Authors: Oludotun Ode, David V. Anderson, Omer T. Inan

Licence: CC BY 4.0

TL;DR

Chest PPG is weakly perfused, so the pulsatile AC is small against a large DC baseline. Built-in coarse DC cancellation on commercial AFEs injects step artifacts. (1) Hardware–software co-design: the sensor's DC-cancellation input is overridden by smoothed software control; a real-time EWMA calibrator estimates and cancels DC. (2) Linear-predictive-coding artifact elimination for the remaining steps. (3) Softmax combiner fusing multiple PPG channels into one composite waveform.

  • Form factor: other

Abstract

Although chest-mounted photoplethysmography (PPG) has proven useful in clinical applications, it nevertheless remains vulnerable to dynamic-range challenges. Because the chest is less well perfused than traditional PPG measurement sites, it yields weaker pulsatile signals that may be masked by large DC baselines. Some commercial sensors feature coarse built-in DC cancellation, but they inject considerable uncompensated artifacts. First, this paper presents a hardware–software codesign using a commercial sensor with a DC cancellation feature, which is overridden by smoothed, software-defined control inputs. A real-time calibrator estimates and cancels DC components using an exponentially weighted moving average. Second, a linear predictive coding-based artifact-elimination method is introduced. Third, a multi-channel data fusion method using a softmax combiner is developed to improve PPG signal quality. System performance was evaluated using experimental data from six human subjects in a seated laboratory environment. Results indicate that the methods remove DC offsets and step artifacts, producing waveforms that are similar to ground-truth references. Quantitative results also show that the proposed combiner achieved a signal-to-noise ratio of 8.47±3.14 dB on its composite waveform, providing modest improvements (p=0.031) compared to traditional equal-gain (7.91±2.86 dB) and maximal-ratio (8.03±2.87 dB) combiners.

Verbatim from the publisher's DOI record (opens in a new tab) · Publisher licence: https://creativecommons.org/licenses/by/4.0/

Results on ChatPPG leaderboards

All leaderboards

No results from this paper on ChatPPG leaderboards yet. Browse the leaderboards.

Tasks & signals

Tasks

Signal qualitySignal denoising

Signals

Keywords

signal-qualityDC-cancellationdynamic-rangechest-PPGhardware-software-codesignlinear-predictive-codingartifact-removalmultichannel-fusionsoftmax-combinerGeorgia-TechInan

How it works

How Method works: Setup → Method → TargetSetup: Signals (PPG); Form factor (other). Method: Method. Target: Signal quality; Signal denoising.SetupSignals: PPGSignalsPPGForm factor: otherForm factorotherMethodMethodTargetSignal qualitySignal qualitySignal denoisingSignal denoising

Cite this paper

APA

Ode, O., Anderson, D. V., & Inan, O. T. (2026). Methods for Optimizing Photoplethysmography Signal Quality. https://doi.org/10.3390/s26206669

BibTeX

@misc{ode2026methods,
  title         = {{Methods for Optimizing Photoplethysmography Signal Quality}},
  author        = {Ode, Oludotun and Anderson, David V. and Inan, Omer T.},
  year          = {2026},
  doi           = {10.3390/s26206669}
}