FDA 510(k) · Biobeat

K241066: BB-613-BPM (Biobeat Technologies , Ltd.)

Biobeat Technologies , Ltd. received FDA 510(k) K241066 for BB-613-BPM, with a decision date of Jan 14, 2025 (substantially equivalent).

Prescription or OTC: Unknown (no FDA document we parsed states it).

FDA record for K241066FDA summary or statement (PDF)

FDA facts

Official fields for K241066
FDA numberK241066
Pathway510(k)
ApplicantBiobeat Technologies , Ltd.
Device nameBB-613-BPM
Product codeDXN: System, Measurement, Blood-Pressure, Non-Invasive
Regulation (21 CFR)870.1130
Device classClass 2
Date receivedApr 18, 2024
Decision dateJan 14, 2025
DecisionSubstantially Equivalent (SESE)
510(k) typeTraditional

Source: openFDA device 510(k) data, data through Oct 26, 2026.

Indications for use parsed from FDA documents

The Biobeat BB-613-BPM is a noninvasive blood pressure monitor which provides a notification to the user every 15 minutes over a 24 hour period for user- initiated measurements. It includes a device and a platform (BPM) for use in adults by healthcare professionals, for collection of pulse rate and tracking changes in blood pressure based on Pulse Wave Transit Time (PWTT) in hospitals, clinics, long-term care and at home. The BB-613-BPM tracks changes in blood pressure and requires calibration using an FDA-cleared oscillometric blood pressure monitor. The data from the Biobeat BB-613-BPM can be accessed through the Biobeat platform for review by a healthcare professional or can be downloaded as a report. The device is not intended to be used on critical care patients.

Extracted automatically from FDA's 510(k) summary. It can contain extraction errors and is not a verified quote. Read the wording in the FDA PDF before relying on it.

Predicates parsed from FDA documents

FDA numbers found next to the word "predicate" in the summary. The list can include reference devices; check the PDF for which device is the primary predicate.

  • K190792 BB-613WP (Biobeat)
  • K222010 Biobeat Platform-2 and BB-613WP Patch (Biobeat)

ChatPPG interpretation

A 510(k) decision of "substantially equivalent" means FDA found the device as safe and effective as a legally marketed predicate. FDA's word for this is cleared, not approved. The clearance covers the device and indications in this submission; it does not extend to other features or versions of the same product that were not part of it. See FDA cleared vs approved.

Product code DXN is "System, Measurement, Blood-Pressure, Non-Invasive", class 2, 21 CFR 870.1130.

Our labels for filtering: claim Blood pressure, sensor Other. These are heuristics, not FDA fields.