Paperdoi:10.5646/ch.2026.32.e41

Real-world performance of a ring-type cuffless blood pressure device compared with ambulatory blood pressure monitoring

Venue listed2026Added Oct 12, 2026

  • KH
  • KB
  • SJ
  • SH
  • K

Authors: Kim HJ, Kim BS, Shin JH, Shin J (Hanyang University Guri / Seoul Hospitals, Korea)

Clinical Hypertension 32:e41 (2026-10-01)

TL;DR

Real-world observational study, n = 154 adults (mean age 62.1 ± 14.2 y, 41.6 % men), simultaneous 24-h cuff ABPM and CART BP pro. Ring period summaries read lower SBP than ABPM: −7.1 mmHg 24-h, −7.8 mmHg daytime, −1.7 mmHg nighttime. DBP ran higher on the ring, especially at night (+4.8 mmHg).

  • Form factor: ring

Abstract

BackgroundCuffless blood pressure (BP) monitoring devices are convenient alternatives to conventional ambulatory blood pressure monitoring (ABPM). We evaluated the real-world agreement between a ring-type cuffless BP device (continuous ambulatory ring-type blood pressure monitor, CART BP pro) and ABPM and identified the clinical and physiological factors associated with measurement discrepancies.MethodsThis real-world observational study included 154 adults (mean age, 62.1 ± 14.2 years; 41.6% men) who underwent simultaneous 24-hour ABPM and CART BP pro after exclusion of predefined clinical conditions and insufficient recordings. CART BP pro measurements were calibrated at baseline to ABPM in the contralateral arm. BP values from the 2 devices were compared for 24-hour, daytime, and nighttime periods, as well as for the classification of nocturnal hypertension and dipping status. Predictors of close systolic agreement (|24-hour systolic blood pressure [SBP] difference| ≤ 5 mmHg) were identified using multivariable logistic regression.ResultsCART BP pro-derived period summaries were lower than ABPM for systolic BP (-7.1 mmHg for 24-hour SBP) and showed a modest positive bias for diastolic BP. Only 33.1% of measurements met the ≤ 5 mmHg accuracy criterion. Daytime SBP showed a similar bias (-7.8 mmHg), whereas nighttime SBP differences were smaller (-1.7 mmHg). By contrast, diastolic BP was consistently higher with the CART BP pro, particularly at night (+4.8 mmHg). Older age, male sex, lower body mass index, and higher heart rate independently predicted a closer systolic agreement. The agreement for nocturnal hypertension was moderate (κ = 0.516), with an overall concordance of 76.0%. However, dipping concordance was limited, with the CART BP pro showing blunted nocturnal dipping compared with ABPM (mean, 3.7% vs. 7.8%).ConclusionsThis exploratory, real-world study found that CART BP pro-derived period summaries showed lower SBP values and smaller nocturnal BP declines than ABPM. Because period-level aggregates were compared rather than simultaneous point-to-point measurements, these findings should not be interpreted as a formal accuracy validation of the device. Further studies are needed to establish device-specific diagnostic thresholds and prospectively validate the clinical interpretation of CART BP pro-derived period summaries for hypertension diagnosis and circadian BP phenotyping.

Verbatim from Europe PMC (opens in a new tab) · Abstract via Europe PMC; © the authors/publisher

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

Blood pressure

Signals

Keywords

cuffless-BPring24h-ABPMnocturnal-hypertensiondippingCART-BP-proSky-Labsreal-worldKorea

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG); Form factor (ring). Method: Method. Target: Blood pressure.SetupSignals: PPGSignalsPPGForm factor: ringForm factorringMethodMethodTargetBlood pressureBlood pressure

Cite this paper

APA

Kim, H. J., Kim, B. S., Shin, J. H., & Shin, J. (2026). Real-world performance of a ring-type cuffless blood pressure device compared with ambulatory blood pressure monitoring. Clinical Hypertension 32:e41. https://doi.org/10.5646/ch.2026.32.e41

BibTeX

@article{kim2026realworld,
  title         = {{Real-world performance of a ring-type cuffless blood pressure device compared with ambulatory blood pressure monitoring}},
  author        = {Kim, H. J. and Kim, B. S. and Shin, J. H. and Shin, J.},
  year          = {2026},
  journal       = {Clinical Hypertension 32:e41},
  doi           = {10.5646/ch.2026.32.e41}
}

Built from the hub record. Check it against the publisher before submitting.

Added to the ChatPPG research hub