Validation of photoplethysmography-derived short-term heart rate variability using a wearable device (Bora Band vs 12-lead ECG)
Scientific Reports (2026-05-18)
TL;DR
Prospective single-centre validation of the wrist PPG Bora Band for 5.5-min resting HRV against simultaneous high-resolution 12-lead ECG in 66 participants in sinus rhythm; 16 HRV metrics. Strong agreement (biweight ρ, Cliff's δ, TOST): mean HR ρ=1.0, SDNN ρ=0.98, deceleration capacity ρ=0.95, CVNN ρ=0.98, Poincaré SD2 ρ=0.99. Moderate: DFA α2-type long-term scaling, VLF/LF/HF power.
- Form factor: wrist
Abstract
Heart rate variability (HRV) reflects cardiac autonomic function and cardiovascular health. While electrocardiography (ECG) remains the gold standard for HRV assessment, photoplethysmography (PPG)-based wearables offer an alternative for monitoring. This study aimed to validate the wrist-worn PPG device Bora band® for short-term HRV assessment against standard ECG and to characterize domain-specific agreement patterns. In this prospective, single-center observational study, 66 participants in sinus rhythm underwent simultaneous high-resolution 12-lead ECG and wrist-based PPG recording for 5 min and 30 s. PPG-derived interbeat intervals were extracted and compared to ECG-derived R-R intervals. Sixteen HRV metrics from time-, frequency-, and non-linear domains were computed for both modalities. Agreement was assessed using biweight mid-correlation (ρ), Cliff's delta (δ), and equivalence testing (TOST). Strong agreement was observed for mean heart rate (ρ = 1.0, δ = 0.104, TOST p < 0.001), standard deviation of normal-to-normal intervals (ρ = 0.98, δ = 0.037, TOST p = 0.003), deceleration capacity of heart rate (ρ = 0.95, δ = 0.071, TOST p = 0.014), coefficient of variation of normal-to-normal intervals (ρ = 0.98, δ = 0.035, TOST p = 0.002) and Poincaré Plot standard deviation 2 (ρ = 0.99, δ = 0.074, TOST p < 0.001). Moderate agreement was noted for long-term fractal scaling exponent and very low-, low- and high-frequency power. Weaker agreement appeared for short-term variability and entropy metrics. Bland-Altman analysis indicated minimal bias without systematic error. Under controlled resting conditions, wrist-based PPG provides reliable HRV indices compared with ECG-derived HRV. These findings support the use of selected PPG-derived HRV parameters for short-term assessment in clinical settings. Further validation in real-world settings will be necessary.
Verbatim from Europe PMC (opens in a new tab) · Licence: cc by-nc-nd
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Signals
Keywords
How it works
Cite this paper
Zuern, C. S., Felkel, M., Tilquin, F., Le Guillou, Y., Dervieux, E., Hämmerle, P., Kaplan, E., Mahfoud, F., Speich, B., Briel, M., Labhardt, N. D., & Zhou, Q. (2026). Validation of photoplethysmography-derived short-term heart rate variability using a wearable device (Bora Band vs 12-lead ECG). Scientific Reports. https://doi.org/10.1038/s41598-026-52700-7
Zuern, Christine S., Maximilian Felkel, Florian Tilquin, Yann Le Guillou, Emmanuel Dervieux, Peter Hämmerle, and Emel Kaplan, et al. 2026. "Validation of photoplethysmography-derived short-term heart rate variability using a wearable device (Bora Band vs 12-lead ECG)." Scientific Reports. https://doi.org/10.1038/s41598-026-52700-7.
Zuern, C. S., Felkel, M., Tilquin, F., Le Guillou, Y., Dervieux, E., Hämmerle, P. and Kaplan, E. et al. (2026) 'Validation of photoplethysmography-derived short-term heart rate variability using a wearable device (Bora Band vs 12-lead ECG)', Scientific Reports. doi: 10.1038/s41598-026-52700-7.
Zuern, Christine S., et al. "Validation of photoplethysmography-derived short-term heart rate variability using a wearable device (Bora Band vs 12-lead ECG)." Scientific Reports, 2026. https://doi.org/10.1038/s41598-026-52700-7.
C. S. Zuern et al., "Validation of photoplethysmography-derived short-term heart rate variability using a wearable device (Bora Band vs 12-lead ECG)," Scientific Reports, 2026, doi: 10.1038/s41598-026-52700-7.
@article{zuern2026validation,
title = {{Validation of photoplethysmography-derived short-term heart rate variability using a wearable device (Bora Band vs 12-lead ECG)}},
author = {Zuern, Christine S. and Felkel, Maximilian and Tilquin, Florian and Le Guillou, Yann and Dervieux, Emmanuel and Hämmerle, Peter and Kaplan, Emel and Mahfoud, Felix and Speich, Benjamin and Briel, Matthias and Labhardt, Niklaus Daniel and Zhou, Qian},
year = {2026},
journal = {Scientific Reports},
doi = {10.1038/s41598-026-52700-7}
}Added to the ChatPPG research hub