Clinical Validation of Photoplethysmography-Based Finger Cuffless Blood Pressure Monitoring Following European Society of Hypertension Recommendations
Journal of the American Heart Association 15(15):e050501 (2026; epub 2026-07-25)
TL;DR
Phase 1 (n = 768): validation vs invasive BP, office BP and 24-h ABPM across the 7 ESH-recommended scenarios (invasive, static, awake/asleep, device position, treatment, exercise, recalibration). Phase 2 (n = 405 hypertensive patients): PPG- vs ABPM-derived mean SBP, SBP time-in-target range (TTR) and variability vs hypertension-mediated organ damage and CV complications.
- Form factor: finger
Abstract
Background
Cuffless blood pressure (BP) monitoring remains challenging for clinical hypertension management. Although photoplethysmography offers promising solutions for cuffless BP measurement, its clinical validation is limited. This study evaluated the accuracy and clinical utility of photoplethysmography‐based finger cuffless BP monitoring following European Society of Hypertension recommendations.
Methods
The 2‐phase study enrolled 1173 participants. In phase 1 (n=768), a novel photoplethysmography‐based finger cuffless BP monitor was validated against invasive BP, office BP, and 24‐hour ambulatory BP monitoring across 7 European Society of Hypertension–recommended scenarios. In phase 2 (n=405 hypertensive patients), the clinical utility of photoplethysmography‐ versus ambulatory BP monitoring–derived mean systolic BP (SBP), SBP time‐in‐target range, and variability was assessed by examining their associations with hypertension‐mediated organ damage and cardiovascular complications, including arterial stiffness, endothelial dysfunction, coronary heart disease, and stroke.
Results
The photoplethysmography‐based finger cuffless BP monitor fulfilled all validation criteria, with mean differences ≤5 mm Hg and standard deviations ≤8 mm Hg of both systolic and diastolic BP across 7 scenarios, including invasive, static, awake/asleep, device position, treatment, exercise, and recalibration test. Both ambulatory BP monitoring–derived time‐in‐target range and photoplethysmography‐derived mean SBP and time‐in‐target range were significantly associated with endothelial injury and arterial stiffness. Notably, photoplethysmography‐derived mean SBP and time‐in‐target range, but not ambulatory BP monitoring–derived metrics, showed a significant association with coronary artery disease. Furthermore, photoplethysmography‐derived SBP variability was significantly and positively associated with endothelial injury.
Conclusions
Our data demonstrated that the novel photoplethysmography‐based finger cuffless BP monitor reached the criteria of clinical application following European Society of Hypertension recommendations and provided additional clinical implications for the evaluation of hypertension‐mediated organ damage and cardiovascular complications.
Verbatim from the publisher's DOI record (opens in a new tab) · Abstract as deposited by the publisher with Crossref; © the authors/publisher
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Tasks
Signals
Keywords
How it works
Cite this paper
APA
Qiu, Y., Zhou, Z., Zhang, J., He, J., Xu, S., Zhang, X., Sun, R., Su, C., Xia, W., & Tao, J. (2026). Clinical Validation of Photoplethysmography-Based Finger Cuffless Blood Pressure Monitoring Following European Society of Hypertension Recommendations. Journal of the American Heart Association 15:e050501. https://doi.org/10.1161/JAHA.126.050501
BibTeX
@article{qiu2026clinical,
title = {{Clinical Validation of Photoplethysmography-Based Finger Cuffless Blood Pressure Monitoring Following European Society of Hypertension Recommendations}},
author = {Qiu, Yumin and Zhou, Zhe and Zhang, Jianning and He, Jiang and Xu, Shiyue and Zhang, Xiaoyu and Sun, Ruochen and Su, Chen and Xia, Wenhao and Tao, Jun},
year = {2026},
journal = {Journal of the American Heart Association 15:e050501},
doi = {10.1161/JAHA.126.050501}
}Built from the hub record. Check it against the publisher before submitting.
Added to the ChatPPG research hub