Robust synchronization of cardiac-linked biosensors under peak detection failures: a modality-agnostic approach via structural repair and Huber-IRLS regression

Venue listed2026Added Nov 9, 2026

  • KO

Authors: Khaled Ben Abdessalem (single author per OpenAlex)

Measurement Science and Technology (IOP), published 2026-11-06

First page or project page of “Robust synchronization of cardiac-linked biosensors under peak detection failures: a modality-agnostic approach via structural repair and Huber-IRLS regression” (opens in a new tab)

Khaled Ben Abdessalem, "Robust synchronization of cardiac-linked biosensors under peak detection failures: a modality-agnostic approach via structural repair and Huber-IRLS regression", doi:10.1088/1361-6501/aeb820, CC BY 4.0

TL;DR

Problem: timing alignment between independently clocked sensors (e.g. ECG and PPG) using inter-beat intervals only, without waveform alignment. Classic estimators assume one-to-one beat correspondence, which missed beats, motion-induced false peaks and gross interval errors break. Pipeline: (i) Dynamic Repair of abnormally short split intervals caused by false peaks; (ii) greedy peak-failure-aware mapping with q ∈ {1, 2} that models isolated missed beats as merged intervals; (iii) Huber-IRLS regression with MAD-based residual gating.

  • Form factor: other

Abstract

Cooperative sensing in clinical instruments, distributed medical devices, and body-area networks requires accurate relative timing between independently clocked sensors, such as electrocardiography (ECG) and photoplethysmography (PPG). Interval-only synchronization operates directly on inter-beat intervals and avoids waveform alignment, but conventional estimators generally assume one-to-one reference–device beat correspondence. Missed beats, motion-induced false peaks, and gross interval errors violate this assumption by creating merged or split intervals and index shifts. We propose a robust interval-only synchronization pipeline comprising: (i) Dynamic Repair of abnormally short split intervals caused by extra (false) peaks; (ii) greedy peak-failure-aware mapping with

q ∈

1 , 2

to represent isolated missed beats as merged intervals; and (iii) Huber-iteratively reweighted least squares regression with median absolute deviation-based residual gating. Approximate uncertainty is quantified using an OLS-style Newey–West HAC covariance estimate (approximate and conditional when gating) and Fieller confidence sets for the ratio-defined delay-dynamics parameter. Monte Carlo ablation on synthetic ECG/PPG data (ORACLE and SIGNAL pipelines) shows that structural peak-detection failures can be catastrophic for OLS (e.g.

RMS

E

after , all

≈

227

ms

under Extra 10% and

≈ 232

ms

under a mixed extreme regime). In contrast, the proposed pipeline maintains robustness; in the mixed extreme regime it achieves

RMS

E

after , in

≈ 37.4

ms

(SIGNAL) and ≈35.9 ms (ORACLE). The real-data experiment was performed as a semi-synthetic feasibility validation using matched ECG/PPG beats from a real recording, with known clock skew imposed on the PPG timestamps and controlled peak failures injected into the resulting device interval sequence.

Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://creativecommons.org/licenses/by/4.0/

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

Signal denoisingPulse arrival time

Signals

Keywords

ECG PPG SynchronizationInter Beat IntervalMotion ArtifactMissed BeatsRobust RegressionHuber IrlsClock SkewPulse Arrival Time

How it works

How Method works: Setup → Method → TargetSetup: Signals (ECG, PPG); Form factor (other). Method: Method. Target: Signal denoising; Pulse arrival time.SetupSignals: ECG, PPGSignalsECG, PPGForm factor: otherForm factorotherMethodMethodTargetSignal denoisingSignal denoisingPulse arrival timePulse arrival time

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Abdessalem, K. B. (2026). Robust synchronization of cardiac-linked biosensors under peak detection failures: a modality-agnostic approach via structural repair and Huber-IRLS regression. Measurement Science and Technology. https://doi.org/10.1088/1361-6501/aeb820
Abdessalem, Khaled Ben. 2026. "Robust synchronization of cardiac-linked biosensors under peak detection failures: a modality-agnostic approach via structural repair and Huber-IRLS regression." Measurement Science and Technology. https://doi.org/10.1088/1361-6501/aeb820.
Abdessalem, K. B. (2026) 'Robust synchronization of cardiac-linked biosensors under peak detection failures: a modality-agnostic approach via structural repair and Huber-IRLS regression', Measurement Science and Technology. doi: 10.1088/1361-6501/aeb820.
Abdessalem, Khaled Ben "Robust synchronization of cardiac-linked biosensors under peak detection failures: a modality-agnostic approach via structural repair and Huber-IRLS regression." Measurement Science and Technology, 2026. https://doi.org/10.1088/1361-6501/aeb820.
K. B. Abdessalem, "Robust synchronization of cardiac-linked biosensors under peak detection failures: a modality-agnostic approach via structural repair and Huber-IRLS regression," Measurement Science and Technology, 2026, doi: 10.1088/1361-6501/aeb820.
@article{abdessalem2026robust,
  title         = {{Robust synchronization of cardiac-linked biosensors under peak detection failures: a modality-agnostic approach via structural repair and Huber-IRLS regression}},
  author        = {Abdessalem, Khaled Ben},
  year          = {2026},
  journal       = {Measurement Science and Technology},
  publisher     = {IOP},
  doi           = {10.1088/1361-6501/aeb820}
}

Added to the ChatPPG research hub