Paperdoi:10.64898/2026.04.08.26350411
Apnea-hypopnea index estimation with wrist-worn photoplethysmography
medRxiv preprint
TL;DR
Preprint describing a neural network that detects respiratory events from wrist reflective PPG and PPG-derived sleep stages, pre-trained on finger PPG from 3,111 PSG recordings and fine-tuned on 969 concurrent wrist recordings. Estimated AHI agreed with PSG at ICC 0.87 on the development test split and 0.91 on an external hold-out set of 171 recordings.
- Form factor: wrist
- Preprint; development test split 0.87 (N=3,111 development incl. 969 wrist).
Abstract
Objective
Long term monitoring of obstructive sleep apnea (OSA) severity may be relevant for several clinical applications. We developed a method for estimating the apnea-hypopnea index (AHI) using wrist-worn, reflective photoplethysmography (PPG).
Approach
A neural network was developed to detect respiratory events using PPG and PPG-derived sleep stages as input. The development database encompassed retrospective data from three polysomnographic datasets (N=3111), including a dataset with concurrent reflective PPG recordings from a wrist-worn device (N=969). The model was pre-trained with (transmissive) finger-PPG signals from all overnight recordings and then fine-tuned to wrist-PPG characteristics using transfer learning. Validation was performed on the test portion of the development set and on a fourth, external hold-out dataset containing both wrist-PPG and PSG data (N=171). Performance was evaluated in terms of AHI estimation accuracy and OSA severity classification.
Main Results
The fine-tuned wrist-PPG model demonstrated strong agreement with the PSG-derived gold-standard AHI, achieving intra-class correlation coefficients of 0.87 in the test portion of the development set and 0.91 in the external hold-out validation set. Diagnostic performance was high, with accuracies above 80% for all severity thresholds.
Significance
The study highlights the potential of reflective PPG-based AHI estimation, achieving high estimation performance in comparison with PSG. These measurements can be performed with relatively comfortable sensors integrated in convenient wrist-worn wearables, enabling long-term assessment of sleep disordered breathing, both in a diagnostic phase, and during therapy follow-up.
Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: http://creativecommons.org/licenses/by-nc-nd/4.0/
Results on ChatPPG leaderboards · 1
All leaderboards| Leaderboard | Method | Result | Rank |
|---|---|---|---|
| Non-public external wrist+PSG set ICCSleep apnea (OSA) screening | Wrist PPG event detector, finger-to-wrist transferCaveat: Preprint; development test split 0.87 (N=3,111 development incl. 969 wrist). | 0.91ICC, higher is better | Unranked |
1 row is credited to this paper. Ranks apply within each leaderboard only, as reported by the source.
Tasks & signals
Tasks
Signals
Keywords
How it works
Cite this paper
Fonseca, P., Ross, M., van Meulen, F., Asin, J., van Gilst, M. M., & Overeem, S. (2026). Apnea-hypopnea index estimation with wrist-worn photoplethysmography. medRxiv preprint. https://doi.org/10.64898/2026.04.08.26350411
Fonseca, Pedro, Marco Ross, Fokke van Meulen, Jerryll Asin, Merel M. van Gilst, and Sebastiaan Overeem. 2026. "Apnea-hypopnea index estimation with wrist-worn photoplethysmography." medRxiv preprint. https://doi.org/10.64898/2026.04.08.26350411.
Fonseca, P., Ross, M., van Meulen, F., Asin, J., van Gilst, M. M. and Overeem, S. (2026) 'Apnea-hypopnea index estimation with wrist-worn photoplethysmography', medRxiv preprint. doi: 10.64898/2026.04.08.26350411.
Fonseca, Pedro, et al. "Apnea-hypopnea index estimation with wrist-worn photoplethysmography." medRxiv preprint, 2026. https://doi.org/10.64898/2026.04.08.26350411.
P. Fonseca, M. Ross, F. van Meulen, J. Asin, M. M. van Gilst, and S. Overeem, "Apnea-hypopnea index estimation with wrist-worn photoplethysmography," medRxiv preprint, 2026, doi: 10.64898/2026.04.08.26350411.
@misc{fonseca2026apneahypopnea,
title = {{Apnea-hypopnea index estimation with wrist-worn photoplethysmography}},
author = {Fonseca, Pedro and Ross, Marco and van Meulen, Fokke and Asin, Jerryll and van Gilst, Merel M. and Overeem, Sebastiaan},
year = {2026},
howpublished = {medRxiv preprint},
doi = {10.64898/2026.04.08.26350411}
}Added to the ChatPPG research hub