Paperdoi:10.1055/a-2998-9516
Accuracy and Reliability of Pulse Oxygen Saturation Measured Using Wrist-Worn Oximeters (Moreigneaux et al., Int J Sports Med 2026)
International Journal of Sports Medicine (Thieme), online ahead of print 2026-11-23
TL;DR
Twenty-six healthy people did rest and mild exercise at sea level and at simulated altitudes of 3,000-5,500 m while wearing a Garmin Forerunner 965, a Polar Vantage 3 and a Checkme O2. All three gave SpO2 values that differed from the criterion; more than half of readings fell outside ±3%, and concordance was poor for all devices at every altitude. The authors conclude none gave trustworthy SpO2 at sea level or altitude.
Abstract
This study aimed to evaluate the accuracy of pulse oxygen saturation measured using wrist-worn devices (Garmin Forerunner 965, Polar Vantage 3 and Checkme O2), from sea level to high altitude. Twenty-six healthy subjects underwent five tests under normoxia and hypoxia conditions (simulated altitudes from 3,000 to 5,500 m), consisting of a 5-minute rest phase, followed by 5 minutes of mild exercise. Bland–Altman parameters, absolute error (around±3% of the criterion), mean absolute percentage error, and the lower bounds of intraclass correlation coefficient and concordance correlation coefficient were calculated. Device-measured pulse oxygen saturation values were different from the criterion values (from p<0.01 to p<0.001) and biases were different from 0 (positive for Garmin and CheckmeO2 and negative for Polar); absolute error rates exceeded 50% for all devices. Intraclass correlation coefficient criteria for Polar and Garmin were poor but good for Checkme O2, whereas concordance correlation coefficient were poor for all devices, overall and at every altitude. All devices failed to provide trustworthy pulse oxygen saturation values at both sea level and altitude, where a pronounced arterial oxygen desaturation could lead to acute mountain sickness in hypoxia-sensitive subjects, and its life-threatening complications.
Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://creativecommons.org/licenses/by-nc-nd/4.0/
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Tasks
Signals
Keywords
How it works
Cite this paper
Moreigneaux, O., Ceugniez, M., Lhuissier, F. J., Denis, L., Bourse, J., Gille, T., & Hermand, E. (2026). Accuracy and Reliability of Pulse Oxygen Saturation Measured Using Wrist-Worn Oximeters (Moreigneaux et al., Int J Sports Med 2026). International Journal of Sports Medicine. https://doi.org/10.1055/a-2998-9516
Moreigneaux, O, M Ceugniez, F J Lhuissier, L Denis, J Bourse, T Gille, and E Hermand. 2026. "Accuracy and Reliability of Pulse Oxygen Saturation Measured Using Wrist-Worn Oximeters (Moreigneaux et al., Int J Sports Med 2026)." International Journal of Sports Medicine. https://doi.org/10.1055/a-2998-9516.
Moreigneaux, O., Ceugniez, M., Lhuissier, F. J., Denis, L., Bourse, J., Gille, T. and Hermand, E. (2026) 'Accuracy and Reliability of Pulse Oxygen Saturation Measured Using Wrist-Worn Oximeters (Moreigneaux et al., Int J Sports Med 2026)', International Journal of Sports Medicine. doi: 10.1055/a-2998-9516.
Moreigneaux, O, et al. "Accuracy and Reliability of Pulse Oxygen Saturation Measured Using Wrist-Worn Oximeters (Moreigneaux et al., Int J Sports Med 2026)." International Journal of Sports Medicine, 2026. https://doi.org/10.1055/a-2998-9516.
O. Moreigneaux et al., "Accuracy and Reliability of Pulse Oxygen Saturation Measured Using Wrist-Worn Oximeters (Moreigneaux et al., Int J Sports Med 2026)," International Journal of Sports Medicine, 2026, doi: 10.1055/a-2998-9516.
@article{moreigneaux2026accuracy,
title = {{Accuracy and Reliability of Pulse Oxygen Saturation Measured Using Wrist-Worn Oximeters (Moreigneaux et al., Int J Sports Med 2026)}},
author = {Moreigneaux, O. and Ceugniez, M. and Lhuissier, F. J. and Denis, L. and Bourse, J. and Gille, T. and Hermand, E.},
year = {2026},
journal = {International Journal of Sports Medicine},
publisher = {Thieme},
doi = {10.1055/a-2998-9516}