Beyond average accuracy: motion-related failure patterns in wrist PPG heart-rate monitoring (Xu, Theor. Nat. Sci. 2026)

Venue listed2026Added Nov 5, 2026

  • AX

Authors: Angel Xu

Theoretical and Natural Science (EWA Publishing), 2026-09-22

TL;DR

Uses PPG-DaLiA (15 participants, Empatica E4 wrist PPG + chest ECG). 8-s windows, 2-s step; HR from a dominant-frequency PPG method. A large error is >10 bpm vs ECG.

  • Form factor: wrist

Abstract

Wrist-worn photoplethysmography (PPG) enables unobtrusive heart rate monitoring, but motion-related errors tend to cluster under specific conditions and are often masked by aggregate accuracy metrics. This study characterized those failure patterns and assessed whether wrist-sensor features could identify unreliable estimates in previously unseen participants. PPG-DaLiA recordings from 15 participants, acquired with an Empatica E4 and synchronized chest electrocardiography during daily-life activities, were segmented into 8-s windows with a 2-s step. Heart rate was estimated using a transparent dominant-frequency PPG method. A large error was defined a priori as an absolute deviation of more than 10 beats per minute (bpm) relative to the ECG-derived reference. Motion intensity, acceleration-band power, PPG-acceleration spectral overlap, peak dominance and spectral entropy were used as reliability predictors. Of 47,182 non-transition windows, 14,622 (31.0%) met the large-error criterion. Failure prevalence differed substantially by activity, from 5.8% during sitting to 63.8% during stair climbing. In leave-one-subject-out validation, a motion-only model achieved an AUROC of 0.743, wrist-signal features increased AUROC to 0.851, and activity context increased it to 0.864. Wrist PPG reliability is therefore condition dependent, and an interpretable signal-based reliability flag may complement point heart-rate estimates. External prospective validation is required before clinical or consumer deployment.

Verbatim from the publisher's DOI record (opens in a new tab) · Abstract as deposited by the publisher with Crossref; © the authors/publisher

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

Signal denoisingSignal qualityHeart rate

Signals

Keywords

Motion ArtifactSQIReliability FlagHeart RatePPG DaliaEmpatica E4Loso

Datasets used

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG, ACC, ECG); Form factor (wrist). Method: Method. Target: Signal denoising; Signal quality; Heart rate.SetupSignals: PPG, ACC, ECGSignalsPPG, ACC, ECGForm factor: wristForm factorwristMethodMethodTargetSignal denoisingSignal denoisingSignal qualitySignal qualityHeart rateHeart rate

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Xu, A. (2026). Beyond average accuracy: motion-related failure patterns in wrist PPG heart-rate monitoring (Xu, Theor. Nat. Sci. 2026). Theoretical and Natural Science. https://doi.org/10.54254/2753-8818/2026.es37117
Xu, Angel. 2026. "Beyond average accuracy: motion-related failure patterns in wrist PPG heart-rate monitoring (Xu, Theor. Nat. Sci. 2026)." Theoretical and Natural Science. https://doi.org/10.54254/2753-8818/2026.es37117.
Xu, A. (2026) 'Beyond average accuracy: motion-related failure patterns in wrist PPG heart-rate monitoring (Xu, Theor. Nat. Sci. 2026)', Theoretical and Natural Science. doi: 10.54254/2753-8818/2026.es37117.
Xu, Angel "Beyond average accuracy: motion-related failure patterns in wrist PPG heart-rate monitoring (Xu, Theor. Nat. Sci. 2026)." Theoretical and Natural Science, 2026. https://doi.org/10.54254/2753-8818/2026.es37117.
A. Xu, "Beyond average accuracy: motion-related failure patterns in wrist PPG heart-rate monitoring (Xu, Theor. Nat. Sci. 2026)," Theoretical and Natural Science, 2026, doi: 10.54254/2753-8818/2026.es37117.
@article{xu2026beyond,
  title         = {{Beyond average accuracy: motion-related failure patterns in wrist PPG heart-rate monitoring (Xu, Theor. Nat. Sci. 2026)}},
  author        = {Xu, Angel},
  year          = {2026},
  journal       = {Theoretical and Natural Science},
  doi           = {10.54254/2753-8818/2026.es37117}
}

Added to the ChatPPG research hub