In Vitro Evaluation of a Non-Invasive Photoplethysmography Based Intracranial Pressure Sensor

Venue listed2022Added Nov 5, 2026

  • TA
  • JP
  • CU
  • MR
  • PK

Authors: Tomas Y. Abay, Justin P. Phillips, Christopher Uff, Maria Roldan, Panicos A. Kyriacou

Applied Sciences

First page or project page of “In Vitro Evaluation of a Non-Invasive Photoplethysmography Based Intracranial Pressure Sensor” (opens in a new tab)

Tomas Y. Abay et al., "In Vitro Evaluation of a Non-Invasive Photoplethysmography Based Intracranial Pressure Sensor", doi:10.3390/app13010534, CC BY 4.0

TL;DR

Intracranial pressure (ICP) is an important measurement in the treatment of Traumatic Brain Injury (TBI). Currently, ICP can only be measured invasively, which exposes patients to operative risk and can only be performed by neurosurgeons. Hence, there is a significant need for a non-invasive ICP technology.

Abstract

Intracranial pressure (ICP) is an important measurement in the treatment of Traumatic Brain Injury (TBI). Currently, ICP can only be measured invasively, which exposes patients to operative risk and can only be performed by neurosurgeons. Hence, there is a significant need for a non-invasive ICP technology. This paper describes the evaluation of a novel non-invasive intracranial pressure (nICP) monitor which uses the Photoplethysmogram (PPG) to measure the ICP. The monitor was evaluated in an in vitro model that simulated cerebral haemodynamics and allowed the controlled manipulation of ICP. A number of features from the PPG were extracted and utilised in a machine learning model to estimate ICP. Three separate measurements in which the ICP was varied were performed, and the estimated ICP (nICP) was compared with reference (invasive) ICP measurements. The ICP estimated by the nICP monitor was highly correlated with reference ICP measurements (Pearson’s correlation coefficient between 0.95 and 0.98). The nICP monitor also showed a low Root Mean Square Error from the reference ICP measure (3.12, 1.48, and 1.45 mmHg). Analysis of agreement by Bland and Altman also revealed good agreement between the two techniques. The optical nICP monitor was able to estimate the ICP non-invasively from an in vitro model simulating intracranial hypertension. The non-invasive ICP monitor showed very promising results which can set the base for further investigations. This work contributes significantly to the quest for non-invasive ICP monitoring in Traumatic Brain Injury (TBI), and paves the way for further research in this field.

Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://creativecommons.org/licenses/by/4.0/

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Signals

Keywords

Kyriacou BibliographyBrain Monitoring Icp

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG). Method: Method.SetupSignals: PPGSignalsPPGMethodMethod

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Abay, T. Y., Phillips, J. P., Uff, C., Roldan, M., & Kyriacou, P. A. (2022). In Vitro Evaluation of a Non-Invasive Photoplethysmography Based Intracranial Pressure Sensor. Applied Sciences. https://doi.org/10.3390/app13010534
Abay, Tomas Y., Justin P. Phillips, Christopher Uff, Maria Roldan, and Panicos A. Kyriacou. 2022. "In Vitro Evaluation of a Non-Invasive Photoplethysmography Based Intracranial Pressure Sensor." Applied Sciences. https://doi.org/10.3390/app13010534.
Abay, T. Y., Phillips, J. P., Uff, C., Roldan, M. and Kyriacou, P. A. (2022) 'In Vitro Evaluation of a Non-Invasive Photoplethysmography Based Intracranial Pressure Sensor', Applied Sciences. doi: 10.3390/app13010534.
Abay, Tomas Y., et al. "In Vitro Evaluation of a Non-Invasive Photoplethysmography Based Intracranial Pressure Sensor." Applied Sciences, 2022. https://doi.org/10.3390/app13010534.
T. Y. Abay, J. P. Phillips, C. Uff, M. Roldan, and P. A. Kyriacou, "In Vitro Evaluation of a Non-Invasive Photoplethysmography Based Intracranial Pressure Sensor," Applied Sciences, 2022, doi: 10.3390/app13010534.
@article{abay2022vitro,
  title         = {{In Vitro Evaluation of a Non-Invasive Photoplethysmography Based Intracranial Pressure Sensor}},
  author        = {Abay, Tomas Y. and Phillips, Justin P. and Uff, Christopher and Roldan, Maria and Kyriacou, Panicos A.},
  year          = {2022},
  journal       = {Applied Sciences},
  doi           = {10.3390/app13010534}
}

Added to the ChatPPG research hub