Paperdoi:10.1002/epi4.70242
Pulse-based photoplethysmography quality assessment improves wearable seizure detection performance
Epilepsia Open (2026-06-17)
TL;DR
Applies a new pulse-based PPG quality assessment (PQA) that scores each pulse and drops bad ones before HR estimation, versus HR-outlier QA (HRQA) and no QA, on data from the PROMISE home trial of the NightWatch device in children with refractory epilepsy. 741 alarms (135 true seizures, 606 non-seizure) in 28 children (46% female, 9.1 ± 3.3 y). Averaged over HR-increase thresholds 30–80%: PQA raised sensitivity by +0.248, PPV +0.025, F1 +0.082 vs no QA (HRQA: +0.081 / +0.040 / +0.056).
- Form factor: upper arm (NightWatch)
Abstract
Objective
Wearable photoplethysmography (PPG) is increasingly used for seizure detection due to its ability to unobtrusively estimate heart rate (HR). However, the vulnerability of PPG to body movements often leads to unreliable HR estimates, compromising accurate seizure detection. Although this limitation is well recognized, methods for rejecting artifactual HR measures—and their impact on seizure detection performance—have not yet been studied.
Methods
We sourced data from the PROMISE trial, a home‐based study in children with refractory epilepsy that evaluated the performance of the NightWatch (NW) in detecting nocturnal motor seizures. We applied a novel pulse‐based quality assessment (PQA) method that evaluates the quality of each individual PPG pulse and excludes compromised ones prior to HR estimation. We compared the performance of PQA with two alternatives: HR‐based quality assessment (HRQA), rejecting extreme outliers in consecutive inter‐beat intervals, and no quality assessment (noQA). We assessed the impact of PQA, HRQA, and noQA on seizure detection performance by applying HR increase thresholds from 30% to 80% relative to the baseline. A permutation test with bootstrapping was applied for all statistical comparisons.
Results
We analyzed 741 NW alarms, comprising 135 true seizures and 606 non‐seizure events in 28 children (46% female; age: 9.1 ± 3.3 years). Averaged across all evaluated detection thresholds, both PQA and HRQA significantly improved seizure detection performance compared to noQA ( p
Significance
Implementing PPG quality assessment—particularly pulse‐based—improves the performance of wearable HR‐based seizure detection.
Plain Language Summary
This study aimed to improve seizure detection in wearable devices by introducing a new method for assessing the quality of the heart rate (HR) data derived from a light sensor called photoplethysmography (PPG). By applying our method to data from children with epilepsy, we showed that discarding low‐quality PPG data before HR estimation leads to better seizure detection, allowing more seizures to be correctly identified. This improvement is important because it enhances seizure detection performance through a software‐only upgrade.
Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: http://creativecommons.org/licenses/by-nc-nd/4.0/
Results on ChatPPG leaderboards
All leaderboardsNo results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.
Tasks & signals
Tasks
Signals
Keywords
How it works
Cite this paper
Shahbakhti, M., Bosch, A. T., Long, X., van Dijk, J. P., & Thijs, R. D. (2026). Pulse-based photoplethysmography quality assessment improves wearable seizure detection performance. Epilepsia Open. https://doi.org/10.1002/epi4.70242
Shahbakhti, Mohammad, Anemoon T. Bosch, Xi Long, Johannes P. van Dijk, and Roland D. Thijs. 2026. "Pulse-based photoplethysmography quality assessment improves wearable seizure detection performance." Epilepsia Open. https://doi.org/10.1002/epi4.70242.
Shahbakhti, M., Bosch, A. T., Long, X., van Dijk, J. P. and Thijs, R. D. (2026) 'Pulse-based photoplethysmography quality assessment improves wearable seizure detection performance', Epilepsia Open. doi: 10.1002/epi4.70242.
Shahbakhti, Mohammad, et al. "Pulse-based photoplethysmography quality assessment improves wearable seizure detection performance." Epilepsia Open, 2026. https://doi.org/10.1002/epi4.70242.
M. Shahbakhti, A. T. Bosch, X. Long, J. P. van Dijk, and R. D. Thijs, "Pulse-based photoplethysmography quality assessment improves wearable seizure detection performance," Epilepsia Open, 2026, doi: 10.1002/epi4.70242.
@article{shahbakhti2026pulsebased,
title = {{Pulse-based photoplethysmography quality assessment improves wearable seizure detection performance}},
author = {Shahbakhti, Mohammad and Bosch, Anemoon T. and Long, Xi and van Dijk, Johannes P. and Thijs, Roland D.},
year = {2026},
journal = {Epilepsia Open},
doi = {10.1002/epi4.70242}
}Added to the ChatPPG research hub