Paperdoi:10.1007/s10916-022-01872-6
Evaluation of a wearable in-ear sensor for temperature and heart rate monitoring: a pilot study
Journal of Medical Systems, 2022
TL;DR
Pilot study of the Cosinuss One in-ear sensor, which measures ear-canal temperature and PPG heart rate, in patients undergoing non-cardiac surgery. Against urinary bladder temperature the limits of agreement were -0.79 to 0.49 °C (89% of readings within ±0.5 °C), against a zero-heat-flux thermometer -0.78 to 0.34 °C (79%), and heart rate agreed with ECG within -4.81 to 4.27 bpm.
Abstract
In the context of the COVID-19 pandemic, wearable sensors are important for early detection of critical illness especially in COVID-19 outpatients. We sought to determine in this pilot study whether a wearable in-ear sensor for continuous body temperature and heart rate monitoring (Cosinuss company, Munich) is sufficiently accurate for body temperature and heart rate monitoring. Comparing with several anesthesiologic standard of care monitoring devices (urinary bladder and zero-heat flux thermometer and ECG), we evaluated the in-ear sensor during non-cardiac surgery (German Clinical Trials Register Reg.-No: DRKS00012848). Limits of Agreement (LoA) based on Bland–Altman analysis were used to study the agreement between the in-ear sensor and the reference methods. The estimated LoA of the Cosinuss One and bladder temperature monitoring were [-0.79, 0.49] °C (95% confidence intervals [-1.03, -0.65] (lower LoA) and [0.35, 0.73] (upper LoA)), and [-0.78, 0.34] °C (95% confidence intervals [-1.18, -0.59] (lower LoA) and [0.16, 0.74] (upper LoA)) of the Cosinuss One and zero-heat flux temperature monitoring. 89% and 79% of Cosinuss One temperature monitoring were within ± 0.5 °C limit of bladder and zero-heat flux monitoring, respectively. The estimated LoA of Cosinuss One and ECG heart rate monitoring were [-4.81, 4.27] BPM (95% confidence intervals [-5.09, -4.56] (lower LoA) and [4.01, 4.54] (upper LoA)). The proportion of detection differences within ± 2BPM was 84%. Body temperature and heart rate were reliably measured by the wearable in-ear sensor.
Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://creativecommons.org/licenses/by/4.0
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Signals
Keywords
How it works
Cite this paper
Ellebrecht, D., Gola, D., & Kaschwich, M. (2022). Evaluation of a wearable in-ear sensor for temperature and heart rate monitoring: a pilot study. Journal of Medical Systems, 2022. https://doi.org/10.1007/s10916-022-01872-6
Ellebrecht, DB, D Gola, and M Kaschwich. 2022. "Evaluation of a wearable in-ear sensor for temperature and heart rate monitoring: a pilot study." Journal of Medical Systems, 2022. https://doi.org/10.1007/s10916-022-01872-6.
Ellebrecht, D., Gola, D. and Kaschwich, M. (2022) 'Evaluation of a wearable in-ear sensor for temperature and heart rate monitoring: a pilot study', Journal of Medical Systems, 2022. doi: 10.1007/s10916-022-01872-6.
Ellebrecht, DB, et al. "Evaluation of a wearable in-ear sensor for temperature and heart rate monitoring: a pilot study." Journal of Medical Systems, 2022, 2022. https://doi.org/10.1007/s10916-022-01872-6.
D. Ellebrecht, D. Gola, and M. Kaschwich, "Evaluation of a wearable in-ear sensor for temperature and heart rate monitoring: a pilot study," Journal of Medical Systems, 2022, 2022, doi: 10.1007/s10916-022-01872-6.
@article{ellebrecht2022evaluation,
title = {{Evaluation of a wearable in-ear sensor for temperature and heart rate monitoring: a pilot study}},
author = {Ellebrecht, DB and Gola, D. and Kaschwich, M.},
year = {2022},
journal = {Journal of Medical Systems, 2022},
doi = {10.1007/s10916-022-01872-6}
}Added to the ChatPPG research hub