensor (SleepImage).
The ring's own manual describes it as a non-invasive pulse oximeter for adults that measures oxygen saturation and pulse rate for spot checks and continuous data collection, not continuous monitoring. It uses red and infrared light, and the manual lists SpO2 accuracy of ±2% from 80 to 100% saturation, pulse rate accuracy of ±2 bpm or 2%, a battery life of 12 to 14 hours, and an IP22 rating (ring manual).
The ring is only the sensor. The diagnosis-support part is the software.
Is the SleepImage ring FDA cleared?
The SleepImage System software has three FDA 510(k) clearances:
Scroll horizontally if needed
The K182618 indications say the software "analyzes, displays and summarizes Electrocardiogram (ECG) or Plethysmogram (PLETH) data" to aid evaluation of sleep disorders, that the sAHI aids diagnosis and management of sleep-disordered breathing, and that output should not be acted on "without consultation of a qualified healthcare professional". It is marked prescription use (FDA).
We did not find or verify a separate FDA number for the ring hardware. If that matters to you, ask the provider who issues the device.
How a SleepImage sleep study works
SleepImage's patient quick start guide sets out the steps (guide):
- Get a Patient ID. Your provider issues the recorder, and you receive a Patient ID by email, text or in person.
- Install the SleepImage app from the App Store or Google Play. The patient page says phones need iOS 16 or Android 12 or newer.
- Charge the ring. The quick start guide says a full charge takes about 90 minutes; the ring manual says 2 to 3 hours. Charge it fully either way.
- Pair it in the app, not in your phone's Bluetooth settings, by matching the last four digits of the serial number.
- Start the study in the app at bedtime and keep the phone close, for example on a bedside table.
- Stop the study in the app when you wake, then take the ring off. The recording uploads to your clinician.
How to wear the SleepImage ring
- Wear the ring on the thumb or index finger of the left hand, per the quick start guide. The fingertip sensor goes on the left index finger, with nail polish removed.
- Fit matters. The ring manual says that if the ring is too tight for the thumb or index finger, try another finger but avoid the middle finger, and that a loose fit can lead to poor data.
- Avoid strong light and a lot of movement during the recording.
- If you get up in the night, you can take the ring off briefly, but do not stop the study in the app. Put it back on and the recording resumes.
- The manual says not to use the device if you have a cardiac pacemaker, implanted defibrillator or other implanted metallic or electronic device, and to check the finger every 6 to 8 hours.
What the ring measures, and what CPC does with it
The ring records one optical signal, the photoplethysmogram (PPG), plus SpO2 and movement. SleepImage's clinical guide says its software derives pulse rate, pulse rate variability, peripheral arterial tone, a breathing signal (tidal volume variability), oxygen saturation and actigraphy from that, and analyzes them with CPC (clinical guide). If PPG is new to you, start with what PPG is and how PPG estimates breathing rate.
CPC was published in 2005 by Thomas and colleagues at Beth Israel Deaconess Medical Center (doi:10.1093/sleep/28.9.1151). It compares beat-to-beat intervals with the breathing signal in 8.5-minute windows and asks at what frequency the two move together:
- High-frequency coupling (HFC) at the breathing rate marks stable NREM sleep.
- Low-frequency coupling (LFC) in slow 10- to 100-second cycles marks unstable NREM sleep, the pattern sleep apnea and fragmented sleep produce.
- Very-low-frequency coupling (VLFC) marks REM sleep or wake.
1. Beat-to-beat (NN) intervals, ms
2. Respiration surrogate (EDR or PDR)
First 3 minutes of an 8.5-minute window, resampled at 2 Hz.
3. Coherent cross-power (coherence × cross-spectrum)
One 8.5-minute CPC window, simulated: beat-to-beat intervals (top left), the breathing signal derived from the same sensor (bottom left), and their coherent cross-power (right). In stable sleep the shared power sits at the breathing rate, so the window scores as HFC.
The full method, with more interactive figures, is in our cardiopulmonary coupling guide.
What a SleepImage report includes
According to SleepImage's clinical documentation, a report includes total sleep time and sleep efficiency, stable and unstable NREM and REM time, a sleep quality index (SQI), and, with oximetry, the SleepImage apnea-hypopnea index (sAHI) split into central and obstructive events, plus oxygen desaturation measures. The clinician can view raw data and adjust scoring.
Two things are different from a lab polysomnogram:
- No EEG. The ring cannot see brain waves, so its "sleep stages" are CPC states, not the AASM W, N1, N2, N3 and REM stages. SleepImage's guide says stable NREM roughly corresponds to part of N2 plus N3, and unstable NREM to N1 plus part of N2.
- Hypopneas are scored from oxygen. With no airflow sensor, breathing events are inferred from the pulse signal and SpO2, so the hypopnea rule behind the AHI matters when comparing reports.
How accurate is it? The evidence
Scroll horizontally if needed
| Study | Who | Signal | Result vs PSG AHI | Authors |
|---|---|---|---|---|
| Lu et al. 2023 | 39 adults, Charité Berlin | SleepImage ring | r = 0.89; AUC 1.00 / 0.90 / 0.98 at AHI ≥5 / 15 / 30 | Academic sleep center, no vendor authors |
| Hilmisson et al. 2020 | Children in the CHAT trial | PPG + SpO2 | r = 0.954 | Company-affiliated |
| Al Ashry et al. 2021 | 833 adults, APPLES trial | ECG + SpO2 | r = 0.972 | Includes company and inventor authors |
Sources: Lu 2023, Hilmisson 2020, Al Ashry 2021.
The agreement numbers are strong, but read them with two caveats. The only independent ring study is small, and most of the larger datasets were analyzed by people with a financial stake in CPC; the inventors disclose royalties and the SleepImage authors are company staff. Correlation also says less than it seems: a device can track AHI closely across a wide range and still misplace people near the 5, 15 and 30 cut-offs.
At scale, SleepImage authors analyzed one million nights of home recordings. They found the CPC sleep quality index declined with age much less steeply than EEG deep sleep usually does, and was higher in women in every age group (doi:10.1016/j.sleep.2025.108728).
SleepImage ring vs a consumer sleep ring
People searching "sleep study ring" often have a consumer ring in mind. They do different jobs:
- Purpose. The SleepImage ring is a recorder for prescription diagnostic software. Consumer rings are wellness trackers.
- Output. SleepImage reports CPC sleep stability and an sAHI for a clinician. Consumer rings estimate light, deep and REM sleep and readiness-style scores; see how accurate PPG sleep staging is.
- Use. SleepImage studies are usually one to a few nights, ordered by a provider. Consumer rings are worn every night.
For a broader look at apnea detection from optical sensors, see PPG sleep apnea detection. For every company we found that uses CPC, including ECG-based systems from Largan and DynaDx, see the CPC vendor directory.
Where SleepImage shows up under other names
Vivos Therapeutics offers SleepImage-based home testing to dental practices as VivoScore, under a 2021 agreement with MyCardio LLC disclosed in an SEC filing. Earlier, MyCardio and Embla built a single-lead ECG recorder, the M1, for CPC analysis (K092003).
The bottom line
The SleepImage ring records a finger PPG and oxygen; the cleared software applies cardiopulmonary coupling to report how stable your sleep was and, with oximetry, how often your breathing was disrupted. It does not measure brain waves, it is prescription only, and its results go to a clinician. If you want to understand the method behind the report, read the cardiopulmonary coupling guide.
Frequently Asked Questions
- What is the SleepImage ring?
- It is a finger-worn pulse oximeter that records the pulse waveform (pleth), oxygen saturation and movement overnight and sends them over Bluetooth to the SleepImage app. The SleepImage System software then analyzes the night with cardiopulmonary coupling (CPC) and produces a sleep study report for a clinician. SleepImage lists two ring sizes (PO2 large, PO4 small) and a fingertip recorder (PO6).
- Is the SleepImage ring FDA cleared?
- The SleepImage System software is FDA cleared under K182618 (2019) to analyze ECG or pleth data for children, adolescents and adults and, when oximetry is present, to report an apnea-hypopnea index (sAHI) that aids diagnosis of sleep-disordered breathing. It is prescription use. We did not verify a separate FDA number for the ring hardware itself.
- How do you wear the SleepImage ring?
- SleepImage's patient quick start guide says to wear the ring on the thumb or index finger of the left hand. The ring's manual says fit matters: if it is too tight for the thumb or index finger, try another finger but avoid the middle finger, and a loose fit can spoil the data. Charge it fully first, start the study in the app at bedtime, keep the phone nearby, and stop the study in the app before taking the ring off in the morning.
- What if I take the SleepImage ring off during the night?
- Both the quick start guide and the ring manual say you can remove it briefly, for example to use the bathroom, but should not stop the recording in the app. Put the ring back on and recording resumes.
- What does a SleepImage sleep study measure?
- From the pulse waveform it derives heart rate, pulse rate variability, a breathing signal and pulse-amplitude changes; from the oximeter it gets SpO2; from the accelerometer it gets movement. CPC analysis turns those into stable NREM, unstable NREM and REM/wake time, a sleep quality index, and, with oximetry, apnea-hypopnea and oxygen desaturation indices.
- How accurate is the SleepImage ring for sleep apnea?
- In an independent study at Charité Berlin, 39 adults wore the ring during polysomnography. Ring AHI correlated with PSG AHI at r = 0.89, with areas under the curve of 1.00, 0.90 and 0.98 at AHI cut-offs of 5, 15 and 30 (Lu et al. 2023, doi:10.1007/s13534-023-00304-9). Larger studies of the same software by company-affiliated authors report similar agreement. A small sample is not a guarantee for every patient.
- Can I buy a SleepImage ring without a prescription?
- SleepImage's patient pages describe a provider-led process: your provider issues the recorder and a Patient ID, and the results go to the provider for review. The software's FDA clearance is for prescription use.
- What is VivoScore?
- VivoScore is the name under which Vivos Therapeutics offers SleepImage-based home sleep testing through dental practices, under a 2021 agreement with MyCardio LLC disclosed in an SEC filing.
Related articles
Bilateral Ring PPG: What a 1,810-Person Study Shows About HR, BP, and Left–Right Differences
A large-scale bilateral ring PPG study measured both hands simultaneously in 1,810 participants. The results reveal where ring sensors excel (heart rate), where they show modest utility (age), and where limitations remain (blood pressure).
Overnight Oximetry vs Sleep Study: What Oximetry Can Show, What It Misses, and Why False Reassurance Happens
Overnight oximetry can be useful for screening and triage, but it is not a full sleep study. This guide explains what oximetry measures, where it fits clinically, and why normal results can still miss meaningful sleep apnea.
Oura Ring Sleep Tracking Accuracy: Polysomnography Validation Review
Oura Ring sleep tracking validation: de Zambotti polysomnography study results, sleep stage accuracy by stage, total sleep time error, and comparison vs Fitbit and Apple Watch.
Does Fitbit measure blood pressure? Which models get BP trends
No Fitbit shows mmHg. Blood pressure trends runs only on Fitbit Edge, Air and Pixel Watch 3–5, not Charge 6, Versa 4 or Inspire 3. What a High trend means.