PaperarXiv:2609.30122

Melanin- and linewidth-corrected pulse-oximetry

Preprint2026Added Sep 26, 2026

  • GB
  • JF
  • RD
  • SF
  • VT

Authors: Giles Blaney, Jodee Frias, Ravi Durbha, Sergio Fantini, Valencia Koomson (Tufts)

arXiv preprint (eess / biomedical optics)

Open code

TL;DR

Fits three SpO2 calibration forms (empirical linear; mBLL-derived with offset; melanin- and linewidth-aware mBLL) on the public OpenOximetry cohort — 98 patients with paired ABG SaO2, recoverable RoR from raw red/IR PPG, and melanin volume fraction M from reflectance. Overall accuracy is nearly identical (~2.42% Arms), but residual dependence on M falls from about +2.8 → +1.2 → +0.1 pp SaO2 per unit melanin fraction across the three cases (direction consistent with spectral-coloring theory; slopes not statistically resolved at n=98; authors estimate ~10× cohort needed). Melanin-aware case needs a per-patient skin-tone measurement and assumed LED FWHM.

  • Form factor: finger

Abstract

Current pulse oximeters tend to overestimate arterial saturation in darkly pigmented patients, increasing occult hypoxemia rates. One proposed mechanism is spectral coloring: melanin contributes to reshaping the detected spectrum of a broad-linewidth source, making the wavelength-dependent parameters behind calibration skin-tone-dependent. Diffuse optics theory allows for a melanin- and linewidth-corrected calibration equation, which to our knowledge has never been tested in vivo. In this work, we fit the empirical linear equation, the diffuse optics theoretical equation, and its corrected form to the public OpenOximetry Repository—considering 98 patients with paired arterial blood-draws, raw photoplethysmograms, and skin reflectance spectra. Accuracy is practically indistinguishable between the three approaches, yet the saturation error's dependence on melanin falls from +2.8 to +1.2 to +0.1 percentage points per unit melanin volume fraction—consistent with the spectral-coloring theory. None of these dependencies on melanin are statistically resolved here—a cohort roughly ten times larger is needed—so the correction is backed by theory and its expected direction. All three calibration approaches carry two degrees of freedom: the theoretical equation drops into existing calibration procedures, and its melanin-aware form needs a per-patient skin-tone measurement. This work demonstrates how the theoretical pulse oximetry model can be applied in calibration.

Verbatim from arXiv (opens in a new tab) · arXiv metadata (abstracts are released under CC0)

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

SpO₂

Keywords

fairnessskin-tonemelaninSpO2opticsOpenOximetrypulse-oximetry

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG, ABG SaO2, skin reflectance); Wavelengths (red, IR); Form factor (finger). Method: Method. Target: SpO₂.SetupSignals: PPG, ABG SaO2, skin reflectanceSignalsPPG, ABG SaO2, skin reflectanceWavelengths: red, IRWavelengthsred, IRForm factor: fingerForm factorfingerMethodMethodTargetSpO₂O₂SpO₂
Wavelengths
red, IR

Cite this paper

APA

Blaney, G., Frias, J., Durbha, R., Fantini, S., & Koomson, V. (2026). Melanin- and linewidth-corrected pulse-oximetry. arXiv preprint. https://doi.org/10.48550/arXiv.2609.30122

BibTeX

@misc{blaney2026melanin,
  title         = {{Melanin- and linewidth-corrected pulse-oximetry}},
  author        = {Blaney, Giles and Frias, Jodee and Durbha, Ravi and Fantini, Sergio and Koomson, Valencia},
  year          = {2026},
  howpublished  = {arXiv preprint},
  doi           = {10.48550/arXiv.2609.30122},
  eprint        = {2609.30122},
  archivePrefix = {arXiv}
}

Built from the hub record. Check it against the publisher before submitting.

Added to the ChatPPG research hub