The effects of optical sensor-tissue separation in endocavitary photoplethysmography
Physiological Measurement
TL;DR
Objective : Intestinal anastomotic failure that occurs mainly due to ischaemia is a serious risk in colorectal cancer patients undergoing surgery. Surgeons continue to rely on subjective methods such as visual inspection to assess intestinal viability during surgery and there are no clinical tools to directly monitor viability postoperatively. A dual-wavelength reflectance optical sensor has been developed for continuous and dynamic monitoring of intestinal viability via the intestinal lumen.
- Published in Physiological Measurement (2018).
Abstract
Objective : Intestinal anastomotic failure that occurs mainly due to ischaemia is a serious risk in colorectal cancer patients undergoing surgery. Surgeons continue to rely on subjective methods such as visual inspection to assess intestinal viability during surgery and there are no clinical tools to directly monitor viability postoperatively. A dual-wavelength reflectance optical sensor has been developed for continuous and dynamic monitoring of intestinal viability via the intestinal lumen. Maintaining direct contact between the sensor and the inner intestinal wall can be difficult in an intraluminal design, therefore impacting on signal acquisition and quality. This paper investigates the effect of direct contact versus variable distances between the sensor and the tissue surface of the buccal mucosa as a surrogate. Approach : The in vivo study involved 20 healthy volunteers to measure the effect of optical sensor-tissue distances on the ability to acquire photoplethysmography signals and their quality. Signals were acquired from the buccal mucosa at five optical sensor-tissue distances. Main results : Distances between 0 mm (contact) to 5 mm were the most optimal, producing signals of high quality and signal-to-noise ratio, resulting in reliable estimations of the blood oxygen saturation. Distances exceeding 5 mm compromised the acquired signals, and were of poor quality, thereby unreliably estimating the blood oxygen saturation. Significance : The developed optical sensor proved to be reliable for acquiring photoplethysmography signals for cases where distances between the optical sensor-tissue may arise during the assessment of intraluminal intestinal viability.
Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://publishingsupport.iopscience.iop.org/iop-standard/v1
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Signals
Keywords
How it works
Cite this paper
Patel, Z., Thaha, M. A., & Kyriacou, P. A. (2018). The effects of optical sensor-tissue separation in endocavitary photoplethysmography. Physiological Measurement. https://doi.org/10.1088/1361-6579/aacc1d
Patel, Zaibaa, Mohamed A Thaha, and Panayiotis A Kyriacou. 2018. "The effects of optical sensor-tissue separation in endocavitary photoplethysmography." Physiological Measurement. https://doi.org/10.1088/1361-6579/aacc1d.
Patel, Z., Thaha, M. A. and Kyriacou, P. A. (2018) 'The effects of optical sensor-tissue separation in endocavitary photoplethysmography', Physiological Measurement. doi: 10.1088/1361-6579/aacc1d.
Patel, Zaibaa, et al. "The effects of optical sensor-tissue separation in endocavitary photoplethysmography." Physiological Measurement, 2018. https://doi.org/10.1088/1361-6579/aacc1d.
Z. Patel, M. A. Thaha, and P. A. Kyriacou, "The effects of optical sensor-tissue separation in endocavitary photoplethysmography," Physiological Measurement, 2018, doi: 10.1088/1361-6579/aacc1d.
@article{patel2018effects,
title = {{The effects of optical sensor-tissue separation in endocavitary photoplethysmography}},
author = {Patel, Zaibaa and Thaha, Mohamed A. and Kyriacou, Panayiotis A.},
year = {2018},
journal = {Physiological Measurement},
doi = {10.1088/1361-6579/aacc1d}
}Added to the ChatPPG research hub