Paperdoi:10.3390/bioengineering13111269

Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test–Retest Experiment Among Healthy Young Men

Venue listed2026Added Oct 29, 2026

  • TL
  • JB
  • JK

Authors: Tae-Hyung Lee, Jong-Hee Bae, Jae-Myun Ko

Bioengineering 13(11):1269 (2026-10-27)

TL;DR

60 healthy young men, three identical resistance-exercise sessions five days apart. Devices: Apple Watch Series 7, Galaxy Watch 4, Fitbit Charge 5, Garmin Vivosmart 5 vs ECG criterion; displayed HR recorded before each set and 5, 30 and 60 s after.

  • Form factor: wrist

Abstract

Smartwatch heart-rate (HR) displays may be inaccurate during rapid post-exercise transitions. This study evaluated timepoint-specific smartwatch–electrocardiography (ECG) error and its reproducibility across three resistance-exercise sessions. Sixty healthy young men completed three identical sessions five days apart using the Apple Watch Series 7, Galaxy Watch 4, Fitbit Charge 5, and Garmin Vivosmart 5. Displayed HR was recorded before each set and at 5, 30, and 60 s afterward. The primary outcome was participant-level session-mean error across eight 5 s observations. Mean absolute error increased in the range of 1.94–2.32 bpm before exercise to 8.71–25.43 bpm at 5 s and decreased to a range of 3.58–10.06 bpm at 30 s. Session-mean 5 s error intraclass correlation coefficients were 0.892, 0.771, 0.879, and 0.834 for Apple, Galaxy, Fitbit, and Garmin, respectively, indicating moderate-to-good through good-to-excellent reliability. Session effects were not statistically significant. Raw single-observation limits of agreement were substantially wider than session-mean limits, and all devices exhibited proportional bias. Participant-level variance estimates remained similar in exploratory models including ECG-response and wrist-position covariates. Therefore, early post-set error contained a reproducible session-level component, but individual 5 s readings were not interchangeable with ECG measurements. These findings support cautious interpretation of post-set HR displays without establishing a stable participant-specific error trait.

Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://creativecommons.org/licenses/by/4.0/

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

Heart rate

Signals

Keywords

HR ValidationResistance ExerciseMotionTest RetestFitbitGarminECG Reference

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG, ECG); Form factor (wrist). Method: Method. Target: Heart rate.SetupSignals: PPG, ECGSignalsPPG, ECGForm factor: wristForm factorwristMethodMethodTargetHeart rateHeart rate

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Lee, T.-H., Bae, J.-H., & Ko, J.-M. (2026). Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test–Retest Experiment Among Healthy Young Men. Bioengineering, 13(11), 1269. https://doi.org/10.3390/bioengineering13111269
Lee, Tae-Hyung, Jong-Hee Bae, and Jae-Myun Ko. 2026. "Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test–Retest Experiment Among Healthy Young Men." Bioengineering 13 (11): 1269. https://doi.org/10.3390/bioengineering13111269.
Lee, T.-H., Bae, J.-H. and Ko, J.-M. (2026) 'Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test–Retest Experiment Among Healthy Young Men', Bioengineering, 13(11), p. 1269. doi: 10.3390/bioengineering13111269.
Lee, Tae-Hyung, et al. "Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test–Retest Experiment Among Healthy Young Men." Bioengineering, vol. 13, no. 11, 2026, p. 1269. https://doi.org/10.3390/bioengineering13111269.
T.-H. Lee, J.-H. Bae, and J.-M. Ko, "Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test–Retest Experiment Among Healthy Young Men," Bioengineering, vol. 13, no. 11, p. 1269, 2026, doi: 10.3390/bioengineering13111269.
@article{lee2026assessing,
  title         = {{Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test–Retest Experiment Among Healthy Young Men}},
  author        = {Lee, Tae-Hyung and Bae, Jong-Hee and Ko, Jae-Myun},
  year          = {2026},
  journal       = {Bioengineering},
  volume        = {13},
  number        = {11},
  pages         = {1269},
  doi           = {10.3390/bioengineering13111269}
}

Added to the ChatPPG research hub