Paperdoi:10.2147/NSS.S306808
It is All in the Wrist: Wearable Sleep Staging in a Clinical Population versus Reference Polysomnography
Nature and Science of Sleep
TL;DR
Wrist reflective PPG + accelerometry sleep staging trained/validated against PSG in an independent clinical population (244 adults + 48 children/adolescents, ages 3–82) spanning diverse sleep disorders. Four-class average Cohen’s κ 0.62, accuracy 76.4% (children/adolescents κ 0.66 / acc 77.9%). Performance higher in non-REM parasomnias (κ 0.69) and lower in REM parasomnias (κ 0.55); weak negative correlation with age.
- Form factor: wrist
Abstract
PurposeThere is great interest in unobtrusive long-term sleep measurements using wearable devices based on reflective photoplethysmography (PPG). Unfortunately, consumer devices are not validated in patient populations and therefore not suitable for clinical use. Several sleep staging algorithms have been developed and validated based on ECG-signals. However, translation from these techniques to data derived by wearable PPG is not trivial, and requires the differences between sensing modalities to be integrated in the algorithm, or having the model trained directly with data obtained with the target sensor. Either way, validation of PPG-based sleep staging algorithms requires a large dataset containing both gold standard measurements and PPG-sensor in the applicable clinical population. Here, we take these important steps towards unobtrusive, long-term sleep monitoring.MethodsWe developed and trained an algorithm based on wrist-worn PPG and accelerometry. The method was validated against reference polysomnography in an independent clinical population comprising 244 adults and 48 children (age: 3 to 82 years) with a wide variety of sleep disorders.ResultsThe classifier achieved substantial agreement on four-class sleep staging with an average Cohen's kappa of 0.62 and accuracy of 76.4%. For children/adolescents, it achieved even higher agreement with an average kappa of 0.66 and accuracy of 77.9%. Performance was significantly higher in non-REM parasomnias (kappa = 0.69, accuracy = 80.1%) and significantly lower in REM parasomnias (kappa = 0.55, accuracy = 72.3%). A weak correlation was found between age and kappa (ρ = -0.30, pρ = -0.22, pConclusionThis study shows the feasibility of automatic wearable sleep staging in patients with a broad variety of sleep disorders and a wide age range. Results demonstrate the potential for ambulatory long-term monitoring of clinical populations, which may improve diagnosis, estimation of severity and follow up in both sleep medicine and research.
Verbatim from Europe PMC (opens in a new tab) · Licence: cc by-nc
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Tasks
Signals
Keywords
How it works
Cite this paper
APA
Wulterkens, B. M., Fonseca, P., Hermans, L. W., Ross, M., Cerny, A., Anderer, P., Long, X., van Dijk, J. P., Vandenbussche, N., Pillen, S., et al. (2021). It is All in the Wrist: Wearable Sleep Staging in a Clinical Population versus Reference Polysomnography. Nature and Science of Sleep. https://doi.org/10.2147/NSS.S306808
BibTeX
@article{wulterkens2021wrist,
title = {{It is All in the Wrist: Wearable Sleep Staging in a Clinical Population versus Reference Polysomnography}},
author = {Wulterkens, Bernice M. and Fonseca, Pedro and Hermans, Lieke W.A. and Ross, Marco and Cerny, Andreas and Anderer, Peter and Long, Xi and van Dijk, Johannes P. and Vandenbussche, Nele and Pillen, Sigrid and others},
year = {2021},
journal = {Nature and Science of Sleep},
doi = {10.2147/NSS.S306808}
}Built from the hub record. Check it against the publisher before submitting.
Added to the ChatPPG research hub