Paperdoi:10.1007/s12028-010-9349-y
Cerebral Arterial Oxygen Saturation Measurements Using a Fiber-Optic Pulse Oximeter
Neurocritical Care
TL;DR
A pilot investigation was undertaken to assess the performance of a novel fiber-optic cerebral pulse oximetry system. A fiber-optic probe designed to pass through the lumen of a cranial bolt of the type used to make intracranial pressure measurements was used to obtain optical reflectance signals directly from brain tissue. Short-duration measurements were made in six patients undergoing neurosurgery.
- Published in Neurocritical Care (2010).
Abstract
BackgroundA pilot investigation was undertaken to assess the performance of a novel fiber-optic cerebral pulse oximetry system. A fiber-optic probe designed to pass through the lumen of a cranial bolt of the type used to make intracranial pressure measurements was used to obtain optical reflectance signals directly from brain tissue.MethodsShort-duration measurements were made in six patients undergoing neurosurgery. These were followed by a longer duration measurement in a patient recovering from an intracerebral hematoma. Estimations of cerebral arterial oxygen saturation derived from a frequency domain-based algorithm are compared with simultaneous pulse oximetry (SpO2) and hemoximeter (SaO2) blood samples.ResultsThe short-duration measurements showed that reliable photoplethysmographic signals could be obtained from the brain tissue. In the long-duration study, the mean (±SD) difference between cerebral oxygen saturation (ScaO2) and finger SpO2 (in saturation units) was -7.47(±3.4)%. The mean (±SD) difference between ScaO2 and blood SaO2 was -7.37(±2.8)%.ConclusionsThis pilot study demonstrated that arterial oxygen saturation may be estimated from brain tissue via a fiber-optic pulse oximeter used in conjunction with a cranial bolt. Further studies are needed to confirm the clinical utility of the technique.
Verbatim from Europe PMC (opens in a new tab) · Abstract via Europe PMC; © the authors/publisher
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Signals
Keywords
How it works
Cite this paper
Phillips, J. P., Langford, R. M., Chang, S. H., Maney, K., Kyriacou, P. A., & Jones, D. P. (2010). Cerebral Arterial Oxygen Saturation Measurements Using a Fiber-Optic Pulse Oximeter. Neurocritical Care. https://doi.org/10.1007/s12028-010-9349-y
Phillips, J. P., R. M. Langford, S. H. Chang, K. Maney, P. A. Kyriacou, and D. P. Jones. 2010. "Cerebral Arterial Oxygen Saturation Measurements Using a Fiber-Optic Pulse Oximeter." Neurocritical Care. https://doi.org/10.1007/s12028-010-9349-y.
Phillips, J. P., Langford, R. M., Chang, S. H., Maney, K., Kyriacou, P. A. and Jones, D. P. (2010) 'Cerebral Arterial Oxygen Saturation Measurements Using a Fiber-Optic Pulse Oximeter', Neurocritical Care. doi: 10.1007/s12028-010-9349-y.
Phillips, J. P., et al. "Cerebral Arterial Oxygen Saturation Measurements Using a Fiber-Optic Pulse Oximeter." Neurocritical Care, 2010. https://doi.org/10.1007/s12028-010-9349-y.
J. P. Phillips, R. M. Langford, S. H. Chang, K. Maney, P. A. Kyriacou, and D. P. Jones, "Cerebral Arterial Oxygen Saturation Measurements Using a Fiber-Optic Pulse Oximeter," Neurocritical Care, 2010, doi: 10.1007/s12028-010-9349-y.
@article{phillips2010cerebral,
title = {{Cerebral Arterial Oxygen Saturation Measurements Using a Fiber-Optic Pulse Oximeter}},
author = {Phillips, J. P. and Langford, R. M. and Chang, S. H. and Maney, K. and Kyriacou, P. A. and Jones, D. P.},
year = {2010},
journal = {Neurocritical Care},
doi = {10.1007/s12028-010-9349-y}
}Added to the ChatPPG research hub