Accuracy of Photoplethysmography-Derived Pulse Rate Variability Compared with Electrocardiography-Derived Heart Rate Variability: A Systematic Review and Meta-Analysis

Venue listed2026Added Nov 15, 2026

  • SX
  • HL
  • ZL
  • PS
  • ZG

Authors: Shiwen Xu, Hao Liu, Zhengliang Liu, Peng Su, Zhuangzhuang Gu

Sensors 26(16) 5192 (2026-08-17)

TL;DR

43 studies in the qualitative synthesis, 10 in the meta-analysis (8 each for RMSSD and SDNN), healthy non-clinical populations. Pooled absolute standardized error 0.188 (95% CI 0.066–0.309, I² 11.7%) for RMSSD and 0.134 (0.014–0.255, I² 0%) for SDNN. Mostly resting / controlled conditions; the authors warn against generalising to sleep, exercise, stress or free living, or treating PRV and HRV as interchangeable.

  • Form factor: mixed (wearable + camera)

Abstract

Photoplethysmography-derived pulse rate variability (PPG-derived PRV) is increasingly used in wearable and camera-based biomedical sensors as a low-burden alternative to electrocardiography-derived heart rate variability (ECG-derived HRV), but its agreement with ECG-derived measurements remains uncertain. This systematic review and meta-analysis evaluated the accuracy of PPG-derived PRV compared with ECG-derived HRV in healthy or apparently healthy non-clinical populations, focusing on the root mean square of successive differences (RMSSD) and the standard deviation of normal-to-normal intervals (SDNN). Forty-three studies were included in the qualitative synthesis; 33 were summarized narratively, and 10 unique studies provided sufficient data for quantitative synthesis. Eight studies contributed to each RMSSD and SDNN meta-analysis. The pooled absolute standardized error was 0.188 (95% CI: 0.066 to 0.309; I2 = 11.69%) for RMSSD and 0.134 (95% CI: 0.014 to 0.255; I2 = 0%) for SDNN. Sensitivity analyses supported the robustness of RMSSD findings, whereas SDNN estimates were directionally stable but less robust in statistical significance. Because the quantitative synthesis included only 10 studies and was based predominantly on selected resting or controlled conditions, the pooled estimates should not be generalized to sleep, exercise, stress, or free-living settings or interpreted as evidence of interchangeability.

Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://creativecommons.org/licenses/by/4.0/

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Signals

Keywords

PRVHRVRMSSDSDNNMeta AnalysisReview

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG, ECG); Form factor (mixed (wearable + camera)). Method: Method.SetupSignals: PPG, ECGSignalsPPG, ECGForm factor: mixed (wearable + camera)Form factormixed (wearable + camera)MethodMethod

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Xu, S., Liu, H., Liu, Z., Su, P., & Gu, Z. (2026). Accuracy of Photoplethysmography-Derived Pulse Rate Variability Compared with Electrocardiography-Derived Heart Rate Variability: A Systematic Review and Meta-Analysis. Sensors 26 5192. https://doi.org/10.3390/s26165192
Xu, Shiwen, Hao Liu, Zhengliang Liu, Peng Su, and Zhuangzhuang Gu. 2026. "Accuracy of Photoplethysmography-Derived Pulse Rate Variability Compared with Electrocardiography-Derived Heart Rate Variability: A Systematic Review and Meta-Analysis." Sensors 26 5192. https://doi.org/10.3390/s26165192.
Xu, S., Liu, H., Liu, Z., Su, P. and Gu, Z. (2026) 'Accuracy of Photoplethysmography-Derived Pulse Rate Variability Compared with Electrocardiography-Derived Heart Rate Variability: A Systematic Review and Meta-Analysis', Sensors 26 5192. doi: 10.3390/s26165192.
Xu, Shiwen, et al. "Accuracy of Photoplethysmography-Derived Pulse Rate Variability Compared with Electrocardiography-Derived Heart Rate Variability: A Systematic Review and Meta-Analysis." Sensors 26 5192, 2026. https://doi.org/10.3390/s26165192.
S. Xu, H. Liu, Z. Liu, P. Su, and Z. Gu, "Accuracy of Photoplethysmography-Derived Pulse Rate Variability Compared with Electrocardiography-Derived Heart Rate Variability: A Systematic Review and Meta-Analysis," Sensors 26 5192, 2026, doi: 10.3390/s26165192.
@article{xu2026accuracy,
  title         = {{Accuracy of Photoplethysmography-Derived Pulse Rate Variability Compared with Electrocardiography-Derived Heart Rate Variability: A Systematic Review and Meta-Analysis}},
  author        = {Xu, Shiwen and Liu, Hao and Liu, Zhengliang and Su, Peng and Gu, Zhuangzhuang},
  year          = {2026},
  journal       = {Sensors 26 5192},
  doi           = {10.3390/s26165192}
}

Added to the ChatPPG research hub