Paperdoi:10.1038/s41430-026-01848-6

Wearable vs standing bioimpedance vs BIS for total body water in adults with Down syndrome (Esco et al., EJCN 2026)

Venue listed2026Added Nov 13, 2026

Eur J Clin Nutr, published 2026-11-09 (subscription)

TL;DR

Abstract only (full text gated). 16 adults with Down syndrome; TBW from BIS (reference), one wearable bioimpedance device and two standing multi-frequency BIA analyzers. The wearable did not differ from BIS (p = 0.106) and had the smallest bias (-1.07 +/- 4.64 L), highest CCC (range 0.82-0.92 across field devices) and lowest MAPE (range 5.5-8.7%).

  • Form factor: wearable (model not named in abstract)

Abstract

BackgroundWearable bioimpedance technology offers a potentially accessible method for assessing total body water (TBW), yet its agreement with laboratory-based approaches in adults with Down syndrome remains unclear. The purpose of this study was to evaluate whether a wearable bioimpedance device demonstrates agreement with laboratory-based bioelectrical impedance spectroscopy (BIS) estimates of TBW comparable to standing bioimpedance devices in adults with Down syndrome.MethodsSixteen adults with Down syndrome (age: 28.3 ± 9.2 yr; height: 150.9 ± 7.8 cm; body mass: 77.4 ± 20.3 kg) completed TBW assessment using BIS (reference), a wearable bioimpedance device (BIAW), and two standing multi-frequency bioimpedance analyzers (BIAST1, BIAST2).ResultsRepeated measures ANOVA revealed significant differences among devices (F (3,42) = 17.52, p W did not differ from BIS (p = 0.106). Lin's concordance correlation coefficients demonstrated strong agreement between BIS and all field-based devices (CCC = 0.82-0.92), with the highest concordance observed for the wearable device. Bland-Altman analyses showed systematic underestimation for standing devices (-3.12 ± 4.65 L and -2.36 ± 4.64 L), while the wearable exhibited the smallest bias (-1.07 ± 4.64 L). Mean absolute percentage error ranged from 5.5% to 8.7%, with the wearable device demonstrating the lowest relative error.ConclusionThese findings indicate that wearable bioimpedance technology demonstrated closer agreement with TBW estimates from BIS than standing BIA devices.

Verbatim from Europe PMC (opens in a new tab) · Abstract via Europe PMC; © the authors/publisher

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

Hydration

Keywords

ValidationTbwHydrationWrist Electrical SensingGated

How it works

How Method works: Setup → Method → TargetSetup: Signals (bioimpedance); Form factor (wearable (model not named in abstract)). Method: Method. Target: Hydration.SetupSignals: bioimpedanceSignalsbioimpedanceForm factor: wearable (model not named in abstract)Form factorwearable (model not named in abstract)MethodMethodTargetHydrationHydration

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Wearable vs standing bioimpedance vs BIS for total body water in adults with Down syndrome (Esco et al., EJCN 2026). (2026). Eur J Clin Nutr. https://doi.org/10.1038/s41430-026-01848-6
@article{escowearablebiat2026wearable,
  title         = {{Wearable vs standing bioimpedance vs BIS for total body water in adults with Down syndrome (Esco et al., EJCN 2026)}},
  year          = {2026},
  journal       = {Eur J Clin Nutr},
  doi           = {10.1038/s41430-026-01848-6}
}

Added to the ChatPPG research hub