Paperdoi:10.1016/j.burns.2004.10.025

Evaluation of oesophageal reflectance pulse oximetry in major burns patients

Venue listed2005Added Nov 5, 2026

  • SP
  • PK
  • SK
  • SF
  • RE
  • RL
  • DJ

Authors: S.K. Pal, P.A. Kyriacou, S. Kumaran, S. Fadheel, R. Emamdee, R.M. Langford, D.P. Jones

Burns

TL;DR

Pulse oximetry is being used in everyday clinical practice in anaesthesia utilising a peripheral probe. However, it may be unreliable in certain clinical situations such as hypothermia, hypovolemia, vasoconstriction and decreased cardiac output. Similar situations occur in burns patients and, more importantly, burns to extremities which limit the sites available for measurement of peripheral oxygen saturation (SpO(2)).

  • Published in Burns (2005).

Abstract

Pulse oximetry is being used in everyday clinical practice in anaesthesia utilising a peripheral probe. However, it may be unreliable in certain clinical situations such as hypothermia, hypovolemia, vasoconstriction and decreased cardiac output. Similar situations occur in burns patients and, more importantly, burns to extremities which limit the sites available for measurement of peripheral oxygen saturation (SpO(2)). To overcome these limitations, the oesophagus has been investigated as an alternative measurement site, as perfusion may be preferentially preserved centrally. A miniaturised reflectance oesophageal saturation (SpO(2)) probe has been constructed utilising infrared and red photodiodes and a photodetector. Our study was aimed at evaluating the reliability of oesophageal pulse oximetry in major burns patients. Seven adult patients (five males, two females) were studied. They were sedated and ventilated as part of their routine care. Measurable photoplethysmographic (PPG) traces and SpO(2) values were obtained in the oesophagus of all patients at a mean depth of 15.6+/-1.8 cm (measured from the lips). It was found that the oesophageal pulse oximeter results were in good agreement with oxygen saturation measurements obtained by a CO-oximeter. The mean (+/-S.D.) of the differences between the oesophageal oxygen saturation results and those from CO-oximetry was 0.50+/-0.69%. A Bland and Altman analysis showed that the bias and the limits of agreement between the oesophageal and commercial toe pulse oximeters were 0.4% and -3.6% to 4.6%, respectively. This study suggests that the oesophagus can be used as an alternative site for monitoring arterial blood oxygen saturation by pulse oximetry in burned patients.

Verbatim from Europe PMC (opens in a new tab) · Abstract via Europe PMC; © the authors/publisher

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

SpO₂

Signals

Keywords

Kyriacou BibliographyPulse Oximetry Oxygenation

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG). Method: Method. Target: SpO₂.SetupSignals: PPGSignalsPPGMethodMethodTargetSpO₂O₂SpO₂

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Pal, S., Kyriacou, P., Kumaran, S., Fadheel, S., Emamdee, R., Langford, R., & Jones, D. (2005). Evaluation of oesophageal reflectance pulse oximetry in major burns patients. Burns. https://doi.org/10.1016/j.burns.2004.10.025
Pal, S.K., P.A. Kyriacou, S. Kumaran, S. Fadheel, R. Emamdee, R.M. Langford, and D.P. Jones. 2005. "Evaluation of oesophageal reflectance pulse oximetry in major burns patients." Burns. https://doi.org/10.1016/j.burns.2004.10.025.
Pal, S., Kyriacou, P., Kumaran, S., Fadheel, S., Emamdee, R., Langford, R. and Jones, D. (2005) 'Evaluation of oesophageal reflectance pulse oximetry in major burns patients', Burns. doi: 10.1016/j.burns.2004.10.025.
Pal, S.K., et al. "Evaluation of oesophageal reflectance pulse oximetry in major burns patients." Burns, 2005. https://doi.org/10.1016/j.burns.2004.10.025.
S. Pal et al., "Evaluation of oesophageal reflectance pulse oximetry in major burns patients," Burns, 2005, doi: 10.1016/j.burns.2004.10.025.
@article{pal2005evaluation,
  title         = {{Evaluation of oesophageal reflectance pulse oximetry in major burns patients}},
  author        = {Pal, S.K. and Kyriacou, P.A. and Kumaran, S. and Fadheel, S. and Emamdee, R. and Langford, R.M. and Jones, D.P.},
  year          = {2005},
  journal       = {Burns},
  doi           = {10.1016/j.burns.2004.10.025}
}

Added to the ChatPPG research hub