Paperdoi:10.1109/MLSP68619.2026.11772589

ECG Foundation Model Features Improve Atrial Fibrillation Alarm Precision In Intensive Care Monitoring

Venue listed2026Added Nov 22, 2026

  • AL
  • SC
  • ZM
  • RY
  • SK
  • LC
  • SW
  • XH
  • +1

Authors: Andrew Lu, Selene Cho, Ze-Yuan Meng, Run-Ze Yan, Saurabh Kataria, Lisa Castle, Stacy B. Westerman, Xiao Hu, Ran Xiao

IEEE MLSP 2026 (2026-09)

TL;DR

Tests whether ECG foundation-model (ECG-FM) embeddings separate true AF alarms from AF-mimicking false alarms better than engineered ECG features. 711 clinician-annotated ICU AF alarms with matched 30-s ECG (159 false, 22.4%); stratified group cross-validation. ECG-FM embeddings were best: AUROC 0.903, AUPRC 0.963; at ~90% sensitivity PPV rose from the monitor's baseline 0.776 to 0.945, suppressing 130 of 159 false alarms.

  • Form factor: bedside ICU monitor

Abstract

Atrial fibrillation (AF) alarms generated by intensive care unit bedside monitors are designed to be sensitive, but false alarms remain common and contribute to alarm fatigue. We evaluated whether ECG foundation model (ECG-FM) embeddings improve discrimination of true AF alarms from AF-mimicking false alarms compared with engineered ECG features. Clinicians annotated 711 AF alarms with matched 30-second ECG waveforms, of which 159 (22.4%) were false alarms. We compared engineered ECG features, ECG-FM embeddings, and demographic features using stratified group cross-validation. ECG-FM achieved the best performance, with area under the receiver operating characteristic curve (AUROC) of 0.903 and area under the precision-recall curve of 0.963. At approximately 90% sensitivity, ECG-FM increased positive predictive value from the baseline alarm precision of 0.776 to 0.945 and suppressed 130 of 159 false alarms. These findings suggest that ECG foundation model representations can substantially improve AF alarm precision while preserving high sensitivity.

Verbatim from Semantic Scholar (opens in a new tab) · Abstract via Semantic Scholar; © the authors/publisher

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

AF / rhythmStress

Signals

Keywords

Foundation ModelAFAlarm FatigueFalse AlarmICUEmbeddings

How it works

How Method works: Setup → Method → TargetSetup: Signals (ECG); Form factor (bedside ICU monitor). Method: Method. Target: AF / rhythm; Stress.SetupSignals: ECGSignalsECGForm factor: bedside ICU monitorForm factorbedside ICU monitorMethodMethodTargetAF / rhythmAF / rhythmStressStress

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Lu, A., Cho, S., Meng, Z.-Y., Yan, R.-Z., Kataria, S., Castle, L., Westerman, S. B., Hu, X., & Xiao, R. (2026). ECG Foundation Model Features Improve Atrial Fibrillation Alarm Precision In Intensive Care Monitoring. IEEE MLSP 2026. https://doi.org/10.1109/MLSP68619.2026.11772589
Lu, Andrew, Selene Cho, Ze-Yuan Meng, Run-Ze Yan, Saurabh Kataria, Lisa Castle, Stacy B. Westerman, Xiao Hu, and Ran Xiao. 2026. "ECG Foundation Model Features Improve Atrial Fibrillation Alarm Precision In Intensive Care Monitoring." IEEE MLSP 2026. https://doi.org/10.1109/MLSP68619.2026.11772589.
Lu, A., Cho, S., Meng, Z.-Y., Yan, R.-Z., Kataria, S., Castle, L., Westerman, S. B., Hu, X. and Xiao, R. (2026) 'ECG Foundation Model Features Improve Atrial Fibrillation Alarm Precision In Intensive Care Monitoring', IEEE MLSP 2026. doi: 10.1109/MLSP68619.2026.11772589.
Lu, Andrew, et al. "ECG Foundation Model Features Improve Atrial Fibrillation Alarm Precision In Intensive Care Monitoring." IEEE MLSP 2026, 2026. https://doi.org/10.1109/MLSP68619.2026.11772589.
A. Lu et al., "ECG Foundation Model Features Improve Atrial Fibrillation Alarm Precision In Intensive Care Monitoring," IEEE MLSP 2026, 2026, doi: 10.1109/MLSP68619.2026.11772589.
@article{lu2026foundation,
  title         = {{ECG Foundation Model Features Improve Atrial Fibrillation Alarm Precision In Intensive Care Monitoring}},
  author        = {Lu, Andrew and Cho, Selene and Meng, Ze-Yuan and Yan, Run-Ze and Kataria, Saurabh and Castle, Lisa and Westerman, Stacy B. and Hu, Xiao and Xiao, Ran},
  year          = {2026},
  journal       = {IEEE MLSP 2026},
  doi           = {10.1109/MLSP68619.2026.11772589}
}

Added to the ChatPPG research hub