Deep-Tissue Hemodynamic Sensing: Comparing Impedance and Photoplethysmography for Wearable Blood Pressure Estimation
medRxiv preprint
Seamus Thomson et al., "Deep-Tissue Hemodynamic Sensing: Comparing Impedance and Photoplethysmography for Wearable Blood Pressure Estimation", doi:10.64898/2026.06.17.26355894, CC BY 4.0
TL;DR
Custom dual-modality wearable captures co-located IPG and PPG on the forearm in 261 participants (130 hypertensive, 131 not); >150,000 cardiac cycles; unsupervised archetype discovery of beat morphology. IPG shows up to 3 distinct morphology modes per person while PPG converges to uniform profiles. Ventral-forearm IPG amplitude has a significant BP-status effect (p = 0.024) absent in PPG; higher BMI reduces steep-upstroke IPG archetypes (p = 0.035).
- Form factor: wrist
Abstract
A bstract
The pursuit of continuous, cuffless blood pressure (BP) monitoring is constrained by the superficial sensing depth of photoplethysmography (PPG). Impedance plethysmography (IPG) offers deeper tissue penetration, but its comparative value over PPG remains unquantified at scale. In this comparative study of 261 participants (130 hypertensive, 131 non-hypertensive), we utilized a custom dual-modality wearable prototype to capture simultaneous IPG and PPG signals. Over 150,000 cardiac cycles were analyzed using an unsupervised archetype discovery pipeline to quantify beat-to-beat morphological heterogeneity. IPG resolved up to three distinct morphological modes per participant, whereas co-located PPG converged into highly conserved, uniform profiles. IPG captured specific signatures of pathological arterial remodeling and physiological habitus; ventral forearm IPG pulse amplitude exhibited a significant main effect for BP status (p = 0.024), a relationship absent in the co-located PPG signal. Furthermore, increasing body mass index (BMI) significantly attenuated the prevalence of steep-upstroke archetypes in IPG (p = 0.035), quantifying a likely damping effect of adipose tissue. Deep-tissue bioimpedance captures rich, heterogeneous hemodynamic signatures including arterial-dominant morphologies that are invisible to optical sensors. Transitioning from op-tical pulse wave analysis to bioimpedance-based models may offer a promising pathway for accurate wearable cardiovascular monitoring.
Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: http://creativecommons.org/licenses/by/4.0/
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Tasks
Signals
Keywords
How it works
Cite this paper
APA
Thomson, S., Jung, S., Pantelopoulos, A., Deshpande, A., Cai, L., Blanchard, E., Wasson, J., & Mukherjee, D. (2026). Deep-Tissue Hemodynamic Sensing: Comparing Impedance and Photoplethysmography for Wearable Blood Pressure Estimation. medRxiv preprint. https://doi.org/10.64898/2026.06.17.26355894
BibTeX
@misc{thomson2026deeptissue,
title = {{Deep-Tissue Hemodynamic Sensing: Comparing Impedance and Photoplethysmography for Wearable Blood Pressure Estimation}},
author = {Thomson, Seamus and Jung, Seobin and Pantelopoulos, Alexandros and Deshpande, Aniket and Cai, Lawrence and Blanchard, Emily and Wasson, Jackie and Mukherjee, Debanjan},
year = {2026},
howpublished = {medRxiv preprint},
doi = {10.64898/2026.06.17.26355894}
}Built from the hub record. Check it against the publisher before submitting.
Added to the ChatPPG research hub