A novel fontanelle probe for sensing oxygen saturation in the neonate

Venue listed2017Added Nov 5, 2026

  • JM
  • PK
  • AP

Authors: J M May, P A Kyriacou, A J Petros

Biomedical Physics & Engineering Express

TL;DR

Monitoring of blood oxygen saturation (SpO 2 ) of the neonate is essential to the quality of health care provided on a neonatal intensive care unit (NICU). Current sensors are usually placed at the hand or foot, which are dependent of a peripheral blood supply. When the peripheral blood circulation of neonates is compromised conventional peripheral pulse oximeters, in many cases, fail to operate accurately or at all.

  • Published in Biomedical Physics & Engineering Express (2017).
  • DOI 10.1088/2057-1976/aa5946.

Abstract

Monitoring of blood oxygen saturation (SpO 2 ) of the neonate is essential to the quality of health care provided on a neonatal intensive care unit (NICU). Current sensors are usually placed at the hand or foot, which are dependent of a peripheral blood supply. When the peripheral blood circulation of neonates is compromised conventional peripheral pulse oximeters, in many cases, fail to operate accurately or at all. A new reflectance anterior fontanelle (ANTF) SpO 2 sensor and instrumentation has been developed to investigate SpO 2 s from the neonatal fontanelle. The hypothesis is that perfusion at a central site should be preserved at times of compromised peripheral circulation. Fifteen neonates on an NICU (9 male, 6 female) with a median age of 7 d (IQR = 41.5 d) were selected for monitoring. ANTF photoplethysmographic (PPG) signals were monitored for a maximum period of 2 h. The developed system and custom made sensors were successful at acquiring good quality signals at both wavelengths necessary for pulse oximetry calculations. ANTF SpO 2 s, estimated from the acquired PPGs, were in broad agreement with SpO 2 s obtained from the commercial foot pulse oximeter. A Bland and Altman analysis of the differences between SpO 2 s from the fontanelle PPG sensor and the commercial device show a relatively small mean difference (

d = ± 2.2 %

), but with a wide variation (

2 s = ± 17.4 %

) this observation may be due to the varied levels of ill health patients and is backed up by comparing the commercial device SpO 2 readings at the same moment a blood gas sample was taken (

d = 4.8 % , 2 s = ± 15.8 %

).

Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://publishingsupport.iopscience.iop.org/iop-standard/v1

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

SpO₂

Keywords

Kyriacou BibliographyPulse Oximetry Oxygenation

How it works

How Method works: Setup → Method → TargetMethod: Method. Target: SpO₂.MethodMethodTargetSpO₂O₂SpO₂

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

May, J. M., Kyriacou, P. A., & Petros, A. J. (2017). A novel fontanelle probe for sensing oxygen saturation in the neonate. Biomedical Physics & Engineering Express. https://doi.org/10.1088/2057-1976/aa5946
May, J M, P A Kyriacou, and A J Petros. 2017. "A novel fontanelle probe for sensing oxygen saturation in the neonate." Biomedical Physics & Engineering Express. https://doi.org/10.1088/2057-1976/aa5946.
May, J. M., Kyriacou, P. A. and Petros, A. J. (2017) 'A novel fontanelle probe for sensing oxygen saturation in the neonate', Biomedical Physics & Engineering Express. doi: 10.1088/2057-1976/aa5946.
May, J M, et al. "A novel fontanelle probe for sensing oxygen saturation in the neonate." Biomedical Physics & Engineering Express, 2017. https://doi.org/10.1088/2057-1976/aa5946.
J. M. May, P. A. Kyriacou, and A. J. Petros, "A novel fontanelle probe for sensing oxygen saturation in the neonate," Biomedical Physics & Engineering Express, 2017, doi: 10.1088/2057-1976/aa5946.
@article{may2017novel,
  title         = {{A novel fontanelle probe for sensing oxygen saturation in the neonate}},
  author        = {May, J. M. and Kyriacou, P. A. and Petros, A. J.},
  year          = {2017},
  journal       = {Biomedical Physics \& Engineering Express},
  doi           = {10.1088/2057-1976/aa5946}
}

Added to the ChatPPG research hub