Paperdoi:10.1038/s41746-025-01555-9

Ambulatory atrial fibrillation detection and quantification by wristworn AI device compared to standard holter monitoring

Venue listed2025Added Sep 26, 2026

  • SN

Authors: See npj Digital Medicine article (ClinicalTrials.gov NCT05899959)

npj Digital Medicine (2025)

TL;DR

Head-to-head of CardioWatch 287-2 wrist PPG + single-lead ECG AF algorithms vs 24-h Holter in 150 participants (mean age 64±12; 41% female), then 27 additional days of ambulatory wear. PPG and single-lead ECG algorithms achieved specificity ≥98% and sensitivity ≥95% for AF detection, and ~99% correlation for AF burden vs Holter. AF prevalence rose from 14.7% (24-h Holter) to 26.7% over 28-day CW2 wear — longer monitoring finds more paroxysmal AF.

  • Form factor: wrist

Abstract

Timely detection of atrial fibrillation (AF) is crucial for the prevention of serious consequences such as stroke and heart failure, yet it remains challenging due to its often asymptomatic or paroxysmal nature. Wearable devices with artificial intelligence algorithms offer promising solutions. AF detection by the CardioWatch 287-2 (CW2), a wrist-worn photoplethysmography (PPG) and single-lead ECG device, was compared to 24-h Holter. Patient compliance, AF prevalence and AF burden were evaluated for 27 additional days. Data from 150 participants (mean age 64 ± 12 SD; 41% female) were analysed. The CW2's PPG and single-lead ECG algorithms achieved a specificity ≥98% and sensitivity ≥95% for AF detection, and 99% correlation for AF burden, compared to 24-h Holter. AF prevalence increased from 14.7% (24-h Holter) to 26.7% (28-day CW2). Thus, the wrist-worn device showed promising performance in detecting AF and determining AF burden. The trial was registered on ClinicalTrials.gov (NCT05899959) on June 2, 2023.

Verbatim from Europe PMC (opens in a new tab) · Licence: cc by-nc-nd

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

AF / rhythm

Keywords

AFibwristclinical-validationHolterAF-burden

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG, single-lead ECG, Holter ref); Form factor (wrist). Method: Method. Target: AF / rhythm.SetupSignals: PPG, single-lead ECG, Holter refSignalsPPG, single-lead ECG, Holter refForm factor: wristForm factorwristMethodMethodTargetAF / rhythmAF / rhythm

Cite this paper

APA

article, S. N. D. M. (2025). Ambulatory atrial fibrillation detection and quantification by wristworn AI device compared to standard holter monitoring. npj Digital Medicine. https://doi.org/10.1038/s41746-025-01555-9

BibTeX

@article{article2025ambulatory,
  title         = {{Ambulatory atrial fibrillation detection and quantification by wristworn AI device compared to standard holter monitoring}},
  author        = {article, See npj Digital Medicine},
  year          = {2025},
  journal       = {npj Digital Medicine},
  doi           = {10.1038/s41746-025-01555-9}
}

Built from the hub record. Check it against the publisher before submitting.

Added to the ChatPPG research hub