Paperdoi:10.1109/embc.2012.6346255

Photoplethysmographic and SpO<inf>2</inf> readings from the neonatal anterior fontanelle: A case Study

Venue listed2012Added Nov 5, 2026

  • JM
  • PK
  • MH
  • AP

Authors: J. M. May, P. A. Kyriacou, M. Honsel, A. J. Petros

2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society

TL;DR

There is a need for a more reliable, non-invasive and alternative measurement site for the monitoring of arterial blood oxygen saturation in critically ill neonates when peripheral perfusion is poor. The anterior fontanelle, a unique anatomical feature of the neonate, has been presented as an alternative site for the estimation of arterial blood oxygen saturation (SpO(2)). A new fontanelle photoplethysmographic sensor and processing system has been developed to investigate fontanelle photoplethysmographic (PPG) signals and estimate SpO(2) values at this anatomical location.

  • Published in 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society (2012).

Abstract

There is a need for a more reliable, non-invasive and alternative measurement site for the monitoring of arterial blood oxygen saturation in critically ill neonates when peripheral perfusion is poor. The anterior fontanelle, a unique anatomical feature of the neonate, has been presented as an alternative site for the estimation of arterial blood oxygen saturation (SpO(2)). A new fontanelle photoplethysmographic sensor and processing system has been developed to investigate fontanelle photoplethysmographic (PPG) signals and estimate SpO(2) values at this anatomical location. Preliminary clinical trials have shown that good quality PPG signals with large amplitudes and high signal to noise ratio can be obtained from the neonatal fontanelle. The estimation of SpO(2) values from the fontanelle were in broad agreement with a commercial foot pulse oximeter.

Verbatim from Europe PMC (opens in a new tab) · Abstract via Europe PMC; © the authors/publisher

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

SpO₂

Signals

Keywords

Kyriacou BibliographyPulse Oximetry Oxygenation

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG). Method: Method. Target: SpO₂.SetupSignals: PPGSignalsPPGMethodMethodTargetSpO₂O₂SpO₂

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

May, J. M., Kyriacou, P. A., Honsel, M., & Petros, A. J. (2012). Photoplethysmographic and SpO<inf>2</inf> readings from the neonatal anterior fontanelle: A case Study. 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society. https://doi.org/10.1109/embc.2012.6346255
May, J. M., P. A. Kyriacou, M. Honsel, and A. J. Petros. 2012. "Photoplethysmographic and SpO<inf>2</inf> readings from the neonatal anterior fontanelle: A case Study." 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society. https://doi.org/10.1109/embc.2012.6346255.
May, J. M., Kyriacou, P. A., Honsel, M. and Petros, A. J. (2012) 'Photoplethysmographic and SpO<inf>2</inf> readings from the neonatal anterior fontanelle: A case Study', 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society. doi: 10.1109/embc.2012.6346255.
May, J. M., et al. "Photoplethysmographic and SpO<inf>2</inf> readings from the neonatal anterior fontanelle: A case Study." 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2012. https://doi.org/10.1109/embc.2012.6346255.
J. M. May, P. A. Kyriacou, M. Honsel, and A. J. Petros, "Photoplethysmographic and SpO<inf>2</inf> readings from the neonatal anterior fontanelle: A case Study," 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2012, doi: 10.1109/embc.2012.6346255.
@inproceedings{may2012photoplethysmographic,
  title         = {{Photoplethysmographic and SpO<inf>2</inf> readings from the neonatal anterior fontanelle: A case Study}},
  author        = {May, J. M. and Kyriacou, P. A. and Honsel, M. and Petros, A. J.},
  year          = {2012},
  booktitle     = {2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society},
  doi           = {10.1109/embc.2012.6346255}
}

Added to the ChatPPG research hub