AI-Enhanced Smartwatch AHI Estimation and AI-Scored Polysomnography for Obstructive Sleep Apnea: Real-World Validation

Venue listed2025Added Oct 29, 2026

  • DK
  • JH
  • HJ
  • YS
  • CL
  • GR
  • SH
  • HK
  • +1

Authors: Donghyeok Kim, Jeong Yup Han, Hyunjun Jung, Da Yeun Song, Changhee Lee, Gwanghui Ryu, Sang Duk Hong, Hyo-Yeol Kim, Yong Gi Jung (Samsung Medical Center; Han and Jung are with Samsung Electronics Mobile eXperience)

Nature and Science of Sleep, 2025

TL;DR

Real-world validation of a smartwatch AHI algorithm against Level 1 polysomnography in Korean adults (53 of 90 with enough valid watch data). Watch eAHI agreed with expert-scored AHI at ICC 0.82, and detected moderate-to-severe OSA with 92.3% sensitivity and 92.6% specificity against AI-scored PSG, while underestimating AHI in mild OSA.

  • Form factor: wrist
Caveats
  • Vs expert PSG (0.87 vs AI-scored PSG); 41% excluded for invalid data; Samsung co-authors.
  • Reference is AI-scored PSG, not expert scoring; Samsung co-authors.

Abstract

ObjectiveThis study validated the accuracy of an artificial‑intelligence (AI) smartwatch algorithm that directly estimates the apnea-hypopnea index (AHI) by comparing its performance with AI-scored Level 1 polysomnography (PSG) in Korean adults. The model was trained in South‑American cohorts, allowing inter‑ethnic validation.MethodsA total of 90 adults underwent simultaneous Level 1 PSG and smartwatch recording. Fifty‑three datasets with ≥ 3 hours of valid watch data were analyzed. AHI values were obtained as follows: expert‑scored PSG (pAHI), AI‑scored PSG (aiAHI), and smartwatch output (eAHI). Agreement was assessed with Spearman correlation, intraclass correlation coefficients, and receiver‑operating‑characteristic curves.ResultseAHI correlated strongly with aiAHI (ρ = 0.88, ICC = 0.87) and pAHI (ρ = 0.85, ICC = 0.82). For detecting moderate‑to‑severe OSA (aiAHI ≥ 15 events/h), the smartwatch algorithm yielded 92.3% sensitivity, 92.6% specificity, and 92.5% overall accuracy. Bland-Altman analysis revealed systematic underestimation of actual AHI by the smartwatch, particularly in mild OSA.ConclusionThis study demonstrates that the evaluated smartwatch-based AHI estimation algorithm shows high concordance with PSG-derived values, particularly for the detection and classification of moderate to severe OSA. However, it should be noted that this smartwatch algorithm tends to underestimate the AHI of OSA due to limitations in scoring unit and recording duration calculation. These findings support the clinical utility of wearable technology as a practical and scalable tool for early identification and longitudinal monitoring of OSA in real-world environments, while highlighting the need for further optimization to accurately detect mild cases.

Verbatim from Europe PMC (opens in a new tab) · Licence: cc by-nc

Results on ChatPPG leaderboards · 3

All leaderboards
LeaderboardMethodResultRank
Samsung Medical Center cohort ICCSleep apnea (OSA) screeningSmartwatch eAHI algorithmCaveat: Vs expert PSG (0.87 vs AI-scored PSG); 41% excluded for invalid data; Samsung co-authors.0.82ICC, higher is betterUnranked
Samsung Medical Center cohort sensitivitySleep apnea (OSA) screeningSmartwatch eAHI algorithmCaveat: Reference is AI-scored PSG, not expert scoring; Samsung co-authors.92.3%sensitivity, higher is betterUnranked
Samsung Medical Center cohort specificitySleep apnea (OSA) screeningSmartwatch eAHI algorithmCaveat: Reference is AI-scored PSG, not expert scoring; Samsung co-authors.92.6%specificity, higher is betterUnranked

3 rows are credited to this paper. Ranks apply within each leaderboard only, as reported by the source.

Tasks & signals

Signals

Keywords

Sleep ApneaOSAeAHIWristVendor CoauthoredKorea

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG); Form factor (wrist). Method: Method. Target: Sleep apnea.SetupSignals: PPGSignalsPPGForm factor: wristForm factorwristMethodMethodTargetSleep apneaSleep apnea

Cite this paper

Cite this paper with the fields recorded here. Check it against the publisher before submitting.

Kim, D., Han, J. Y., Jung, H., Song, D. Y., Lee, C., Ryu, G., Hong, S. D., Kim, H.-Y., & Jung, Y. G. (2025). AI-Enhanced Smartwatch AHI Estimation and AI-Scored Polysomnography for Obstructive Sleep Apnea: Real-World Validation. Nature and Science of Sleep, 2025. https://doi.org/10.2147/NSS.S540460
Kim, Donghyeok, Jeong Yup Han, Hyunjun Jung, Da Yeun Song, Changhee Lee, Gwanghui Ryu, Sang Duk Hong, Hyo-Yeol Kim, and Yong Gi Jung. 2025. "AI-Enhanced Smartwatch AHI Estimation and AI-Scored Polysomnography for Obstructive Sleep Apnea: Real-World Validation." Nature and Science of Sleep, 2025. https://doi.org/10.2147/NSS.S540460.
Kim, D., Han, J. Y., Jung, H., Song, D. Y., Lee, C., Ryu, G., Hong, S. D., Kim, H.-Y. and Jung, Y. G. (2025) 'AI-Enhanced Smartwatch AHI Estimation and AI-Scored Polysomnography for Obstructive Sleep Apnea: Real-World Validation', Nature and Science of Sleep, 2025. doi: 10.2147/NSS.S540460.
Kim, Donghyeok, et al. "AI-Enhanced Smartwatch AHI Estimation and AI-Scored Polysomnography for Obstructive Sleep Apnea: Real-World Validation." Nature and Science of Sleep, 2025, 2025. https://doi.org/10.2147/NSS.S540460.
D. Kim et al., "AI-Enhanced Smartwatch AHI Estimation and AI-Scored Polysomnography for Obstructive Sleep Apnea: Real-World Validation," Nature and Science of Sleep, 2025, 2025, doi: 10.2147/NSS.S540460.
@article{kim2025aienhanced,
  title         = {{AI-Enhanced Smartwatch AHI Estimation and AI-Scored Polysomnography for Obstructive Sleep Apnea: Real-World Validation}},
  author        = {Kim, Donghyeok and Han, Jeong Yup and Jung, Hyunjun and Song, Da Yeun and Lee, Changhee and Ryu, Gwanghui and Hong, Sang Duk and Kim, Hyo-Yeol and Jung, Yong Gi},
  year          = {2025},
  journal       = {Nature and Science of Sleep, 2025},
  doi           = {10.2147/NSS.S540460}
}

Added to the ChatPPG research hub